PHscan10L Pocket pH Tester Thank you for selecting the PHscan10L pocket pH tester. This manual provides a step-by-step guide to help you operate the meter, please carefully read the following instructions before use. Installing the Batteries • Remove the battery compartment lid.
PHscan10L Pocket pH Tester Prior to Use • Remove the electrode protective cap from the tester. • If the glass sensitive membrane on the electrode dries out, soak the electrode in 3M KCL solution or tap water for at least 15 minutes. DO NOT use distilled or deionized water, it will render the electrode useless.
Page 4
Calibration PHscan10L pocket pH tester must be calibrated prior to first use or new electrode replaced. To ensure accuracy, regular calibration is recommended. DO NOT reuse the pH buffer solution after calibration, contaminants in solution will affect the calibration and eventually the accuracy of the measurement.
PHscan10L Pocket pH Tester Rinse the electrode with distilled water. Place the electrode into the pH4.01 (or 10.01) standard buffer solution. Stir the tester gently. Press the key. Wait for the reading to stabilize, the calibration value will flash Enter three times and the CAL icon disappears from the display.
Electrode Selection Guide PHscan10L pocket pH tester installed with a long-reach pH electrode that is used to measure the pH of the small volume samples. If this electrode can not meet your measurement requirements, please refer to the table below to select an applicable sensor.
PHscan10L Pocket pH Tester Specifications Model PHscan10L Range 0.0~14.0pH Accuracy ± 0.1pH Resolution 0.1pH Calibration Points 2 points pH Buffer Options USA (pH4.01/7.00/10.01) Automatic Buffer Recognition Sensor Type E-PHSCAN-L Operating Temperature 0~60° C, 32~140° F Relative Humidity < 80% Power Requirements 3 ×...
Page 8
Hazardous Substance Statement Bante Instruments is committed to the reduction and eventual elimination of all hazardous substances in both the manufacturing process and finished products we supply. We have an active manufacturing and procurement program to minimize and eliminate the use of harmful heavy metals such as cadmium, lead, mercury and the like.