Page 2
Except as expressly provided herein, Beckman Coulter, Inc. makes no other warranty, whether oral or written, expressed or implied, as to any matter whatsoever, including but not limited to those concerning merchantability and fitness for a particular purpose, nor is freedom from any patent owned by Beckman Coulter, Inc. or by others to be inferred.
Warranty and Returned Goods Requirements All standard Beckman Coulter, Inc. policies governing returned goods apply to this product. Subject to the exceptions and upon the conditions stated below, the Company warrants that the products sold under this sales agreement shall be free from defects in...
Please pay special attention to the specific safety information associated with these symbols. WARNING: If the equipment is used in a manner not specified by Beckman Coulter, Inc., the protection provided by the equipment may be impaired. Warning and Caution Definitions All Warnings and Cautions in this document include an exclamation point, a lightning bolt, or a light burst symbol framed within a triangle.
To prevent electrically related injuries and property damage, properly inspect all electrical equipment prior to use and immediately report any electrical deficiencies. Contact a Beckman Coulter Service Engineer for any servicing of equipment requiring the removal of covers or panels unless specifically stated otherwise.
Safety Information Chemical and Biological Safety Normal operation of the Shaking Peltier Device may involve the use of materials that are toxic, flammable, or otherwise biologically harmful. When using such materials, observe the following precautions: • Handle infectious samples according to good laboratory procedures and methods to prevent the spread of disease.
Beckman Coulter Service Engineer or returning parts to Beckman Coulter for repair. Beckman Coulter will NOT accept any items which have not been decontaminated where it is appropriate to do so. If any parts are returned, they must be enclosed in a sealed plastic bag stating that the contents are safe to handle and are not contaminated.
The WARNINGs and CAUTIONs found within this document are listed below. WARNING: If the equipment is used in a manner not speci- fied by Beckman Coulter, Inc., the protection provided by the equipment may be impaired. CAUTION: Use only the cables that were included with the Shaking Peltier Device.
Figure 1-1. Shaking Peltier Device....................1-1 Figure 1-2. Installing new feet on Biomek 3000 instrument ............1-4 Figure 1-3. Attaching the mounting bracket to the Biomek 3000 ..........1-5 Figure 1-4. Adjusting the feet on the mounting bracket .............. 1-6 Figure 1-5.
Shaking Peltier Device Overview The Shaking Peltier Device (Figure 1-1) is a labware positioner with microplate shaking and heating and cooling capabilities from a connected controller. Each Peltier may be equipped with an adapter plate which allows for different labware types to be used with the Peltier device.
3000 in Biomek Software. Refer to the original manuals from the device manufacturer, Inheco, for more information not included in this manual: • Thermoshake Heating and Cooling Unit Manual • TEC Control 96 User’s Manual • Biomek 3000 User’s Manual • Biomek Software User’s Manual Beckman Coulter, Inc.
The Shaking Peltier Device is installed on a special mounting bracket that is installed on the left side of the Biomek 3000 deck in place of the side module. The Shaking Peltier Device may be installed on either of two positions on the mounting bracket.
1. Tilt the Biomek 3000 instrument on its back feet. 2. While one person holds the Biomek 3000, a second person screws a new foot into each of the threaded holes adjacent to the two front feet (Figure 1-2).
To mount the mounting bracket on the left side of the instrument: 1. If the Biomek 3000 Side Module is installed on the left side of the Biomek 3000 instrument, remove it (refer to the Biomek® 3000 Laboratory Automation Workstation User’s Manual, Section 1.3.4.1, Removing Installed Side Modules).
Figure 1-4. Adjusting the feet on the mounting bracket 2. Using a bubble level, adjust the feet on the Biomek 3000 instrument to make sure the deck is level. Turn the feet to the left to raise the feet or to the right to lower them.
Installing the Shaking Peltier Device The mounting bracket has two positions in which devices or positioners may be placed. Either a Static Peltier Device, Shaking Peltier Device, or Biomek 3000 Position Holder, in any combination, may be installed in these two positions.
7. Connect the power cable from the main power port on the back of the Watlow controller to a power supply (Figure 1-6). 8. Power on the Watlow controller using the power switch on the back of the Watlow controller just above the power cable (Figure 1-6). Beckman Coulter, Inc.
Shaking Peltier Device 1.2.5 Installing Adapter Plates Before using the Shaking Peltier Device, one of three available adapter plates must be installed. Adapter plates should be chosen based on what type of microplates are used on the Shaking Peltier Device. Each adapter plate provides a different heating and cooling profile that is optimized for specific types of plates or applications.
2. Using a 1.5 mm Allen wrench, adjust the corner guide screws to control how tightly labware is held in place. The microplate should fit snugly in the position but still move freely in and out of the plate holder. Beckman Coulter, Inc.
Setting Up the Shaking Peltier Device in Biomek Software After physically installing the Shaking Peltier Device on the Biomek 3000 Laboratory Automation Workstation, it is must be set up to use in Biomek Software. Setting up the Shaking Peltier Device in Biomek Software includes: •...
7. Select Simulation only if using the software on a computer that is not attached to an instrument. This allows methods to be created without the controller attempting to communicate with the device. 8. Choose Next to install the Shaking Peltier SILAS Module. The installation begins. Beckman Coulter, Inc.
Shaking Peltier Device 9. When the installation completes, Figure 1-10 appears. Figure 1-10. Installation of Shaking Peltier SILAS Module complete 10. Choose Finish to close the installation program. The Shaking Peltier SILAS Module has been installed successfully. Shaking Peltier Device Integration Manual...
Adding the Shaking Peltier Device to the Deck in the Deck Editor After physically installing the Shaking Peltier Device on the deck of the Biomek 3000 instrument, it must be added to the deck in the Deck Editor. To add the Shaking Peltier Device to the Deck Editor: 1.
Shaking Peltier Device 6. Choose More to expand Position Properties (Figure 1-12). Figure 1-12. Position Properties for Shaking Peltier Device — expanded 7. In Device, select the ShakingPeltier device. 8. Choose Manual Teach to frame the position (refer to Section 1.4, Framing Instructions).
The Shaking Peltier Device must be framed before using it in a method. Framing allows the Biomek 3000 to accurately move labware to and from the Shaking Peltier Device and pipette to and from microplates positioned on the Shaking Peltier Device.
(Figure 1-13). Note: Make sure that the arrow marked FRONT on the framing tool is pointing towards the front of the Biomek 3000 deck. 3. Release the button when the framing tool is properly seated. Shaking Peltier Device Integration Manual...
Manual Teach is a wizard-type interface that is used to manually frame deck positions on the Biomek 3000 workstation. Using Manual Teach helps to ensure that the tips can access the wells without causing any damage to the tips, tools, or labware positioners.
Figure 1-15. Position Properties for a Shaking Peltier Device 4. In Name, verify that the position is assigned a unique name. 5. Choose Manual Teach. Biomek 3000 Manual Teaching opens with a Warning (Figure 1-16). Note: On the left side of Manual Teaching, a list of steps required to complete the teaching process is displayed.
7. Choose Next to continue to the next screen. Biomek 3000 Manual Teaching displays Load Tool. Figure 1-18. Loading tool during manual teach 8. If not already installed, install the framing tool on the Biomek 3000 head assembly (refer to Section 1.4.1, Attaching the Manual Framing Tool).
(from step 9). This selection must match the tip type physically loaded on the framing tool. 11. Choose Next to continue to the next screen. Biomek 3000 Manual Teaching displays Select Target (Figure 1-19). Figure 1-19. Selecting target to manual teach position 12.
Z-axis until it is approximately 1 mm above the top of the microplate. Note: Since tip height is set in the next step in the Biomek 3000 Manual Teaching process, it is safe to move the head assembly to any height to make aligning the tips with the target position easier.
Shaking Peltier Device 1.4.2.1 Using the Graphic Alignment Tool The graphic alignment tool (Figure 1-21) is used to instruct the Biomek Software of the current position of the framing tool tip to the center of well A1 on the microplate. The Biomek Software uses the information it is given to move the head assembly so the framing tool tip is directly above the center of the well.
Using the Delta Value and Directional Buttons). 4. Once the tip is aligned with the center of well A1 on the microplate, choose Next to continue. Biomek 3000 Manual Teaching displays Teach Z (refer to Section 1.4.2.3, Framing the Z-Axis).
Page 45
(refer to Section 1.4.2.1, Using the Graphic Alignment Tool). 4. Once the tip is aligned with the target position on the microplate, choose Next to continue. Biomek 3000 Manual Teaching displays Teach Z (refer to Section 1.4.2.3, Framing the Z-Axis).
The directional buttons displayed in Manual Teach parallel the keys on the numeric keypad. More specifically, Fwd. correlates to the ‘1’ on the numeric keypad, while Down is found on the ‘2’, Left is found on the ‘4’, Right on ‘6’, Up on ‘8’, and Back on ‘9’. Beckman Coulter, Inc.
Page 47
Shaking Peltier Device 3. Once the framing tool tip is at the same height as the top of the well, select Finish. The pod moves up to its maximum height in the Z-axis, Manual Teaching closes, and Position Properties appears (Figure 1-15). 4.
8. Press the green button to enter the User menu. 9. Use the arrow keys to scroll to CAL1. 10. Enter the calibration offset value CAL determined in step 5 above. ∞ 11. Press the (infinity symbol) button to set the calibration offset. Beckman Coulter, Inc.
Shaking Peltier Device Using the Shaking Peltier Device in a Method Using the Shaking Peltier Device in a Biomek method requires using the Peltier Step step that is displayed on the Peltier step palette (Figure 1-24). The Move Labware step from the Basic Step Palette is used to move labware to and from the Shaking Peltier Device (refer to the Biomek Software User’s Manual, Section 16.4, Move Labware Step).
Note: The incubation timer starts when the step is executed, even if it is not at the specified temperature. 5. For Incubate, to set a specified temperature, ensure Set Temperature to is checked and enter a temperature in degrees Celsius (4.0 to 50.0). Beckman Coulter, Inc.
Page 51
Shaking Peltier Device 6. For Incubate, select the Shake checkbox to shake the microplate during incubation. 7. For Incubate or Start Shaking, in Shake at, enter the rate of shaking in rotations per minute (RPM). 8. For Incubate, select Continue Shaking after incubation until device receives Open command to continue shaking the microplate after the incubation is complete until the position is accessed for gripping or pipetting.
4. In Communications Port, select the comm port on the Biomek Controller PC to which the Shaking Peltier Device is connected. 5. Choose OK to save the changes and close ShakingPeltier Options, or Cancel to close ShakingPeltier Options without saving changes. Beckman Coulter, Inc.
Shaking Peltier Device 1.7.2 Performing Actions in the Device Editor To manually control a Peltier action outside a method: 1. From the Instrument menu, choose Device Editor. Device Editor appears. 2. In Device, choose ShakingPeltier. 3. Choose Action Commands. ShakingPeltier appears (Figure 1-28). Figure 1-28.
Page 54
This shake sequece repeats until the incubation is complete, the position is accessed, or a Stop Shaking command is sent. 14. Choose Run Now to perform the configured Action. 15. Select Update after completing the action configuration. The ShakingPeltier module action configuration closes. Beckman Coulter, Inc.
Shaking Peltier Device Preventive Maintenance WARNING: ELECTRICAL SHOCK HAZARD! Checking and changing the coolant requires removing the covers from the Shaking Peltier Device, exposing internal electrical components. Remove covers with care and use caution when servicing the device. Check the coolant fluid level every seven weeks and refill as necessary. Refer to the Thermoshake Heating and Cooling Unit Manual for specific instructions.
Refer to the Thermoshake Heating and Cooling Unit Manual and the TEC Control 96 User’s Manual for additional troubleshooting information. Do not attempt to repair the unit without first contacting a Beckman Coulter Service Engineer. Table 1-1. Troubleshooting the Shaking Peltier Device...