Orbis AirSTAR Operation Manual page 44

Table of Contents

Advertisement

End conditions:
End conditions determine the end of the test. According to the type of end
condition the test result will be reported.
Press "+" to add an end condition. Multiple conditions can be added later.
Image 21
Pressing the �� symbol gives a few different end condition options (image 21):
• FILL: if during any aspiration the pipette is not entirely filled within the
set number of seconds, the end of the test has reached.
• EMPTY: if during/after any aspiration the pipette is not entirely empty
within the set number of seconds, the end of the test has reached.
• EMPTY before next aspiration: if during/after any aspiration the pipette is
not entirely empty before the next aspiration is due, the end of the test
has reached with the remark "not empty before next aspiration". This is
also considered CFPP2.
• SAMPLE: the test ends when the set sample temperature is reached.
• BATH: the test ends when the set bath temperature is reached (applies to
linear / ramping cooling profiles only)
After entering the settings, press DONE to create the step and exit the menu.
�� Auto number = 0:
If this line still appears in your latest version of the software: please ignore it as
it is a leftover line from an old feature that is not present nor relevant anymore.