SOMNI 3 Operation Manual

Precision medical vaporizer designed for veterinary use only

Advertisement

Quick Links

SOMNI 3
VETERINARY ANESTHETIC
VAPORIZER OPERATIONS MANUAL
SOMNI PRODUCT MANUAL

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 3 and is the answer not in the manual?

Questions and answers

Summary of Contents for SOMNI 3

  • Page 1 SOMNI 3 VETERINARY ANESTHETIC VAPORIZER OPERATIONS MANUAL SOMNI PRODUCT MANUAL...
  • Page 2: Table Of Contents

    TABLE OF CONTENTS User Responsibility Warnings and Cautions Principles of Operation Examination and Preparation For Use Installation Operating Instructions Filling Instructions Funnel Fill Draining Instructions Funnel Fill Specifications Warranty and Service...
  • Page 3: User Responsibility

    USER RESPONSIBILITY The SOMNI 3 is a precision medical vaporizer designed for Veterinary Use Only. This vaporizer was designed to mix the vapors of specified liquid anesthetic gas agents with anesthetic delivery gases. The vaporizer will continue to provide reliable performance only if the manufacturer’s operating and maintenance instructions are followed.
  • Page 4: Warnings And Cautions

    If unit is accidentally tipped on its side, please call the manufacturer for specific instructions. • Do not have vaporizer serviced by anyone other than a SOMNI Scientific authorized service center. • Do not operate vaporizer prior to leak testing the anesthetic equipment, ensuring secure connections to prevent unnecessary anesthetic exposure •...
  • Page 5: Principles Of Operation

    PRINCIPLES OF OPERATION VAPORIZER SUMP AND VALVE ASSEMBLY This anesthetic vaporizer is comprised of a vaporizing chamber and duct system (located within the sump cover), rotary valve and concentration dial. The Concentration Setting Dial concentration dial is connected to the rotary valve underneath. The rotary valve contains ducts and a vapor control channel.
  • Page 6: Examination And Preparation For Use

    1. Examine shipping carton for signs of external damage. 2. Remove contents from carton and inspect for visible damage such as dents or missing parts. 3. If damage or missing parts are discovered or suspected, notify customer service immediately at 1-877-637-3625. 4. Check that control dial operates freely.
  • Page 7: Installation

    The SOMNI 3 is supplied with mounting hardware consisting of a spacer and 3-6mm bolts. While supporting the weight of the vaporizer, install the spacer between the vaporizer and the mounting surface and secure the vaporizer using the mounting bolts provided.
  • Page 8 INSTALLATION IMPORTANT! Vaporizer Spacer Placement • The direction of gas flow must be from “inlet to outlet” (i.e. from left to right) when viewing the vaporizer from the front. • Ensure the liquid (which may accumulate in the breathing circuit or the CO2 absorber) can not enter the vaporizer while: -in use -during disassembly of the circuit...
  • Page 9: Operating Instructions

    1. Observe all instructions and warnings on the vaporizer 2. Fill only with agent indicated with vaporizer in OFF position 3. Perform a leak test prior to first use (See instructions below- “10 Second Test”) 4. Depress locking button to turn dial from OFF position 5.
  • Page 10 10 SECOND PRESSURE TEST Before each use, “leak test” the machine, and also ensure that the waste gases have a clear way through your evacuation system. Before each use, “leak test” the anesthesia system and ensure the waste gases have a patent way through the evacuation system.
  • Page 11: Filling Instructions Funnel Fill

    Drain Port agent other than specified could prove to be dangerous to a patient. 3. Verify that the drain valve on the right side of the fill assembly is closed by turning clockwise until finger tight. 4. Remove the funnel cap and pour agent slowly into opening.
  • Page 12: Draining Instructions Funnel Fill

    2. Hold an empty, properly labeled container under the tube. 3. Open the drain valve located on the right side of the fill assembly. After all liquid is drained, close the valve finger tight to seal. 4. Discard used anesthetic agent per proper protocol.
  • Page 13 ASSURING PERFORMANCE OF YOUR VAPORIZER To assure the continued performance of your vaporizer, the manufacturer recommends full factory preventative maintenance be performed every 2 years. Accurate and efficient anesthetic gas delivery is a primary consideration in patient care. Anesthetic agent vapors are extremely potent, and a very small error in concentration could be hazardous.
  • Page 14: Specifications

    SPECIFICATIONS Approximate rate of agent consumption = 3 x 2 x 6 = 36 ml/hour. The above figures are approximate and intended for clinical Calibration guidance only. Figures will vary depending on flow meter type Vaporizers are calibrated at 21° C. The variation in output (and other varying factors).
  • Page 15 The valve design and temperature compensation system of causing the magnitude and rate of cycling pressure the Somni 3 vaporizers work to reduce the effects to levels fluctuations to decrease. considered not significant by clinical standards.
  • Page 16 SPECIFICATIONS Effects of Variables Pressure Vaporizers are graduated in v/v percentage at 760 mm Hg. Temperature If the ambient pressure changes the v/v % will change so Temperature variation effects are typically negligible that at an ambient pressure P mm Hg the delivered percentage (D % v/v)- at common dial setting and ambient temperature combinations.
  • Page 17: Warranty And Service

    THIS IS THE EXCLUSIVE from defects in materials and workmanship under normal REMEDY. SOMNI shall not be liable for and hereby disclaims any use, if maintained in accordance with SOMNI’s guidelines direct, incidental, consequential or special damages or delays,...