Condec DLR3110 Technical Manual

Rack-mountable pneumatic pressure calibration console
Table of Contents

Advertisement

Quick Links

DLR3110
Rack-mountable Pneumatic
Pressure Calibration Console
Technical Manual
• Operation
• Calibration
• Maintenance
• Service
CONDEC Sales Phone No.: (888) 295-8475
CONDEC Web Site: WWW.4CONDEC.COM
74843

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the DLR3110 and is the answer not in the manual?

Questions and answers

Summary of Contents for Condec DLR3110

  • Page 1 DLR3110 Rack-mountable Pneumatic Pressure Calibration Console Technical Manual • Operation • Calibration • Maintenance • Service CONDEC Sales Phone No.: (888) 295-8475 CONDEC Web Site: WWW.4CONDEC.COM 74843...
  • Page 3: Table Of Contents

    Contents About This Manual ........................... 1 Introduction..........................1 Installation and Wiring ........................ 3 2.1 Unpacking and Inspection ............3 2.2 Repacking .
  • Page 4 Auto Selection Default Tables ....................63 10.0 Conversion Factors ........................66 11.0 Options, Replacement Kits ......................67 12.0 Specifications ..........................68 DLR3110 Warranty and Return Policy....................69 DLR3110 Return Material Authorization Form..................70 DLR3110 Rack Mount Digital Pressure Indicator Manual...
  • Page 5: About This Manual

    (GAGE models only). Introduction Using microprocessor technology, the rack-mountable DLR3110 instrument offers a combination of features, performance, versatility and reliability not previously available in a single, rack-mount pressure calibration instrument. Some of the features of the DLR3110 are listed below: •...
  • Page 6 (on GAGE models). Over-pressure protection, based on model, is provided via a fully adjustable pressure regulator which is manually set for each new device being tested. Figure 1-1 provides an overview of the DLR3110’s function. VERNIER ADJUST...
  • Page 7: Installation And Wiring

    Keep original shipping carton for repacking in case repair is necessary. Repacking If the DLR3110 must be returned for repair, recalibration, or modification, be sure that it is properly cushioned and packed and that a description of the work to be performed is included. The original shipping carton should be kept for this purpose.
  • Page 8: Hardware Configuration And Wiring

    (5). In the normal operating mode this key is used to select and view all active displays that were enabled in the Configuration mode (i.e., Pressure data, MIN/MAX data, Freeze mode data, and delta P display mode data). DLR3110 Rack Mount Digital Pressure Calibrator Manual...
  • Page 9: Freeze Mode

    • function key (6). The AUTO TARE (Auto.tr) feature is enabled in the Configuration mode under TARE the ZERO SET selection. When AUTO TARE is enabled, pressing the key automatically tares out TARE the value of the current pressure display reading and shifts the display to the P (delta P) mode. The resultant ZERO reference reading is useful in cases where it is desirable to subtract the atmospheric pressure from an absolute reading to obtain a quick gauge reference.
  • Page 10: Freeze Mode Specifications

    The following describes the setup for connecting the optional six-digit remote display (PN 74683). The display duplicates the DLR3110 six-digit front panel display for all ranges and units selected (see Figure 2-5 on page 7). NOTE: Remote display does not show current unit or mode designation.
  • Page 11 JUMPER PLUGS SHOWN ON S6 & S7 CONFIGURED AS RS232 WIRE\PCOLOR BLACK P2-4 ISO GND SIG COM P2-5 RS232 RXD REMOTE DISPLAY Figure 2-5. DLR3110 CPU to Remote Display PN 74683 Wiring Installation and Wiring...
  • Page 12: Operation

    4. Connect the male end of the test hose to the TEST PORT (6) fitting. NOTE: DLR3110 may be ordered with Test port on rear of unit. See Section 2.3 on page 3. 5. Connect the swivel fitting end (7/16-20) of the test (output) hose to the DUT using adapters if required.
  • Page 13 3. If pressure measurements above and below atmospheric pressure are required, connect a VACUUM PUMP to the VACUUM/VENT port (2) as shown in Figure 2-1 on page 3. 4. Open the VENT valve (2), close the INPUT valve (1) and apply power to the vacuum pump and allow it to evacuate the system for several minutes or until the digital display reading reaches equilibrium near 0 PSIA.
  • Page 14: Re-Calibration Only

    However, it must be emphasized that when performing these tests, the computer within the DLR3110 is actually being re-programmed and as such, it is imperative that the pressure standard being used is in satisfactory operating condition and that the technician fully understands its operating characteristics and methods of usage.
  • Page 15 10. The alpha-numeric display reads . Save the calibration data by opening rocker switch FILTER 5 position . DLR3110 returns to normal mode. 11. Vent system and remove test standard. "ON-OFF" 2-WAY VALVE VENT PNEUMATIC PRESSURE...
  • Page 16: Configuration And Calibration

    Pneumatic Calibration Set-Up Figure 5-1 defines a typical calibration set-up using a floating piston type dead weight tester. Any precision standard is acceptable to use providing its basic accuracy is at least twice that of the DLR3110’s. Instrument Calibration Set-Up Check that the INPUT valve (1) is closed (rotate clockwise until it stops) and that the VENT valve (2) is open (two turns counter-clockwise from its stop).
  • Page 17: Main Set-Up Mode Menu Selections

    DIAG DIAG MAIN MAIN EXPANDER FRONT OF DLR3110 Figure 5-2. Rocker Switch S1 Location Main Set-Up Mode Menu Selections Enter the main setup mode by closing switch located on the main board (see Section 5.3 on page 12). All other switch positions on must be open.
  • Page 18: Auto Set-Up Configuration Mode Selection

    AUTO ZERO DISPLAY TARE PRINT ENTER Figure 5-3. Auto Set-Up Mode Auto Set-Up Configuration Mode Selection The auto set-up automatically sets the following conditions in the DLR3110: • configuration parameters SET-UP - CONFIG • conversion parameters SET-UP - CONV •...
  • Page 19: Auto Set-Up Menu (Gage)

    Returns to XXX/CAPACITY Table 5-4. Auto Set-Up Mode (Gage) Parameters Selections NOTE: If you continue to press NO key past highest full scale range, DLR3110 will go to exit menu. Hit YES key at exit menu (see Table 5-3 on page 14).
  • Page 20: Auto Set-Up Menu (Absolute)

    Return to XXX/CAPACITY Table 5-5. Auto Set-Up Mode (Absolute) Parameters Selections NOTE: If you continue to press NO key past highest full scale range, DLR3110 will go to exit menu. Hit YES key at exit menu (see Table 5-3 on page 14).
  • Page 21: Calibration Mode Menu

    **XXX represents the present capacity of the DLR3110. Use the following steps to set calibration data in the DLR3110 using the Calibration mode menu. 1. Verify that the INPUT valve (1) is closed (rotate clockwise until it stops) and that the VENT valve (2) is open (two turns counter-clockwise from its stop).
  • Page 22 7. Save the calibration data by opening rocker switch position . The DLR3110 returns to normal mode (see Section 5.3 on page 12). 8. Vent system and remove test standard and/or vacuum pump. DLR3110 Rack Mount Digital Pressure Calibrator Manual...
  • Page 23: Modifying Data

    Modifying Data Data values previously set are easily altered by pressing either of the horizontal arrow keys to enter the modification mode. Upon entering the mode, the least significant digit begins to flash. A flashing digit (LSD) indicates the current position in the data that can be changed. Press the arrow key to select a LEFT RIGHT...
  • Page 24 To overwrite a span calibration point or hysteresis, step to the desired point using the DOWN arrow key and then modify and enter the new value as described in Section 5.6.1 on page 17. CONFIG ZERO TARE PRINT DISPLAY ENTER Figure 5-7. Configuration Set-Up Mode DLR3110 Rack Mount Digital Pressure Calibrator Manual...
  • Page 25: Configuration Set-Up Mode Selection

    See Figure 5-7 on page 20 fo scrolls continuously until the key is pressed. r an Exit? YES or NO example of the DLR3110’s display. LED Display Alpha-Numeric Display Key Pressed Remarks SET-UP CONFIG ?
  • Page 26: 5.12 Conversion Set-Up Mode Selection

    • PASCAL • • The following is a list of the inoperative conversions (allocated for future use) as they appear on the alpha-numeric display: • CONV.SP1 • CONV.SP2 • CONV.SP1 • CONV.SP2 DLR3110 Rack Mount Digital Pressure Calibrator Manual...
  • Page 27: 5.13 I/O Set-Up Mode Selection

    Time and date - sets the onboard clock for time and date (allows time and date to be transmitted in the serial output). • Parallel output - configures DLR3110 for BCD output. This is an option requiring the installation of the parallel output option board. •...
  • Page 28: Port 1 Set-Up Menu

    PORT 1 SIMPLEX See Table 5-17 on page 28 ---- EXIT ? YES key Return to Table 5-10 on page 24 NO key Return to PORT 1? Table 5-11. Port 1 Set-Up Menu DLR3110 Rack Mount Digital Pressure Calibrator Manual...
  • Page 29: Port 1 Printer Menu

    5.13.2 Port 1 Printer Menu Table 5-12 outlines the parameters of the Port 1 Printer menu. LED Display Alpha-Numeric Display Key Pressed Remarks PORT 1 PRINTER DOWN arrow Enters Printer set-up SET P1 AUTO PR? YES key Enters AUTO PR? menu. Use DOWN arrow key to scroll PRINT 1 –...
  • Page 30: Port 1 Duplex Menu

    Allocated for future use PORT 1 EXIT ? YES key Return to menu SET-UP/PORT 1 ?. See Table 5-11 on page 24. NO key Return to PORT 1/DUPLEX Table 5-14. Port 1 Duplex Menu DLR3110 Rack Mount Digital Pressure Calibrator Manual...
  • Page 31 Port 1 Duplex Modify Menu Table 5-15 outlines the parameters of the Port 1 Duplex Modify menu. LED Display Alpha-Numeric Display Key Pressed Remarks PORT 1 MODIFY DOWN arrow Enters Modify menu PORT 1 PTR OFF UP or DOWN Press LEFT or RIGHT arrow keys to arrow change the current state of the selection.
  • Page 32: Port 1 Rs485 Menu

    Allocated for future use PORT 1 EXIT ? YES key Return to menu SET-UP/PORT 1 ?. See Table 5-10 on page 24. NO key Return to PORT 1/SIMPLEX Table 5-17. Port 1 Simplex Parameters DLR3110 Rack Mount Digital Pressure Calibrator Manual...
  • Page 33: Port 2 Set-Up

    Port One Simplex Modify Menu Table 5-18 outlines the parameters of the Port 1 Simplex Modify menu. LED Display Alpha-Numeric Display Key Pressed Remarks PORT 1 MODIFY DOWN arrow Enters Modify menu DELAY 1 AUTO UP or DOWN Press LEFT or RIGHT arrow keys to arrow change the current state of the selection.
  • Page 34: Time And Date Set-Up Menu

    5.13.7 Time and Date Set-Up Menu Table 5-19 outlines the Time and Date Set-Up menu parameters of the DLR3110. LED Display Alpha-Numeric Display Key Pressed Remarks SET-UP TIME&D ? YES key Enters Time and Date menu STANDARD UP or DOWN...
  • Page 35: Print And Zero Wiring Configration Example

    5.13.9 Print and Zero Wiring Configration Example Table 5-21 illustrates a typical wiring configuration to operate the PRINT and ZERO commands remotely. LED Display Alpha-Numeric Display Remarks INPUT.1 ZERO INPUT 1 assigned as a ZERO input INPUT.2 PRINT INPUT 2 assigned as a PRINT input Table 5-21.
  • Page 36: Serial Input/Output

    Serial Input/Output The DLR3110 offers easy access and configuration of two serial communication ports. Both communication ports are ASCII-compatible and thus the serial data output formats are compatible with most printers, scoreboards, or other remote serial devices. Port One may be configured for bi-directional full duplex or half duplex (duplex = two way communication) mode of operation while Port Two is restricted to a simplex output mode only (simplex = one way communication).
  • Page 37: Printer (Port One Or Port Two)

    A more extensive document detailing the full operational parameters for serial duplex protocol is available upon request from CONDEC. Ask for Duplex Serial Protocol Manual, PN 74845. •...
  • Page 38: Full Duplex Protocol

    Weight calibration (Zero Cal, etc.) Data request code = < R > character (3FH). “R” commands are requests for data from the host to the Examples: • Weight/pressure data • Status • Setup data • Calibration data DLR3110 Rack Mount Digital Pressure Calibrator Manual...
  • Page 39: Basic Rs485 Commands

    Basic RS485 Commands Unit address 01 is used in Table 6-2 for illustration. NOTE: A more extensive document detailing the full RS485 command codes is available upon request from CONDEC. Ask for Duplex Serial Protocol Manual, PN 74845. Polling Command Function <*><01><00><PDR><cr>...
  • Page 40 The DLR3110 comes equipped with a 25 pin “D” female serial I/O connector (J6) mounted at the rear of the RS485 hook up. enclosure which provides access to Port One - RS232 or JUMPER PLUGS SHOWN ON S6 & S7 CONFIGURED AS RS232...
  • Page 41 JUMPER PLUGS SHOWN ON S6 & S7 CONFIGURED AS RS232 SIG COM SIG COM TXD (OUT) RXD (IN) REMOTE DEVICE Figure 6-5. Port One RS232 (Full Duplex) - Direct CPU Terminal Block (TB3) Wiring JUMPER PLUGS SHOWN ON S8 & S9 CONFIGURED AS PORT 1 RS485 -RX (-IN) -OUT +RX (+IN)
  • Page 42 JUMPER PLUGS SHOWN ON S8 & S9 CONFIGURED AS PORT 1 RS485 -RX (-IN) -OUT +RX (+IN) +OUT REMOTE DEVICE -OUT +OUT Figure 6-9. Port Two RS232 Active 20mA Current Loop (Simplex) - Direct CPU Terminal Block (TB3) Wiring DLR3110 Rack Mount Digital Pressure Calibrator Manual...
  • Page 43: Maintenance And Service

    Only those persons who are formally trained as skilled technicians should attempt to repair these units. All relevant safety precautions should be observed due to the presence of electrical components and high-pressure components. NOTE: Based on model, the following may give information on components that are not part of your DLR3110. 7.2.1 Panel/Chassis Removal and Installation...
  • Page 44: Orion-2C (Pn 55283) Or Orion-3A (Pn 55287) Manifold Removal

    8. Remove the valve stem seats (8) and valve needle seats (9) using the needle-nose pliers. 9. Remove the inner and outer O-rings (28, 27) and back-up rings (31, 30) from the valve stem seats and wash all parts in solvent (denatured alcohol). DLR3110 Rack Mount Digital Pressure Calibrator Manual...
  • Page 45: Orion Manifold - Vernier Control Disassembly

    10. To remove valve seats (7) from either the coarse (pressure), vent or isolation valves, try blowing compressed air through the inlet and outlet fittings. Otherwise, the center holes have to be drilled and a tap used to extract the seat (Steps 11-14). 11.
  • Page 46: Orion Manifold - Vernier Control Reassembly

    5. For input (pressure) and vent valves (two outer valves), disassemble the valve needle (11) from its housing (10) and check for any burrs or dirt on the threads which might interfere with smooth operation. DLR3110 Rack Mount Digital Pressure Calibrator Manual...
  • Page 47: Orion Manifold - Panel Installation

    6. Clean both the needle (11) and housing (10) in solvent, dry the parts and apply a small amount of fluorinated Krytox grease to the needle threads before reassembly. 7. Assemble the Valve Needle (11) into the Valve Needle Housing (10) and turn it until it stops. 8.
  • Page 48: Orion-2C Manifold - Valve Adjustment Procedure

    7.2.8 ORION-2C Manifold - Valve Adjustment Procedure For DLR3110 models 2500 PSI and below. ORION- 2C: Refer to Section 7.3 on page 53 and Figure 7-1 on page 55. Tools required: hex wrench (.050") hex wrench (.061") SNOOP, leak gas detector (PN 64781) NOTE: * Denotes reference to Figure 3-1 on page 9.
  • Page 49: Orion-3A Manifold - Valve Adjustment Procedure

    7.2.9 ORION-3A Manifold - Valve Adjustment Procedure For DLR3110 5000 PSI and above models. ORION- 3A: Refer to Section 7.4 on page 57 and Figure 7-2 on page 59. Tools required: hex wrench (.050") hex wrench (.061") SNOOP, leak gas detector (PN 64781) NOTES: 1.
  • Page 50: Pressure Limit Control (Standard Pneumatic), Regulator Removal

    1/2" from the panel surface, then tighten the cap screw on the mounting collar. 4. Using a pressure source set input pressure to approximately 100% of full scale rating of DLR3110 and check all fittings for leaks.
  • Page 51: Pressure Limit Control (Tescom), Regulator Removal

    Hold knob in position and install the locknut. 5. Close Pressure Limit Monitor, by turning regulator knob counter-clockwise. 6. Using a pressure source set input pressure to approximately 100% of full scale rating of DLR3110 and check all fittings for leaks.
  • Page 52: Panel Gauge Removal

    Connector on to the gauge. 4. Attach the tubing section that connects to the gauge fitting. 5. Using a pressure source set input pressure to approximately 100% of full scale rating of DLR3110 and check all fittings for leaks. 6. Install panel/chassis assembly in its enclosure as described in Section 7.2.1 on page 39.
  • Page 53: Test Port Filter (Pn 54188) Removal And Installation - 250 And 2500 Full Scale Psi

    Installation 1. To reinstall, reverse the order of steps 2 and 3 of Section 7.2.17 above. 2. Using a pressure source set input pressure to approximately 100% of full scale rating of DLR3110 and check all fittings for leaks. 3. Install panel/chassis assembly in its enclosure as described in Section 7.2.1 on page 39.
  • Page 54: Test Port (Output) Hose Quick-Connect Fitting And Filter Removal And Installation (Pn 56991)- 5000 And 10000 Full Scale Psi

    10. Using a pressure source set input pressure to approximately 100% of full scale rating of DLR3110 and check all fittings for leaks. 11. Install panel/chassis assembly in its enclosure as described in Section 7.2.1 on page 39. 7.2.19 Test Port (Output) Hose Quick-Connect Fitting and Filter Removal and Installation (PN 56991)- 5000 and 10000...
  • Page 55: Ac Fuse (Pn 58076) - Removal And Installation

    1. To reinstall, reverse the order of steps 2, 3 and 4 of the “Input Port Filter Removal” procedure above. 2. Using a pressure source set input pressure to approximately 100% of full scale rating of DLR3110 and check all fittings for leaks.
  • Page 56: Power Switch (Pn 58878) - Removal And Installation

    3. Tighten the switch mounting nut and lock washer from the rear of panel. Do not over tighten if using wrench. Doing so could result in damage to the switch. Caution 4. Install panel/chassis assembly in its enclosure as described in Section 7.2.1 on page 39. DLR3110 Rack Mount Digital Pressure Calibrator Manual...
  • Page 57: Orion-2C Valve Assembly Parts List

    ORION-2C Valve Assembly Parts List Number Description Quantity 57482 Nut,Valve Needle Housing 9/16-18, Width Across Flats=.75,Thickness=.12 54401 Locknut 58079 Knob 57889 Insert knob 57256 Spur 40-tooth gear 59233 Spur 18-tooth gear 55896 Valve seat 59387 Stem valve seat 59045 Needle seat valve 54540 Valve needle housing 59551...
  • Page 58 DLR3110 Rack Mount Digital Pressure Calibrator Manual...
  • Page 59 O-RING, ITEM 27, .239 ID O-RING, ITEM 28, .042 ID O-RING, ITEM 29, .799 ID O-RING, ITEM 32, .424 ID THESE COMPONENTS TO BE INSTALLED AT LOCATIONS B & C Figure 7-1. ORION-2C Exploded View DLR3110 Rack Mount Digital Pressure Calibrator Manual...
  • Page 60 DLR3110 Rack Mount Digital Pressure Calibrator Manual...
  • Page 61: Orion-3A Valve Assembly Parts List

    Packing backup retainer 55615 O-ring, Fluorocarbon (Viton) color black w/white dot 58314 Ball, chrome, steel 59731 Male connector, 1/8 tube x 1/8 NPT, stainless steel Table 7-4. ORION 3-A Valve Assembly Parts List DLR3110 Rack Mount Digital Pressure Calibrator Manual...
  • Page 62 Maintenance and Service...
  • Page 63 O-RING, ITEM 28, .042 ID O-RING, ITEM 29, .799 ID O-RING, ITEM 32, .424 ID O-RING, ITEM 40, .145 ID THESE COMPONENTS TO BE INSTALLED AT LOCATIONS B & C Figure 7-2. ORION-3A Exploded View DLR3110 Rack Mount Digital Pressure Calibrator Manual...
  • Page 64 DLR3110 Rack Mount Digital Pressure Calibrator Manual...
  • Page 65 Figure 7-3. Tescom and Standard Pneumatic Regulator Mounting Maintenance and Service...
  • Page 66: Model Numbering System

    C - Standard unit plus selectable scalable analog output 0-5VDC, 1-6VDC, 0-10VDC, 0-20mA, and 4-20mA D - Standard unit plus parallel BCD output NOTE: DLR3110 is available with test port on back of chassis. DLR3110 Rack Mount Digital Pressure Calibrator Manual...
  • Page 67: Auto Selection Default Tables

    Auto Selection Default Tables Auto Selection (Gage) Default Table Parameter 3 PSIG 10 PSIG 15 PSIG 25 PSIG 50 PSIG 100 PSIG 250 PSIG 500 PSIG CAPACITY CT-BY 0.0005 0.001 0.002 0.002 0.005 0.01 0.02 0.05 UNITS (Cal) TYPE GAGE GAGE GAGE GAGE...
  • Page 68 ZERO+/-% AZM +/- MOTION MIN PRES MAX PRES • mmHg0 • inHg0 mBAR kg/cm • • ftseaH TORR metrsH PASCAL • Table 9-2. Auto Selection (Gage) Default Table (1000 PSIG – 10000 PSIG) DLR3110 Rack Mount Digital Pressure Calibrator Manual...
  • Page 69 Auto Selection (Absolute) Default Table 1000 2500 10000 PARAMETER PSIA PSIA PSIA PSIA PSIA PSIA PSIA PSIA PSIA CAPACITY 1000 2500 10000 CT-BY 0.002 0.002 0.005 0.01 0.02 0..05 UNITS (CAL) TYPE CAL SPAN 1 REF CAL SPAN 2 1000 CAL SPAN 3 2000 CAL SPAN 4...
  • Page 70: Conversion Factors

    Kg/cm • 27.71 2.771000e+01 O 60 • 27.729 2.772900e+01 O 68 27.68 2.768000e+01 • ftseaH 2.252 2.252000e+00 TORR 51.7149 5.171490e+01 metrsH 0.7030696 7.030696e-01 PACAL 6894.757 6.894757e+03 • 703.07 7.030700e+02 Table 10-1. Conversion Factors DLR3110 Rack Mount Digital Pressure Calibrator Manual...
  • Page 71: Options, Replacement Kits

    11.0 Options, Replacement Kits There are numerous replacement part numbers mentioned throughout manual that may be ordered. ORION-2C and ORION-3A O-Ring Replacement Kit (data sheet # 65308) • Fluorocarbon “Viton” (standard for ORION-3A)...........PN 55277 • Nitrile Buna-N (standard for ORION-2C)............PN 58499 •...
  • Page 72: Specifications

    All other parts 300 series stainless steel *Over-pressure Rupture Disk: Rating: 3000 PSIG, nominal Type: Stainless steel outer case *Relief Valve: Type: Adjustable, atmospheric bleed Setting: Adjustable to 10% above highest calibrated pressure Material: 300 series stainless steel DLR3110 Rack Mount Digital Pressure Calibrator Manual...
  • Page 73: Dlr3110 Warranty And Return Policy

    DLR3110 Warranty and Return Policy If possible, please save original packing material, which is specifically designed for the unit. Should it be necessary to ship the unit back to the factory, a suitable shipping container must be used along with sufficient packing material.
  • Page 74: Dlr3110 Return Material Authorization Form

    COMPANY NAME: STREET: CITY, STATE, ZIP: ATTN: CONDEC • 3 SIMM LANE • DOOR D, UNIT 2A • NEWTOWN, CT 06470 ATTN: PRESSURE PRODUCTS/REPAIR LAB TEL: 888-295-8475 • FAX: 203-364-1556 or 715-234-6967 WEB SITE: www.4condec.com DLR3110 Rack Mount Digital Pressure Calibrator Manual...