NoiseMeters CEL24x Meter Controls 1. Microphone 2. Power On/Off Key 3. Display 4. Left Soft Key 5. Right Soft Key 6. USB Port 7. Auxiliary socket (2.5mm stereo)
NoiseMeters CEL24x Power Observe polarity when fitting batteries Do not mix battery types Fitting Batteries 1. Open the Battery Cover 2. Check the compartment is clean and dry 3. Fit three new AA Alkaline or NiMH 4. Close the battery cover...
NoiseMeters CEL24x Display A. Battery Indicator B. Over Range Indicator C. Sound Level D. Analog Display Bar E. Measurement Scale F. Maximum Sound Level G. Left Key - Instrument Settings H. Right Key - Reset Measurement Average Level (CEL244 and CEL246)
NoiseMeters CEL24x Settings The settings can be accessed from the Measurement Screen. Press the settings key (Left Key) repeatedly to cycle through the settings screens. Press the Right key to make changes to the settings shown. Recording (CEL242 and CEL246 Only) ●...
Page 9
NoiseMeters CEL24x Settings Frequency Weightings - Select either “A” or “C” weighting. Check your regulations, but most measurements are “A” weighted. Average Setting (CEL242 and CEL246) Either Leq (for Q=3) or Lavg (Q=5) depending on regulations. Logging Interval (CEL246) - Setting the logging interval for between 1 and 10s average value.
NoiseMeters CEL24x Calibration Calibration should be performed using a CEL120/2 or similar Calibrator before making any important measurements. 1. Select 60-130dB range. 2. Switch Calibrator on to 114dB setting. 3. Fit Calibrator over the microphone. The meter automatically detects the Calibrator tone and enters Calibration Mode.
NoiseMeters CEL24x Specifications Standards IEC61672-1 Class 2 ANSI S1.4 Type 2 CEL244/246: IEC60804 Type 2 and ANSI S1.43 Type 2 Range 30-100dB and 60-130dB Freq Wtg “A” and “C” weightings Noise Floor Typically <33 dB(A) Time Wtg Fast, Slow and Impulse...