APO-go PUMP
PROGRAMMING GUIDE
PUMP
apomorphine hydrochloride

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the APO-go and is the answer not in the manual?

Questions and answers

Carmen Manrique Castillo
April 25, 2025

Me sale en pantalla Er.1 y no puedo quitarlo. La máquina se ha bloqueado y pita.

1 comments:
Mr. Anderson
May 13, 2025

Error code Er.1 on the Crono APO-go machine means an incorrect initial setting, possibly due to the battery being inserted wrongly. To resolve it, remove the battery and reinsert it after 15 seconds.

This answer is automatically generated

Summary of Contents for Crono APO-go

  • Page 1 APO-go PUMP PROGRAMMING GUIDE PUMP apomorphine hydrochloride...
  • Page 2: Table Of Contents

    The APO-go MARK III PUMP has two available programming modes: APo1 and APo3. The APo1 mode is used for a single flow rate, whereas APo3 allows the programming of four different flow rates, for variable flow rate. The different functionalities and a brief explanation of these is included below.
  • Page 3: Locking/Unlocking The Apo-Go Pump

    Locking/unlocking the APO-go PUMP Setting the bolus/priming dose and syringe volume Locking/unlocking APo1 MODE APo3 MODE This task requires the pressing of two buttons simultaneously Display screen L1 = locked. Padlock symbol on display Pump set to and unlocked (L0) L0 = unlocked.
  • Page 4: Switching The Apo-Go Pump On

    6.40 1.60 12.30 6.15 1.70 11.46 5.53 1.80 11.07 5.34 1.90 10.31 5.16 10.0 2.00 10.00 5.00 Settings between 0.05ml/hr to 5.00ml/hr are available. Values shown here are 0.4ml/hr to 2.0ml/hr. Solution used = 5mg/ml strength (i.e. APO-go Pre-filled syringe).
  • Page 5: Priming The Line In Apo1 Mode

    Step 5: Setting the flow rate Priming the line in APo1 mode The screen will show F1 with a number next to it. This is your first flow rate. 0.60 APo1 MODE Referring to your flow rate record set the flow rate for F1 using buttons.
  • Page 6: Administering The Bolus Dose

    Administering the bolus dose Error functions Error functions for MARK I PUMP only Error Message Description Action While the pump is on, press and hold until you hear a long beep - then release Unfeasible operation No action required The screen will show the bolus dose being infused with flashing.
  • Page 7 APO-go Technical Helpline 0844 880 1327 24 hours a day, 365 days a year Part of the Package of Care provided by Britannia Pharmaceuticals Adverse events should be reported. Reporting forms and information can be found at www.mhra.gov.uk/yellowcard Adverse events should also be reported to Medical Information on 0870 851 0207 or dso@britannia-pharm.com...