  • Page 1 FluxGage FluxGage ™ ™ Calibration Unit FGC100 User Manual Ophir Optronics Solutions Ltd.
  • Page 2: Table Of Contents

    Acronyms ..........................3 Introduction ..........................4 Mechanical Drawings ......................6 Setting Up The FGC100 ......................7 Operating the FGC100 ..................... 9 Performing Calibration of FluxGage Using FGC100 ............... 10 Calibration of the spectrum measurement and the total flux measurement ....11 Calibration of the total flux measurement only ............
  • Page 3: Acronyms

    1 Acronyms DUT – Device Under Test Note: The FGC100 calibration lamp complies with standard IEC 62471:2006 (First Edition) "Photobiological safety of lamps and lamp systems" and is classified as Low risk (Risk Group 1) when used as a standalone luminaire. When the FGC100 calibration lamp is connected to the FluxGage system it is classified as Exempt Group (RG0).
  • Page 4: Introduction

    2 Introduction The FGC100 calibration system is used to calibrate the FluxGage LED luminaire system. The FGC100 includes:  12VDC adapter  Control unit  Lamp unit with protective cap  D-type cable  100-230VAC mains cable  Disk on key 100-230VAC D-type Mains...
  • Page 5 The control unit includes Front panel  Fixed current source for the LED  Temperature stabilization circuit  General on/off switch  Hour meter  LED on/off push button  LED temperature indicator light (Green/Red) Back Panel  D-type connector ...
  • Page 6: Mechanical Drawings

    3 Mechanical Drawings FGC100 User Manual...
  • Page 7: Setting Up The Fgc100

    4 Setting Up The FGC100  Place the FluxGage box on a flat surface.  Place the lid over the FluxGage. Make sure the lid is placed such that the round cover is over the spectrometer diffuser.  Remove the round cover from the lid ...
  • Page 8 Figure 5. FGC100 connected to FuxGage FGC100 User Manual...
  • Page 9: Operating The Fgc100

    4.1 Operating the FGC100 Figure 6. FGC100 Control Unit - Front Turn the control unit on using the on/off switch (on the left) Press the LED on/off push button to turn on the lamp unit The temperature indicator will be red initially and will turn to green after a few minutes when the temperature of the lamp unit will be stabilized.
  • Page 10: Performing Calibration Of Fluxgage Using Fgc100

    5 Performing Calibration of FluxGage Using FGC100 FGC100 User Manual...
  • Page 11: Calibration Of The Spectrum Measurement And The Total Flux Measurement

    5.1 Calibration of the spectrum measurement and the total flux measurement Follow these steps:  To calibrate the FluxGage click ‘Calibration’ on the top banner of the FluxGage software.  In the setting area, enter the total luminous flux of the FGC100. The value is provided with the Disk-On-Key included in the FGC100.
  • Page 12: Calibration Of The Spectrometer Only

    5.3 Calibration of the spectrometer only Follow these steps:  To calibrate the FluxGage click ‘Calibration’ on the top banner of the FluxGage software.  In the setting area, enter the total luminous flux value zero  Click ‘Open Light File’ and select the spectral calibration file provided with the Disk-On- Key included in the FGC100.