2118 CONGRATULATIONS ON YOUR NEW CALAJECT ™ Please read these instructions thoroughly before you start using your CALAJECT™ INTENDED USE CALAJECT™ is intended for performing dental local analgesia. CALAJECT™ MAY ONLY BE USED BY TRAINED PERSONNEL authorized to perform dental injections. For this reason, this manual does not include specific in- structions in injection techniques.
Page 3
The cartridge barrel should also be replaced if it is damaged or if the thread appears to be worn. Replacement barrels (3 pcs. per pack) can be ordered through your CALAJECT™ dealer. PLACING CALAJECT™ should be placed on a stable base during use. Ensure good access to all sockets and plugs during use and charging. CHARGING CALAJECT™...
Unplug by pulling the grooved sliding ring backwards (do not turn). Do not touch patient and connectors of control unit / battery charger at the same time. • Turn on CALAJECT™ by touching the on/off key at the front panel. By this, the piston will automatically return to starting position.
Page 5
• Select program. When the foot switch is activated again, the chosen injection program will be active. • CALAJECT™ will stop automatically, when the cartridge is empty - a long beep will be heard. Return the piston rod by pressing “R” at the display. Make sure to let the piston retract completely, before you unscrew the barrel from the handpiece.
PDL technique, we refer to the published literature on the subject. If the pressure has become so high that CALAJECT™ stops, the needle opening may have be- come blocked and it is recommended to rotate the needle slightly in order to obtain a good flow.
Page 7
Recommended for regional nerve block analgesia • Activate the footswitch and hold it down throughout the injection. CALAJECT™ begins slowly, 0.006 ml/sec. By releasing/re-activating the foot control in one swift movement, the injection speed will increase gradually over the next 5 seconds to high speed (approx.
Read instructions for wires, plugs, terminals or possible. Check the correct connection / foot switch. function by using removal of plug. Avoid another CALAJECT squeezing damages foot switch. Send in for and sharp bending of repair. cables. Premature illumination Blocked needle Change the needle of pressure indicator.
Page 9
CONTRA-INDICATIONS AND PHYSIOLOGICAL HAZARDS Contra-indications: No specific for CALAJECT™ beyond the contra-indications described in the product page and package insert of the applied analgesic solution purchased separately. Possible physiological hazards: In the event of the most serious technical failure that can be foreseen, the dental cartridge will be emptied in 30 seconds at maximum speed.
Page 10
EMC Technical Information Table 1 Electromagnetic emissions The “CALAJECT™” is intended for use in the electromagnetic environment specified below. The customer or the user of the “CALAJECT™” should assure that it is used in such an environment. Emissions test Compliance...
Page 11
Table 3 Electromagnetic immunity The “CALAJECT™” is intended for use in the electromagnetic environment specified below. The customer or the user of the “CALAJECT™” should assure that it is used in such an environment. Immunity test IEC 60601-1 Test level...
Page 12
Explanation of symbols used in CALAJECT™ labelling. CE mark with ID no. of notified body. 0543 Type B applied part. Interference may occur in the vicinity of equipment marked with this symbol. To be disposed of as electronic waste according to WEEE Directive 2012/19/EU.