BOVlE SPECIALIST ELECTROSURGlCAL UNIT
4. Common Power Settings
ELECTRICAL SPECIFICATIONS
This equipment is intended to operate at an
input voltage and frequency range as indicated on
the nameplate attached to the back of the unit.
lnput Voltage:
.................
Rated Input
115 Vac, 60 Hz.
Operational Range
.....
90-130 Vac, 50/60 Hz.
.....................
Input Current
3.2A @ 115V
. . . . . . . . . . . . . . . .
Operating Frequency
1.75 MHz
(Crystal Controlled)
.......
Open Circuit Voltage
4000V Peak to Peak
+20% -10%
................
Output Power'
75
+
10% Watts
(True RMS Power) into a 400 Ohm Non-
Inductive Load (Pure Cut Model)
50
+
10% Watts (True RMS Power)
into a 400 Ohm Load
(Cut with Hemostasis)
36
+
10% Watts (True RMS Power)
into a 400 Ohm Load
(Coagulation)
*All output power ratings are for 115
volts/60 Hz. input supply potential.
MECHANlCAL SPECIFICATIONS
Dimensions:
................................
Width
10.00"
...............................
Height
4.75"
................................
Depth
14.25"
.............................
Weight
20 lbs. net
(Unit and Accessories)
......................
Length of Power Cord
10'
. . . . . . . . . . . . . . . . .
Length of Footswitch Cord
10'
LIMITATIONS OF EQUIPMENT
This unit has been designed to operate indefi-
nitely at 20% operational duty cycle (10 sec. O N and
40 sec. OFF) at half power (5 Dial Setting) pure cut
into a 400 Ohm load.