IRtec Rayomatic 14 MK2 Process IR temperature transmitters Instruction Manual MM850618 ed. 01e...
Page 2
This manual is a complete “USER GUIDE”, providing step-by-step instructions to operate the instrument in each of its designed functions. Eurotron has used the best care and efforts in preparing this book and believes the information in this publication are accurate. The Eurotron products are subjected to continuous improvement, in order to pursue the technological leadership;...
As the infrared sensor Eurotron IRtec Rayomatic 14 consists of two pieces (miniature sensing head with separate electronics) it can easily be installed in a variety of applications, especially if industrial environment requires space-saving installation.
Instruction manual MM850618 ed. 01e Specifications Spectral range: 8 to 14 µm Response time: 150 ms (95 %) - HS: 12ms (90%), 8 ms (63%) Accuracy: ±1 % or ±1 °C whichever is greater (T> -20 °C; ambient temperature +23 °C ±5 °C) Repeatability: ±0,5 % or ±0,5 °C whichever is greater Temperature Coefficient: ±0,05 %/ K or ±0,05 K/ K whichever is greater Signal processing: peak hold, valley hold, average...
Instruction manual MM850618 ed. 01e 1.3.1 CloseFocus-Optics The optional CF-lens allows the measurement of small objects. The minimum spot size depends on the used sensing head: 0.6 mm@ 10 mm with 25:1-head 0.8 mm@ 10 mm with 14:1-head 2.5 mm@ 25 mm with 2:1-head The following optical charts show the diameter of the measuring spot in dependence on the distance between measuring object and sensing head.
PROFI Profibus RELAY Relais interface (2x optically isolated) vent/Data Logging Table D Report of Calibration None Eurotron NIST or EA traceable with data 1.4.1 Accessories Ordering Code Description EE290148 Standard Air Purge EE290151 Laminar flow Air purge (for severe applications) EE290147 90°...
Instruction manual MM850618 ed. 01e PHYSICAL DESCRIPTION IRtec Rayomatic radiation thermometers are designed to use the most advanced sensor technology. The head external case is made of stainless steel with M12x1 thread compatible with a wide range of accessory. The case is equipped with a single lens optic system, optic filter, sensor, and electronics for signal conditioning.
Instruction manual MM850618 ed. 01e PRINCIPLE OF OPERATION Basics of Infrared Thermometry: Depending on the temperature each object emits a certain amount of infrared radiation. A change in the temperature of the object is accompanied by a change in the intensity of the radiation. For the measurement of “thermal radiation”...
Inspect the instrument from scratches, dents, damages to case corners etc. which may have occurred during shipment. If any mechanical damage is noted, report the damage to the shipping carrier and then notify Eurotron directly or its nearest agent, and save the damaged packaging for inspection.
Instruction manual MM850618 ed. 01e INSTALLATION IRtec Rayomatic installation is quite easy. The cylindrical stainless steel case is externally threaded and allows quick installation of all available accessories. The following procedure should help you during the installation procedure of your IRtec Rayomatic thermometer.
Page 14
Instruction manual MM850618 ed. 01e Fig. 4-1: Open IRtec Rayomatic 14 electronic box with screw terminals Designation: +8 ... 36VDC Power supply Ground (0V) of power supply Ground (0V) of internal in- and outputs OUT-TC Analog output thermocouple (J or K) OUT-mV/mA Analog output object temperature (mV or mA) Analog output head temperature (mV)
Instruction manual MM850618 ed. 01e CAUTION: After exchanging a head the calibration code of the new head must be entered into the electronics 5.1.2 Outputs The IRtec Rayomatic 14 has two output channels. The channel Out-mV/mA 1 is used for the object temperature and for the alarms.
Page 16
Instruction manual MM850618 ed. 01e Fig. 5-1: Dimensions IRtec Rayomatic 14 head and electronic box In order to align the bracket with the target, look through the support hole (without the thermometer) and find the best position. It's important for the optic path to be free from any obstruction (steam, dust, etc.). In particular, when the surface is targeted through a hole, the hole diameter has to be big enough according to the distance from the instrument.
Page 17
Instruction manual MM850618 ed. 01e M12x1 EE290148 Standard air purge Brackets All sensors come with a 1 m cable and a mounting nut. The sensor can be installed on using the optional bracket or equivalent system. Ø18.5 R15±0.1 Ø3.2 EE290147 90° adjustable bracket...
By head sensor Baud rate 19200 Bps IRtec Rayomatic 14 MK2 display Use this key escape from the parameter you are setting (MODE key) Use this key to move up into the parameter menu and/or to increase the parameter you are setting...
Page 20
Instruction manual MM850618 ed. 01e The main MENU shows the following parameters: Use the ENTER key to enter the desired function. See the flow chart related to each function...
Instruction manual MM850618 ed. 01e 6.1.2 B: Laser settings 6.1.3. C: Password setting The default password is “333”. It can be modified by menu and this feature let the operator to modify the instrument working settings. If the typed password is wrong (wrong number) the instrument settings can not be modified.
Instruction manual MM850618 ed. 01e 6.1.4 D: Alarms settings The backlight of the display and the relay activation the optional card can be set in this menu. The BLU alarm is normally closed (active) and the RED alarm is normally open (non active). The alarms are defined as “alarms of high”;...
Instruction manual MM850618 ed. 01e Regarding the lower and upper signal limit (RANGE), if the output is set in voltage, the minimum and maximum level are set, for example for the range from 0 to 10 Volts, the set level will be 0V and 10V. Regarding the lower and upper temperature range limit (inactive to the TcJ and TcK outputs), the set parameter are referred to the upper and lower extreme limit for the temperature representation on the output port.
Page 27
Tab. 6-1: Adjustment modes Head exchange: Sensing heads of the IRtec Rayomatic 14 MK2 series are exchangeable. Every head has a specific calibration code, which is printed on the head cable. For a correct temperature measurement and functionality of the IRtec Rayomatic 14 MK2 this calibration data must be stored in the electronic box.
YOU CAN UPGRADE YOUR INSTRUMENT THROUGH ITS SERIAL PORT Factory default setting: To set the IRtec Rayomatic 14 MK2 back to the factory default settings, please press at first the MODE-key and then ENTER-key and keep both pressed for approx. 3 seconds. The display shows RESET for confirmation.
Instruction manual MM850618 ed. 01e 6.2.1.2 Edge mode With this mode the record will be saved when one of the set events is true. The events are alarm based, so this logging mode is only possible if at least one alarm is enabled. The alarms available are: Output channel 1 Output channel 2 Alarm 1 (blue display backlight)
Page 30
Instruction manual MM850618 ed. 01e • Select “LOG” and press ENTER The Block 1 option will be displayed • Press ENTER: the option will blink • Select the desired block between B1 and B2 by using the up or down keys •...
Instruction manual MM850618 ed. 01e 6.2.2.1 Cleaning one block memory HE BLOCK MEMORY WILL BE DELETED BUT THE BLOCK CONFIGURATION WILL BE NOT DELETED NOTHER DATA LOGGING WITH THE SAME CONFIGURATION CAN BE PERFORMED • Select “LOG” and press ENTER The Block 1 option will be displayed.
Time function facilities are useful to retain the signal levels that are more likely to represent the true target temperature. Eurotron IR thermometers can be set with the signal processing facilities by using IRSetup configuration Software. For to set the signal processing parameters Press in IR set up software the button advanced:...
Instruction manual MM850618 ed. 01e Ambient temperature compensation: In dependence on the emissivity value a certain amount of ambient radiation will be reflected from the object surface. To compensate this impact, the IRtec Connect software provides the feature Ambient control (middle of the window Signal processing): internal (Head): The ambient temperature will be taken from the head-internal Pt1000 probe...
Instruction manual MM850618 ed. 01e In the mode Valley Hold the IRtec unit waits for ascending signals. If the signal ascends the algorithm maintains the previous signal valley (minimum) for the specified hold time. Advanced Peak hold: In the mode Advanced Peak Hold the IRtec unit waits for local peak values. The signal must drop below the Threshold value to detect the next peak (which must be ≥...
Page 35
Instruction manual MM850618 ed. 01e Fig. 7-10: Setup dialog – window Output signals Output channel 1: The output channel 1 is used for the object temperature. To define it as alarm output, activate digital. With the selection Normally open/ closed you can define the output as High- or Low-alarm. Edit the threshold in the menu box Alarm to set the trigger of the alarm.
Instruction manual MM850618 ed. 01e Advanced Settings If the setup dialog is open, the window Advanced settings allows the setting of the following parameters: Head-parameters: This function allows the exchange of sensing heads. The 12-digit code contains the calibration data of the head.
8.1. USB-interface: If you use the USB-interface, the IRtec Rayomatic 14 MK2 needs no external power supply for operation – it will be powered by the USB-interface. If an external power supply has already been installed, this will have no effect on the functionality of the IRtec Rayomatic 14 MK2.
Instruction manual MM850618 ed. 01e Relais-interface: After installing the interface board as described above please connect the electrical circuit with the terminal block. A red LED shows the closed current switch. The relais board provides two fully isolated switches, which have the capability to switch max.
Page 39
Instruction manual MM850618 ed. 01e 2. Remove the cable cap 1. Unscrew the cable cap Unscrew the first screw on the top of the cover Unscrew the second screw on the top of the cover Open the top cover Unscrew and remove all the thin cables from the card connector Unscrew and remove all the thin cables from the Unscrew the card fixing screw located on the left...
Page 40
Instruction manual MM850618 ed. 01e 10. Hand the cable with one hand and put the card Pass the cable through the hole into its housing facing up with the screw hole located accordingly to the screw hole of the electronic box. Be sure that the multi-pin connector of the other side of the card will correctly gently...
Instruction manual MM850618 ed. 01e 15. Tighten the nut 16. Close the front cover and tighten it with the two screws. NOTE: R EMEMBER TO ALSO CONNECT THE POWER SUPPLY BEFORE TO CLOSE THE FRONT COVER IF NOT YET CONNECTED Connecting to a PC The instrument can operate with a personal computer through the mini-USB serial port.
Page 42
Instruction manual MM850618 ed. 01e before connecting to Windows Update. Windows™ XP SP2 can have the parameter already set. To verify this, open the Control Panel, select “System”, select the Hardware tab and click “Windows Update”. Please verify that this option is DISABLED. •...
Instruction manual MM850618 ed. 01e • Select the folder where the driver is present on the CD: • D:\\EE26032-SW-DRIVER USB-Serial FTDI, then select the folder referred to your operative system and click “OK”. The driver will be automatically installed on your system. NOTE: I -WHQL INDOWS...
Page 44
Instruction manual MM850618 ed. 01e • Double click "System" to open it. • Select "Hardware"...
Page 45
Instruction manual MM850618 ed. 01e • Select "Device Manager" • Click on the "+" symbol of the "COM & LPT ports" category...
Page 46
Instruction manual MM850618 ed. 01e • The installed COM port will be displayed (for example COM7) • Select the port • Right click on it and select "Properties" to open the port properties • Select "Port Settings" to open the port communication properties...
Page 47
Instruction manual MM850618 ed. 01e • The communication speed of the selected COM port (in Bits per second) will be displayed on the first field WARNING REMEMBER TO SET BOTH THE CORRECT PORT NUMBER AND THE PORT SPEED IN BAUD ON THE SOFTWARE SETTINGS THAT YOU ARE RUNNING OTHERWISE YOU WILL HAVE ANY COMMUNICATION BETWEEN THE INSTRUMENT AND THE COMPUTER...
Instruction manual MM850618 ed. 01e IRSETUP SOFTWARE For over 20 years, Eurotron has combined ISO 9001 requirements, infrared thermometry experience, and best practices to develop a technologically advanced software applications. The last generation of Infrared sensors and transmitters include advanced electronic technologies. This technology allows to improve the standard temperature measurements with temperature processing capability.
Instruction manual MM850618 ed. 01e Monitor Signal Processor Mode PC Com Port selection Mathematical function Parameters settings Range settings Alarm settings Emissivity settings Load configuration from thermometer Press the icon Load to load the configuration data from the thermometers and to □...
Page 50
Instruction manual MM850618 ed. 01e before starting the reading, press Load and the software will find the instrument connected and the below page will show up:...
Instruction manual MM850618 ed. 01e Press Start and the software shows the measured temperature: GRAPHIC MODULE To enter the graphic software utility for the first time, to insert a registration code needed to enhance the Setup Software with this feature. From Rayomatic Setup Software, select “About”...
Page 52
Instruction manual MM850618 ed. 01e for more information to use the software read the relative manual IRsetup.
The air filters cleanliness must be checked at regular intervals. Our suggestion is to check it every day, then, according to your experience, find a correct time interval. If the optic system reaches temperatures higher than the working one, it has to be recalibrate in Eurotron laboratories.
This warranty applies to the original purchaser only. If the unit should malfunction, it must be returned during the warranty period, transportation prepaid, to Eurotron for evaluation. Upon examination, if the unit is found to be defective it will be repaired or replaced at no charge.
Instruction manual MM850618 ed. 01e Visual Alarms Alarm 1 Alarm 2 Alarm status Blue Green Preset normally temperature normally temperature Close Below Open Below Standard visual alarms Close Below Open Above Standard visual alarms Close Above Open Below Standard visual alarms Close Above...
Instruction manual MM850618 ed. 01e How to determine an object emissivity Emissivity is the measure of an object ability to absorb, transmit, and emit infrared energy. It can have a value from 0 (shiny mirror) to 1.0 (blackbody). If a value of emissivity higher than the actual one is set, the output will read low, provided that the target temperature is above the ambient one.