Invasix BodyTite™ Service Manual DO600463F
0344
Page 1 of 52

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the BodyTite and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Invasix BodyTite

  • Page 1 BodyTite-TiteFx Service Manual - DO600463F Invasix BodyTite™ Service Manual DO600463F 0344 Page 1 of 52...
  • Page 2: Table Of Contents

    LASSIFICATION SECTION 3 - SYSTEM INSTALLATION LECTRICAL EQUIREMENTS NVIRONMENTAL EQUIREMENTS QUIPMENT NSTALLATION System parts View: OVING THE YSTEM ISPOSAL OF YSTEM SECTION 4 - DESCRIPTION OF DEVICE PERATOR ONTROL ANEL Invasix BodyTite™ Service Manual DO600463F 0344 Page 2 of 52...
  • Page 3 Power Supply Disassembling Power Supply Assembling RF Card Disassembling RF Card Assembling Fan Assembly Disassembling Fan Assembly Assembling Controller Card Disassembling Top Assy Disassembling Top Assy Assembling Speaker Disassembling Speaker Assembling Invasix BodyTite™ Service Manual DO600463F 0344 Page 3 of 52...
  • Page 4 Error! Bookmark not defined. Top Assy Disassembling Error! Bookmark not defined. RF Card Assembling Error! Bookmark not defined. Controller Card Assembling Error! Bookmark not defined. Connector Bracket Assembling Error! Bookmark not defined. Invasix BodyTite™ Service Manual DO600463F 0344 Page 4 of 52...
  • Page 5 Adding of RF calibration and Testing 14/10/2009 11/07/2011 Update to cover the upgrade to TiteFX hand piece Update according to DEKRA reviewer 16/10/2011 question TDR04/Q11 in report 3900465- TDR04-R0 30,31 Updated Error codes 23/01/2012 Invasix BodyTite™ Service Manual DO600463F 0344 Page 5 of 52...
  • Page 6: Section 1 - Introduction

    The System must be serviced only by Invasix Ltd. qualified personnel. System Overview The BodyTite platform is intended multiple applications and support different hand pieces employing RF energy. The System provides individual adjustment of RF energy to achieve maximum efficiency and safety for each patient.
  • Page 7: Conventions Used In The Manual

    Do not discard in trash. Electronic equipment should be disposed of in an appropriate manner. Fuse Type BF applied part HF Isolated Patient Circuit This equipment intentionally supplies non-ionizing RF energy for physiologic effect Invasix BodyTite™ Service Manual DO600463F 0344 Page 7 of 52...
  • Page 8: Section 2 - Safety

    Personnel should not operate the System until they have been fully educated in its use. Make sure that all treatment personnel are familiar with the System controls and know how to shut down the System instantly. Invasix BodyTite™ Service Manual DO600463F 0344 Page 8 of 52...
  • Page 9: Cautions

    RF generator may cause interference between them. Move the System slowly and carefully. The System weighs approximately 15kg (33lb.) and may cause injury if proper care is not used when moving it. Invasix BodyTite™ Service Manual DO600463F 0344 Page 9 of 52...
  • Page 10: Fire Hazards

     There is brief delay between indication of energy activation and actual emission of RF energy.  System is starting at lower power setting. Invasix BodyTite™ Service Manual DO600463F 0344 Page 10 of 52...
  • Page 11: Active Accessories

     Do not connect wet accessories to the System.  Hand piece are single use only. Don’t try to sterilize and reuse the hand piece. Labels Invasix BodyTite™ Service Manual DO600463F 0344 Page 11 of 52...
  • Page 12: Equipment Classification

    The following is a list of the different equipment used and their classifications.  Electric shock protection: Class I, Type BF  Protection against ingress of liquids: Ordinary equipment  Not suitable for use in presence of flammable substance Invasix BodyTite™ Service Manual DO600463F 0344 Page 12 of 52...
  • Page 13 System is classified as a IIb device defined by the Medical Device Directive (93/42/EEC) for CE marking. Invasix BodyTite™ Service Manual DO600463F 0344 Page 13 of 52...
  • Page 14: Section 3 - System Installation

    (68º-79ºF) and relative humidity of less than 80%. Equipment List The System includes the following:  System platform  TiteFx hand piece  Foot Switch  Operator Manual  Power cord Invasix BodyTite™ Service Manual DO600463F 0344 Page 14 of 52...
  • Page 15: Installation

    Installation System parts View: Main Display Operating Keys On/Off switch TiteFx Connector Figure 3: Front view Invasix BodyTite™ Service Manual DO600463F 0344 Page 15 of 52...
  • Page 16 Software Plug Foot Switch connection Power Inlet Figure 4: Back View Invasix BodyTite™ Service Manual DO600463F 0344 Page 16 of 52...
  • Page 17: Moving The System

    To comply with European Commission Directive 2002/96/EC on Waste Electrical and Electronic Equipment (WEEE) and other country and state regulations, please DO NOT dispose of this equipment in any location other than designated locations. Invasix BodyTite™ Service Manual DO600463F 0344 Page 17 of 52...
  • Page 18: Section 4 - Description Of Device

    Panel consists of On-Off switch, four function keys, an information screen, and hand piece connector (Figure 3.2). Main Display Operating Buttons BodyTite and Fractora connector On/Off switch TiteFx Connector Power On-Off switch Power switch turns system on and off. Invasix BodyTite™ Service Manual DO600463F 0344 Page 18 of 52...
  • Page 19 LCD screen above the each key. LCD screen LCD screen shows information about machine status, treatment parameter and titles of functional keys. LCD treatment Scroll Key Increase Key Decrease Key OK Key Invasix BodyTite™ Service Manual DO600463F 0344 Page 19 of 52...
  • Page 20: Treatment Screens

    The Self-Test screen appears after the Power switch is turned on. Software is loaded from the plug and self-test of the system modules is performed. After the end of the self-test, the Treatment screen appears. Invasix BodyTite™ Service Manual DO600463F 0344 Page 20 of 52...
  • Page 21 This indicator shows the total energy delivered from the beginning of Total Energy the treatment. Volume This function allows the user to adjust system volume. Selection Frame The frame selects parameters that can be changed by functional keys. Invasix BodyTite™ Service Manual DO600463F 0344 Page 21 of 52...
  • Page 22 Standby mode. When the signal from the footswitch is indicated, the system switches to Active mode. Any attempt to change the treatment settings switches the system to Standby mode. Invasix BodyTite™ Service Manual DO600463F 0344 Page 22 of 52...
  • Page 23: Functional Keys

    Pressing 2 central functional keys simultaneously will present measured information on the top of treatment screen. The presented information includes RF power, voltage, current, and impedance, 2 RF voltage before and after the sequence. Invasix BodyTite™ Service Manual DO600463F 0344 Page 23 of 52...
  • Page 24: Hand Piece

    - Applied to the skin surface and shape the skin for optimal heating RF electrodes - Located in the vacuum chamber and delivers RF energy to the treated tissue - Shows measured temperature Screen Invasix BodyTite™ Service Manual DO600463F 0344 Page 24 of 52...
  • Page 25 - It is connected to the control panel of the system. The connector has individual IDs to recognize each handpiece and its usage history Hand pieces in cradle are shown in picture below Invasix BodyTite™ Service Manual DO600463F 0344 Page 25 of 52...
  • Page 26: System Specifications

    Maximum Output Power Output Frequency 1[MHz] ± 2% 600[V] Crest Factor * (Rated Load) 2 ± 2% RF V 2000V 5μsec RF Maximal sequence duration Output Frequency 1.7[MHz] ± 2% Invasix BodyTite™ Service Manual DO600463F 0344 Page 26 of 52...
  • Page 27 Invasix BodyTite™ Service Manual DO600463F 0344 Page 27 of 52...
  • Page 28: Section 5 - System Operation

    If no testing load is connected a “Bad Coupling” message will appear on the upper right side of the screen System Shutdown To shutdown the system turn power switch off. Invasix BodyTite™ Service Manual DO600463F 0344 Page 28 of 52...
  • Page 29: Section 6 - Maintenance

    This chapter describes the routine maintenance: Cleaning device exterior  Clean exterior of the BodyTite system with soft cloth.  Disinfect the exterior parts of the device with hospital grade disinfector if it is required. (Do not use alcohol or spill any liquids on any part of the system, it is not sealed to liquids).
  • Page 30: Section 7 - Error Messages

     Software Start Error: The software plug is present but for any reason the OS returns an error while launching the application software. Reason: Operating system fault (rare, never faced) Invasix BodyTite™ Service Manual DO600463F 0344 Page 30 of 52...
  • Page 31  H004 Used hand piece indicated: Software detected a used Hand Piece Reason: The BodyTite HP has made more than 150KJ or FaceTite HP has made more than 50KJ.  H005 System Memory Fault: The software failed to read system parameters Reason: The system parameters (RF calibration) are corrupted.
  • Page 32: Section 8 - Troubleshooting

    Replace Hand Piece Memory Fault Temperature Replace Hand Piece sensor fault Check wiring from RF card to Hand Piece connector (only for Replace RF card BodyTite) System memory Recalibrate RF card Invasix BodyTite™ Service Manual DO600463F 0344 Page 32 of 52...
  • Page 33 Check controller to footswitch wiring Replace Controller card Hand piece LCD Replace hand piece is blank 2nd RF is not Replace hand piece produced No Vacuum Replace hand piece Invasix BodyTite™ Service Manual DO600463F 0344 Page 33 of 52...
  • Page 34: Section 9 - Assembly Instructions

     Align the fingers to the openings in the cabinet, and insert the fingers into the openings  Push all the way up the cover and tighten it with two screws to the cabinet Invasix BodyTite™ Service Manual DO600463F 0344 Page 34 of 52...
  • Page 35 2 x Screws Figure 5: Front cover removal Fingers GND Connection Figure 6: Front cover assembly Invasix BodyTite™ Service Manual DO600463F 0344 Page 35 of 52...
  • Page 36: Ac Card Disassembling

     Insert the two connectors into the AC card (From power inlet, From On/off switch) From Power entry Connectors AC Filter card 4 x Screws Connector to On/Off Switch Figure 7: AC Card removal & Assembly Invasix BodyTite™ Service Manual DO600463F 0344 Page 36 of 52...
  • Page 37: Power Supply Disassembling

     Insert the GND quick disconnection to the GND terminal on the power supply located to the right of the AC power connector 20 pin Molex connector 4 x Screws Power Supply AC power connector GND Quick disconnection Invasix BodyTite™ Service Manual DO600463F 0344 Page 37 of 52...
  • Page 38: Rf Card Disassembling

     Align the RF box cover to the RF shielding box and attach it using the 11 screws After replacing the RF card the RF must be calibrated see section - 9 Invasix BodyTite™ Service Manual DO600463F 0344 Page 38 of 52...
  • Page 39 Figure 9: RF card box cover RF Card Ring Terminal RF Card – RF out Connector RF Card – Control Signals RF Card Connector RF Card – DC power Connector Figure 10: RF card removal Invasix BodyTite™ Service Manual DO600463F 0344 Page 39 of 52...
  • Page 40: Fan Assembly Disassembling

     Tighten the fan to the honey comb using the four nuts  Connect the DC power connection wiring harness to the Controller card, the RF card and the DC power supply Invasix BodyTite™ Service Manual DO600463F 0344 Page 40 of 52...
  • Page 41 4 x Nuts and washers DC Power connection & Fan Assy wiring DC power out to RF, Controller & Fan Figure 11: DC power & Fan Assy wiring harness Invasix BodyTite™ Service Manual DO600463F 0344 Page 41 of 52...
  • Page 42: Controller Card Disassembling

     Connect the connectors to the card  Connect back the software plug at the back side of the system After replacing the Controller card the RF must be calibrated see section - 9 Invasix BodyTite™ Service Manual DO600463F 0344 Page 42 of 52...
  • Page 43 Hand piece wiring opening TiteFX Comm and power Controller to Key Pad Controller to LCD harness harness EH600567x Controller to H.P. I (not in use) Figure 12: Controller card Assembly Invasix BodyTite™ Service Manual DO600463F 0344 Page 43 of 52...
  • Page 44: Top Assy Disassembling

     Connect the keypad flat cable to the controller card  Rout the LCD flat cable through the ferrite and connect it to the controller card  Attach the top Assy to the top panel using the four screws Invasix BodyTite™ Service Manual DO600463F 0344 Page 44 of 52...
  • Page 45 Figure 13: Top Assy LCD Adaptor Connector Bracket & 2 screws Connector Bracket & 2 screws LCD Ferrite & Cable Key Pad Cable connection Connection Figure 14: Top Assy Connector bracket Invasix BodyTite™ Service Manual DO600463F 0344 Page 45 of 52...
  • Page 46: Speaker Disassembling

     Align the speaker to its standoffs  Attach the speaker to the cabinet using four screws  Connect the speaker connector to the controller card Speaker Connection Controller Speaker Figure 15: Speaker Invasix BodyTite™ Service Manual DO600463F 0344 Page 46 of 52...
  • Page 47 Speaker Screws x 4 Tie wraps Figure 16: Speaker Invasix BodyTite™ Service Manual DO600463F 0344 Page 47 of 52...
  • Page 48 I1 (rms)  Adjust the reading on the display as close as possible to the calculated value, using the up or down keys.  Release the footswitch Invasix BodyTite™ Service Manual DO600463F 0344 Page 48 of 52...
  • Page 49  To start RF calibration press shortly on the footswitch and release The RF calibration is automatic procedure and when done it will display a text box.  Press the down key to save RF settings into system Invasix BodyTite™ Service Manual DO600463F 0344 Page 49 of 52...
  • Page 50 +/- 10% of the “Setup Power”.  Press the OK key to finish the test, a “TEST COMPLETE” screen appears. Write down all parameters showing on the screen. Invasix BodyTite™ Service Manual DO600463F 0344 Page 50 of 52...
  • Page 51 Section 11 – System Block Diagram Figure 17: BodyTite Block diagram Invasix BodyTite™ Service Manual DO600463F 0344 Page 51 of 52...
  • Page 52 Section 12 – System Wiring Figure 18: System wiring diagram Invasix BodyTite™ Service Manual DO600463F 0344 Page 52 of 52...