Page 1
AeroBioCollector User's Manual LCB 71260 La Salle France bioMérieux SA Chemin de l'Orme Distributed by : 69280 Marcy-l'Étoile - France RCS Lyon 673 620 399 Tel. 33 (0)4 78 87 20 00 Fax 33 (0)4 78 87 20 90 www.biomerieux.com...
Page 3
R e v i s i o n s The list of revisions below summarizes replacements or additional pages in your User’s Manual Date of Version Modifications Pages modified printing 2001/01 Bilingual (FR / EN) 2001/04 Multilingual (FR / EN / ES / IT / DE / PT) 2005/11 Update of the manual (FR / EN / ES)
Page 5
To ensure user safety, the instrument must be used in accordance with this manual. T r a d e m a r k s BIOMERIEUX, the blue logo, 3P, Air IDEAL and Count-Tact are used, pending and/or registered trademarks belonging to bioMérieux, or one of its subsidiaries, or one of its companies.
T a b l e o f c o n t e n t s How to use this manual ........................1-1 Finding topics and procedures ............................1-1 Glossary ..................................1-2 Typographic conventions ..............................1-3 ® Air IDEAL 3P™ Traceability keypad ..........................1-3 Functional description .......................... 2-1 Presentation .................................2-1 Principle of use .................................2-1 Operating principle ..............................2-2...
Page 8
Table of contents ® Air IDEAL 3P™ Traceability menus ........................3-16 ® Turning on Air IDEAL 3P™ Traceability ........................ 3-17 Standby .................................. 3-18 ® Turning off Air IDEAL 3P™ Traceability ........................ 3-18 Selection of one of the 4 pre-programmed sample volumes (Menu 1) ..............3-18 Programming a sample volume not in memory (Menu 2) ..................
Page 9
L i s t o f f i g u r e s Note: Screenshots and figures are given as examples only. 2. Functional description Figure 2-1: Principle of the impaction biocollector ......................2-2 3. Using Air IDEAL®3P™ Traceability ® Figure 3-1: Air IDEAL 3P™...
H o w t o u s e t h i s m a n u a l IMPORTANT! Please read the "General safety and regulatory information" booklet provided with the instrument. F i n d i n g t o p i c s a n d p r o c e d u r e s This manual is divided into 7 chapters.
How to use this manual Glossary G l o s s a r y Acrylonitrile butadiene styrene. ® Air IDEAL 3P™ Traceability, instrument used to collect and count viable bacterial and AeroBioCollector or fungal particles in a known and precise volume of air. Air sampler Number of Colony Forming Units read on the agar plate.
How to use this manual Typographic conventions T y p o g r a p h i c c o n v e n t i o n s These conventions are used in the different chapters of the manual. •...
F u n c t i o n a l d e s c r i p t i o n P r e s e n t a t i o n "Pharmaceutical manufacturing evolves from an art to a science". This sentence alone from the FDA Guideline “Pharmaceutical cGMPs for the 21 century –...
Functional description Presentation O p e r a t i n g p r i n c i p l e ® Air IDEAL 3P™ Traceability is an impaction AeroBioCollector used to detect the presence of viable micro-organisms in the environment to be tested, by precise sampling of a given volume of air.
Functional description Presentation C o l l e c t i o n e f f i c i e n c y v a l i d a t e d a c c o r d i n g t o t h e I S O 1 4 6 9 8 s t a n d a r d ®...
Using Air IDEAL®3P™ Traceability Description C o n s u m a b l e s ® Air IDEAL 3P™ Traceability is adapted to all types of Petri dishes available on the market: 65, 70 and 90 mm. Moreover, the use of a large range of ready-to-use culture media – irradiated or classic – developed and manufactured by bioMérieux, the reference in this field, enables you to obtain the best possible performance from the instrument.
Using Air IDEAL®3P™ Traceability Sampling grid K e yp a d Figure 3-5: Keypad S a m p l i n g g r i d C o n f i g u r a t i o n f o r 6 5 , 7 0 o r 9 0 m m P e t r i d i s h e s Sampling grid for 65 or 70 mm Petri dishes Sampling grid for 90 mm Petri dishes ®...
Using Air IDEAL®3P™ Traceability Sampling grid Stainless steel strip for attachment Aspiration orifice Seal Figure 3-7: Sampling head– 65, 70 or 90 mm Petri dishes T r i p o d a s s e m b l y Figure 3-8: Tripod assembly ®...
Using Air IDEAL®3P™ Traceability General specifications G e n e r a l s p e c i f i c a t i o n s E n vi r o n m e n t a l c o n d i t i o n s Normal storage conditions ................
Using Air IDEAL®3P™ Traceability General specifications T e c h n i c a l c h a r a c t e r i s t i c s K e yp a d Key control. D i s p l a y Liquid crystal display (2 x 16 characters).
Using Air IDEAL®3P™ Traceability General specifications E l e c t r i c a l c h a r a c t e r i s t i c s Voltage......................7.4 VDC nominal Power supply ......................1 A max. ®...
Using Air IDEAL®3P™ Traceability Unpacking Air IDEAL®3P™ Traceability ® U n p a c k i n g A i r I D E A L 3 P ™ T r a c e a b i l i t y The instrument is supplied in a carrying case and box.
Page 28
Using Air IDEAL®3P™ Traceability Recommendations for installation and use WARNING! The user’s manual or instruction manual for an intentional or unintentional radiator shall caution the user that changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.
Using Air IDEAL®3P™ Traceability Sampling positions S a m p l i n g p o s i t i o n s Figure 3-9: Sampling positions ® IMPORTANT! Air IDEAL 3P™ Traceability is delivered with a hook for suspending it in a glove box during decontamination phases.
Using Air IDEAL®3P™ Traceability Screwing on the sampling grid S c r e w i n g o n t h e s a m p l i n g g r i d CAUTION! Do not insert any foreign objects into the aspiration orifice located under the grid (see page 3-5).
Using Air IDEAL®3P™ Traceability Electrical power supply L o w b a t t e r y s i g n a l s ∗ When the battery life has expired, the following low battery signals are output: Visual signal on the welcome message display: Flashing symbol Visual signal on the sample in progress display: Flashing symbol...
Using Air IDEAL®3P™ Traceability Electrical power supply ∗ At the end of the operation, Disconnect the charger from the power outlet. • Disconnect the charger from the the instrument jack connector. • Install the protection cap. • Check that the battery life displayed is again 4 hours. •...
Using Air IDEAL®3P™ Traceability Operation in slave mode O p e r a t i o n i n s l a v e m o d e ® − In slave mode, the Air IDEAL 3P™ Traceability is used with a RUID. −...
Using Air IDEAL®3P™ Traceability Operation in manual mode ® A i r I D E A L 3 P ™ T r a c e a b i l i t y m e n u s Turning on. WELCOME Last sample taken.
Page 35
Using Air IDEAL®3P™ Traceability Operation in manual mode ® T u r n i n g o n A i r I D E AL 3 P ™ T r a c e a b i l i t y ∗...
Page 36
Using Air IDEAL®3P™ Traceability Operation in manual mode S t a n d b y ∗ To put the instrument on standby, • <ON/OFF> Briefly press , the indicator light must not go off. • Wait for 5 minutes. ∗ To reactivate the instrument, •...
Page 37
Using Air IDEAL®3P™ Traceability Operation in manual mode Example: ∗ To select stored volume No. 2 from menu 1, • < + > Press twice. Wait 2 seconds. • Note: If you did not wait for the end of the first timeout, in other words if you pressed another button before the 2 seconds, the new value selected ( ) is not recorded.
Page 38
Using Air IDEAL®3P™ Traceability Operation in manual mode ∗ To program a new volume, < > or < > Press until the desired volume is displayed. • < > or < > program a new IMPORTANT! It is imperative to press the buttons immediately to volume.
Page 39
Using Air IDEAL®3P™ Traceability Operation in manual mode ∗ To program a new volume for program 1, < > or < > Press until the desired volume is displayed. • < > or < > IMPORTANT! It is imperative to press the buttons immediately to program a new volume.
Page 40
Using Air IDEAL®3P™ Traceability Operation in manual mode ∗ To program a new volume for program 4, < > or < > Press until the desired volume is displayed. • < > or < > IMPORTANT! It is imperative to press the buttons immediately to program a new volume.
Page 41
Using Air IDEAL®3P™ Traceability Operation in manual mode ∗ To program a delay value, < > or < > Press until the desired delay value is displayed. • < > or < > IMPORTANT! It is imperative to immediately press to program a new volume.
Page 42
Using Air IDEAL®3P™ Traceability Operation in manual mode ∗ Total sampling time must be less than 5 hours otherwise an error message will appear. ∗ To select menu 5, • <MENU> Press 5 times. The following message appears and the 1 line flashes: The unit volume is shown on the lower left and cannot be modified in this menu.
Page 43
Using Air IDEAL®3P™ Traceability Operation in manual mode B a t t e r y a u t o n o m y t e s t ( M e n u 6 ) This menu is used to verify that the battery is still compliant with specifications of new material (battery life ≥...
Using Air IDEAL®3P™ Traceability Starting sampling S t a r t i n g s a m p l i n g ® CAUTION! Before starting, check that Air IDEAL 3P™ Traceability is fitted with a sampling grid and not its protective cover otherwise it could undergo irreversible damage which is not covered by the bioMérieux warranty.
Page 45
Using Air IDEAL®3P™ Traceability Starting sampling − A buzzer indicates the end of sampling (6 short buzzes) and the display returns to the original message with automatic correction of the remaining battery life. The indicator light changes to constant green. The instrument is free to perform a new sampling cycle.
Using Air IDEAL®3P™ Traceability Starting sampling S t o p p i n g t h e m o t o r • <START/STOP> Press <START/STOP> Note: It is always possible to stop a sampling operation in progress by pressing If the motor is stopped during the program, i.e.
Using Air IDEAL®3P™ Traceability Setting Air IDEAL®3P™ Traceability parameters (summary diagram) ® S e t t i n g A i r I D E A L 3 P ™ T r a c e a b i l i t y p a r a m e t e r s ( s u m m a r y d i a g r a m ) Menu 1: Selection of one of Menu 2: Selection of a...
Page 48
Using Air IDEAL®3P™ Traceability Setting Air IDEAL®3P™ Traceability parameters (summary diagram) Menu 4 : Programming Menu 5 : Programming Menu 6 : Battery autonomy delayed start-up. sequenced samplings. test. Press < > or < > until the Press < > or < >...
Using Air IDEAL®3P™ Traceability Maintenance Menu M a i n t e n a n c e M e n u N a vi g a t i o n u s i n g < + > a n d < - > k e ys STORED VOL: ZZZL NEXT CALIBRATION FSE (Field Service...
Page 50
Using Air IDEAL®3P™ Traceability Maintenance Menu Welcome message Date of next calibration (Month/Day/Year) ID of operator who performed calibration Serial number of RUID used for calibration Flow Rate Measurement Bench serial number Date of last calibration ® Air IDEAL 3P™ Traceability firmware version ®...
P r o c e d u r e P r i n c i p l e Implementation and following of a procedure to measure aerobiocontamination corresponds to a process based on prevention, which involves: Evaluating the current level and standard of hygiene and controls. •...
Procedure How to obtain a good quality sample H o w t o o b t a i n a g o o d q u a l i t y s a m p l e P r e c a u t i o n s o f u s e Check the condition of the instrument and sampling grid.
Procedure Sampling plan S a m p l i n g p l a n The sampling plan must be drawn up very carefully and followed strictly. The aim is to guarantee that the values obtained are comparable. Any discrepancy between values should reveal a variation in aerobiocontamination.
M a i n t e n a n c e P r e v e n t i v e m a i n t e n a n c e CAUTION! No preventive maintenance needs to be performed by the user. If an anomaly occurs on your instrument, call bioMérieux or your local distributor.
Maintenance Routine servicing D e c o n t a m i n a t i o n o f e x t e r n a l p a r t s All the external parts of the instrument can be decontaminated with most of the usual disinfectants at the usual effective concentration (70% isopropanol, hydrogen peroxide, 70% ethanol, quaternary ammonium).
Tr o u b l e s h o o t i n g Indicator light Action Action Error message Possible cause status in MANUAL mode in SLAVE mode Press <MENU> to return to Press <MENU> to return to LAST SAMPLING Sampling is Flashing red interrupted in...
Page 58
Troubleshooting Routine servicing Indicator light Action Action Error message Possible cause status in MANUAL mode in SLAVE mode MOTOR JAMMED Sampling has been Inspect the black grid under Inspect the black grid under interrupted as the the Petri dish retaining claws. the Petri dish retaining claws.
A p p e n d i x U s i n g t h e r e a d i n g t a b l e The following reading tables indicate the most probable number of micro-organisms collected per plate (MPN collected) with respect to the number of agglomerates of colonies counted on the agar (CFU read).
Appendix Reading tables The probability pi decreases as i increases. It is therefore particularly relevant to apply the statistical correction using FELLER’s law when high CFU values are read, i.e. close to N, i.e. in the case of agar plates almost completely saturated with colonies.