Vacuum Mixers Wamix-Classic IV and Wamix-Injector IV Wamix-Classic IV Wamix-Injector IV USER MANUAL...
Page 2
Dear customer, Thank you for choosing a product from the Wassermann range. Wassermann Dental- Maschinen incorporates the highest standards of quality and the latest technology. In order to enjoy maximum performance and years of trouble-free operation, please read this user manual carefully before you connect this device and start work, and operate the device according to the recommended guidelines.
Contents 1 Features..........................4 2 Safety symbols used in this manual..................5 3 Safety guidelines........................5 4 Responsibility for operation or damage................6 5 Application..........................6 6 Before starting........................7 6.1 Transport........................7 6.2 Installation........................7 6.3 Storage........................7 6.4 Mounting instruction fastening set for wall mounting..........8 7 Start-up / Installation......................11 8 Operation.........................13 8.1 General operating instructions..................13 8.2 VACUUM MIXING with the Wamix................14...
Features Wamix is a manufacturer of vacuum mixers used not only for mixing all kinds of plasters, embedding media and silicones, but also designed for specialised dental procedures and for vacuum-rinsing of prostheses. When mixing silicones, make sure that the beaker is only filled to the max. height for silicone.
Safety symbols used in this manual Warning! This is a warning of risk situations and dangers. Failure to observe this warning could be life-threatening. These warnings has to be observed. Information! This symbol draws your attention to specific features that has to be observed. Safety guidelines Configuring and operating this equipment requires precise knowledge and observance of the instructions in this user manual.
The unit has to be maintained and operated in accordance with this user manual. Wassermann Dental-Maschinen GmbH is not responsible for damage arising from the nonobservance of these instructions.
Before starting Transport Before transporting the unit, ensure that it has been unplugged from the power socket. Make sure that it is packed correctly in order to avoid accidental damage. Be sure to check for any transport damage when unpacking the goods. Note down any damage if found.
Mounting instruction fastening set for wall mounting Please check the included accessories: Incl. in delivery item.-no. 174358 fastening set Quantity Name Securing bracket Phillips head screw M 4x8 Washer Ø 4,3 Serrated washer Ø 4,3 Slotted screw 4 x 16 Slotted screw 6 x 50 Screw anchor S8 Screw anchor S5...
Page 9
Secure the enclosed screw anchors as shown in Figure 1. Screw the 6x50 screws into the two upper holes; keep a distance of 6-7 mm to the wall! (Figure 2) fig. 1 fig. 2...
Page 10
Screw the securing bracket to the Wamix as shown in Figure 3. Do not yet fully tighten the Phillips head screw, however – you still need to be able to move the securing bracket. fig.3 Washer Serrated washer Phillips head srew Securing angle Screw 4x16 „control window“...
Start-up / Installation Check that the mains voltage and the voltage requirements of the mixer are the same, then plug in the power cable. Only for Wamix-Injector IV: Install a compressed air line in accordance with the technical data. Be sure to use the correct hose size if the line is long (inside diameter min.
Page 12
Start-up / Installation Switch on green master switch (1) (vacuum mixing). Only for Wamix-Injector IV: Open compressed air valve When the device is used for the first time, the vacuum first needs to be adjusted at the throttle valve (for the vacuum generator) using a screwdriver. The recommended input pressure is a minimum 5 bar.
Operation General operating instructions All instructions for using the unit, whether in verbal or written form, are based on our own experience and experimentation and can only be regarded as guidelines. Fill the beaker to the Max marking. The vacuum will draw any excess material into the system.
VACUUM MIXING with the Wamix By default, the timer is set to 60 seconds. 6.2.1 Wamix-Classic IV fig. 5 1. Set the mixing time on the timer (1) (up to 999 sec.) using the buttons (4 and 5). A: Automatic mixing = turning knob (7) you can make a free choice of 200 rpm until 600 rpm.
Page 15
4. Insert the beaker in the unit from below; the vacuum will be released automatically. Hold the beaker until the vacuum builds. 5. The mixer switches on automatically (depending on the settings – see point 1) for the set period of time. At the end of this period, an alert signal sounds for 5 seconds, after which the timer display (6) flashes until the main switch (2) is switched off.
Page 16
Place the mixing materials in the beaker and mix by hand until the powder is well mixed with the liquid. 3. Seal the beaker with a lid. 4. Insert the beaker in the unit from below; the vacuum will be released automatically. Hold the beaker until the vacuum builds.
Flow of airpressure too Check airpressure / adjust low ** Beaker falls out Faulty seal Replace seal (171708) Wamix-Classic IV * / Wamix-Injector IV ** If the above recommendations do not solve the problem, contact your dental depot or our service department.
Care and maintenance 10.1 Cleaning Disconnect the power plug before starting any maintenance work. The identification plate has always to be kept in easily legible condition and has not to be removed. Remove external dirt from time to time with some form of cold cleaner. Use only cold cleaners to avoid damaging the paintwork or removing the lettering.
10.2 Maintenance Allow the condensers to discharge before you open the mixer. Our vacuum mixers do not require servicing. Make sure the unit is kept clean and that the rim seal for the beaker (vacuum seal) is not damaged. 10.2.1 Changing the filter Fig.
Spare Parts If necessary please contact our service hotline phone. 10.5 Service Hotline 0049 (0)40 730 926 -20/ -24 10.6 Scope of delivery/ Accessories Wamix-Classic IV Item no.: 171979 Wamix-Injector IV Item no.: 171978* Incl. in delivery Item no.: Beaker, medium cpl. 350 ml...
10.7 Repairs Servicing or repairs to the unit has only to be carried out by qualified technicians. Only original spare parts are to be used. Responsibility for the product is voided if unauthorised persons alter it or if inappropriate components are installed. Disposing of the Unit The unit has to be disposed by an authorized recycling operation.
Technical data Wamix-Classic IV Item no.: 171979 Wamix-Injector IV Item no.: 171978* Voltage 220–240 V / 50/60 Hz Power consumption 2,4 A (Si 4,0 A), 1,8 A (Si 4,0 A)* Output 220 W, 120 W* W x H x D...
W A S S E R M A N N Product description: Vacuum Mixers Dental-Maschinen GmbH for dental Rudorffweg 15-17 application 21031 Hamburg Germany Model: Wamix-Classic IV Item no.: 171979 Wamix-Injector IV Item no.: 171978 Applicable DIN EN 61010-1 standards: DIN EN 61000-6-3 DIN EN 61000-6-1 DIN 45635-1 DIN EN 60335-1 Hiermit wird bestätigt, dass die oben bezeichnete Maschine den genannten EG-Richtlinien...