PrimoX microscopes, binocular and trinocular
Vareno. 111200_10
Manual for PrimoX
microscopes
11.12.00 PrimoX, binocular
11.12.10 PrimoX, trinocular

Specifications:

Total
Magnification: 40x - 1000x.
Eyepiece:
10x wide field

Objectives:

4x N.A. 0.10
plan
10x N.A. 0.25 plan
20x N.A. 0.40 plan
40x N.A. 0.65 plan
100x N.A. 1.25 plan
Gundlach A/S
Silkeborgvej 765 · DK-8220 Brabrand
Mechanical
stage:
Size:
140 mm x 210 mm.
X-Y range:
75 mm x 50 mm.
Graduation:
1 mm
Focusing:
Graduation for fine
focus is 0.002 mm.
Abbé-
condenser:
N.A. = 1.25.With
built-in diaphragm
and filter holder.
Extra swing-out lens
that is used with the
4x objective.
Tel. +45 8694 1388
gundlach@gundlach.eu
Fax +45 8694 2486
www.gundlach.eu
VE111200_10
Rev. 19.06.2007

Construction of the microscope:

Objectives:
All objectives are produced accor-
ding to DIN-standard. 40x and 100x
objectives have a spring mount to
avoid breakage of specimens and
objectives. The five-hole objective
revolver makes it easy to change
between the different objectives.
Focusing knobs and stage stop
adjustment:
The coaxial coarse and fine adjust-
ment knobs make it easy to attain a
sharp image. The biggest knobs are
used for the coarse adjustment and
the smallest knobs for the fine adju-
stment.
Between the left large focusing
knob and the column is there also a
ring. With this ring it is possible to
adjust the friction when turning it. In
this way it can be adapted to diffe-
rent users' wishes.
On the right side (between the right
large focusing knobs and the colu-
mn) is a stop ring with a small peg.
This is called the stage stop. With
this ring it is possible to settle the
upper limit of the stage movement
to protect the specimen from
touching and/or damaging the ob -
jectives. Furthermore, the stop ring
will make you able to focus more
quickly. The stage stop ring is loo-
sened by turning the ring towards
you/down, when you are sitting in
front of the microscope.

Mechanical stage:

The specimen is placed on the
mechanical stage by pulling back
the movable handle of the speci-
men holder. Gently release the
handle and the specimen is placed
securely. The mechanical stage can
then be moved in the X and Y-axis
with the coaxial mounted knobs
under the right side of the mechani-
cal stage.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PrimoX Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Gundlach PrimoX Series

  • Page 1 X and Y-axis with the coaxial mounted knobs 40x N.A. 0.65 plan under the right side of the mechani- 100x N.A. 1.25 plan cal stage. Gundlach A/S Tel. +45 8694 1388 gundlach@gundlach.eu Silkeborgvej 765 · DK-8220 Brabrand Fax +45 8694 2486 www.gundlach.eu...
  • Page 2 The dioptre ring is the adequate. the clamp screw. 2 / 5 SIDE Gundlach A/S Tel. +45 8694 1388 gundlach@gundlach.eu Silkeborgvej 765 · DK-8220 Brabrand Fax +45 8694 2486 www.gundlach.eu...
  • Page 3 Accessories (not included) has contact with the immersion big clamp screw. Gundlach PrimoX microscopes are, oil. – Close the lamp housing. in the standard version, well suited – Air bubbles in the immersion oil for most routine jobs.
  • Page 4 The adapter is the very important difference that equipped with a relay optic that immersion oil must be used be - 4 / 5 SIDE Gundlach A/S Tel. +45 8694 1388 gundlach@gundlach.eu Silkeborgvej 765 · DK-8220 Brabrand Fax +45 8694 2486 www.gundlach.eu...
  • Page 5 The use of the phase equipment is be simply unscrewing the two sil- of the (black) condenser holder. It is described in its own manual. 5 / 5 SIDE Gundlach A/S Tel. +45 8694 1388 gundlach@gundlach.eu Silkeborgvej 765 · DK-8220 Brabrand Fax +45 8694 2486 www.gundlach.eu...