Acronyms ..........................3 Introduction ..........................4 Mechanical Drawings ......................6 Setting Up The FGC100 ......................7 Operating the FGC100 ..................... 9 Performing Calibration of FluxGage Using FGC100 ............... 10 Calibration of the spectrum measurement and the total flux measurement ....11 Calibration of the total flux measurement only ............
1 Acronyms DUT – Device Under Test Note: The FGC100 calibration lamp complies with standard IEC 62471:2006 (First Edition) "Photobiological safety of lamps and lamp systems" and is classified as Low risk (Risk Group 1) when used as a standalone luminaire. When the FGC100 calibration lamp is connected to the FluxGage system it is classified as Exempt Group (RG0).
2 Introduction The FGC100 calibration system is used to calibrate the FluxGage LED luminaire system. The FGC100 includes: 12VDC adapter Control unit Lamp unit with protective cap D-type cable 100-230VAC mains cable Disk on key 100-230VAC D-type Mains...
Page 5
The control unit includes Front panel Fixed current source for the LED Temperature stabilization circuit General on/off switch Hour meter LED on/off push button LED temperature indicator light (Green/Red) Back Panel D-type connector ...
4 Setting Up The FGC100 Place the FluxGage box on a flat surface. Place the lid over the FluxGage. Make sure the lid is placed such that the round cover is over the spectrometer diffuser. Remove the round cover from the lid ...
Page 8
Figure 5. FGC100 connected to FuxGage FGC100 User Manual...
4.1 Operating the FGC100 Figure 6. FGC100 Control Unit - Front Turn the control unit on using the on/off switch (on the left) Press the LED on/off push button to turn on the lamp unit The temperature indicator will be red initially and will turn to green after a few minutes when the temperature of the lamp unit will be stabilized.
5.1 Calibration of the spectrum measurement and the total flux measurement Follow these steps: To calibrate the FluxGage click ‘Calibration’ on the top banner of the FluxGage software. In the setting area, enter the total luminous flux of the FGC100. The value is provided with the Disk-On-Key included in the FGC100.
5.3 Calibration of the spectrometer only Follow these steps: To calibrate the FluxGage click ‘Calibration’ on the top banner of the FluxGage software. In the setting area, enter the total luminous flux value zero Click ‘Open Light File’ and select the spectral calibration file provided with the Disk-On- Key included in the FGC100.