Contents About This Manual ........................... 1 Introduction..........................1 Installation and Wiring ........................ 3 2.1 Unpacking and Inspection ............3 2.2 Repacking .
Page 4
Auto Selection Default Tables ....................63 10.0 Conversion Factors ........................66 11.0 Options, Replacement Kits ......................67 12.0 Specifications ..........................68 DLR3110 Warranty and Return Policy....................69 DLR3110 Return Material Authorization Form..................70 DLR3110 Rack Mount Digital Pressure Indicator Manual...
(GAGE models only). Introduction Using microprocessor technology, the rack-mountable DLR3110 instrument offers a combination of features, performance, versatility and reliability not previously available in a single, rack-mount pressure calibration instrument. Some of the features of the DLR3110 are listed below: •...
Page 6
(on GAGE models). Over-pressure protection, based on model, is provided via a fully adjustable pressure regulator which is manually set for each new device being tested. Figure 1-1 provides an overview of the DLR3110’s function. VERNIER ADJUST...
Keep original shipping carton for repacking in case repair is necessary. Repacking If the DLR3110 must be returned for repair, recalibration, or modification, be sure that it is properly cushioned and packed and that a description of the work to be performed is included. The original shipping carton should be kept for this purpose.
(5). In the normal operating mode this key is used to select and view all active displays that were enabled in the Configuration mode (i.e., Pressure data, MIN/MAX data, Freeze mode data, and delta P display mode data). DLR3110 Rack Mount Digital Pressure Calibrator Manual...
• function key (6). The AUTO TARE (Auto.tr) feature is enabled in the Configuration mode under TARE the ZERO SET selection. When AUTO TARE is enabled, pressing the key automatically tares out TARE the value of the current pressure display reading and shifts the display to the P (delta P) mode. The resultant ZERO reference reading is useful in cases where it is desirable to subtract the atmospheric pressure from an absolute reading to obtain a quick gauge reference.
The following describes the setup for connecting the optional six-digit remote display (PN 74683). The display duplicates the DLR3110 six-digit front panel display for all ranges and units selected (see Figure 2-5 on page 7). NOTE: Remote display does not show current unit or mode designation.
Page 11
JUMPER PLUGS SHOWN ON S6 & S7 CONFIGURED AS RS232 WIRE\PCOLOR BLACK P2-4 ISO GND SIG COM P2-5 RS232 RXD REMOTE DISPLAY Figure 2-5. DLR3110 CPU to Remote Display PN 74683 Wiring Installation and Wiring...
4. Connect the male end of the test hose to the TEST PORT (6) fitting. NOTE: DLR3110 may be ordered with Test port on rear of unit. See Section 2.3 on page 3. 5. Connect the swivel fitting end (7/16-20) of the test (output) hose to the DUT using adapters if required.
Page 13
3. If pressure measurements above and below atmospheric pressure are required, connect a VACUUM PUMP to the VACUUM/VENT port (2) as shown in Figure 2-1 on page 3. 4. Open the VENT valve (2), close the INPUT valve (1) and apply power to the vacuum pump and allow it to evacuate the system for several minutes or until the digital display reading reaches equilibrium near 0 PSIA.
However, it must be emphasized that when performing these tests, the computer within the DLR3110 is actually being re-programmed and as such, it is imperative that the pressure standard being used is in satisfactory operating condition and that the technician fully understands its operating characteristics and methods of usage.
Page 15
10. The alpha-numeric display reads . Save the calibration data by opening rocker switch FILTER 5 position . DLR3110 returns to normal mode. 11. Vent system and remove test standard. "ON-OFF" 2-WAY VALVE VENT PNEUMATIC PRESSURE...
Pneumatic Calibration Set-Up Figure 5-1 defines a typical calibration set-up using a floating piston type dead weight tester. Any precision standard is acceptable to use providing its basic accuracy is at least twice that of the DLR3110’s. Instrument Calibration Set-Up Check that the INPUT valve (1) is closed (rotate clockwise until it stops) and that the VENT valve (2) is open (two turns counter-clockwise from its stop).
DIAG DIAG MAIN MAIN EXPANDER FRONT OF DLR3110 Figure 5-2. Rocker Switch S1 Location Main Set-Up Mode Menu Selections Enter the main setup mode by closing switch located on the main board (see Section 5.3 on page 12). All other switch positions on must be open.
AUTO ZERO DISPLAY TARE PRINT ENTER Figure 5-3. Auto Set-Up Mode Auto Set-Up Configuration Mode Selection The auto set-up automatically sets the following conditions in the DLR3110: • configuration parameters SET-UP - CONFIG • conversion parameters SET-UP - CONV •...
Returns to XXX/CAPACITY Table 5-4. Auto Set-Up Mode (Gage) Parameters Selections NOTE: If you continue to press NO key past highest full scale range, DLR3110 will go to exit menu. Hit YES key at exit menu (see Table 5-3 on page 14).
Return to XXX/CAPACITY Table 5-5. Auto Set-Up Mode (Absolute) Parameters Selections NOTE: If you continue to press NO key past highest full scale range, DLR3110 will go to exit menu. Hit YES key at exit menu (see Table 5-3 on page 14).
**XXX represents the present capacity of the DLR3110. Use the following steps to set calibration data in the DLR3110 using the Calibration mode menu. 1. Verify that the INPUT valve (1) is closed (rotate clockwise until it stops) and that the VENT valve (2) is open (two turns counter-clockwise from its stop).
Page 22
7. Save the calibration data by opening rocker switch position . The DLR3110 returns to normal mode (see Section 5.3 on page 12). 8. Vent system and remove test standard and/or vacuum pump. DLR3110 Rack Mount Digital Pressure Calibrator Manual...
Modifying Data Data values previously set are easily altered by pressing either of the horizontal arrow keys to enter the modification mode. Upon entering the mode, the least significant digit begins to flash. A flashing digit (LSD) indicates the current position in the data that can be changed. Press the arrow key to select a LEFT RIGHT...
Page 24
To overwrite a span calibration point or hysteresis, step to the desired point using the DOWN arrow key and then modify and enter the new value as described in Section 5.6.1 on page 17. CONFIG ZERO TARE PRINT DISPLAY ENTER Figure 5-7. Configuration Set-Up Mode DLR3110 Rack Mount Digital Pressure Calibrator Manual...
See Figure 5-7 on page 20 fo scrolls continuously until the key is pressed. r an Exit? YES or NO example of the DLR3110’s display. LED Display Alpha-Numeric Display Key Pressed Remarks SET-UP CONFIG ?
• PASCAL • • The following is a list of the inoperative conversions (allocated for future use) as they appear on the alpha-numeric display: • CONV.SP1 • CONV.SP2 • CONV.SP1 • CONV.SP2 DLR3110 Rack Mount Digital Pressure Calibrator Manual...
Time and date - sets the onboard clock for time and date (allows time and date to be transmitted in the serial output). • Parallel output - configures DLR3110 for BCD output. This is an option requiring the installation of the parallel output option board. •...
PORT 1 SIMPLEX See Table 5-17 on page 28 ---- EXIT ? YES key Return to Table 5-10 on page 24 NO key Return to PORT 1? Table 5-11. Port 1 Set-Up Menu DLR3110 Rack Mount Digital Pressure Calibrator Manual...
5.13.2 Port 1 Printer Menu Table 5-12 outlines the parameters of the Port 1 Printer menu. LED Display Alpha-Numeric Display Key Pressed Remarks PORT 1 PRINTER DOWN arrow Enters Printer set-up SET P1 AUTO PR? YES key Enters AUTO PR? menu. Use DOWN arrow key to scroll PRINT 1 –...
Allocated for future use PORT 1 EXIT ? YES key Return to menu SET-UP/PORT 1 ?. See Table 5-11 on page 24. NO key Return to PORT 1/DUPLEX Table 5-14. Port 1 Duplex Menu DLR3110 Rack Mount Digital Pressure Calibrator Manual...
Page 31
Port 1 Duplex Modify Menu Table 5-15 outlines the parameters of the Port 1 Duplex Modify menu. LED Display Alpha-Numeric Display Key Pressed Remarks PORT 1 MODIFY DOWN arrow Enters Modify menu PORT 1 PTR OFF UP or DOWN Press LEFT or RIGHT arrow keys to arrow change the current state of the selection.
Allocated for future use PORT 1 EXIT ? YES key Return to menu SET-UP/PORT 1 ?. See Table 5-10 on page 24. NO key Return to PORT 1/SIMPLEX Table 5-17. Port 1 Simplex Parameters DLR3110 Rack Mount Digital Pressure Calibrator Manual...
Port One Simplex Modify Menu Table 5-18 outlines the parameters of the Port 1 Simplex Modify menu. LED Display Alpha-Numeric Display Key Pressed Remarks PORT 1 MODIFY DOWN arrow Enters Modify menu DELAY 1 AUTO UP or DOWN Press LEFT or RIGHT arrow keys to arrow change the current state of the selection.
5.13.7 Time and Date Set-Up Menu Table 5-19 outlines the Time and Date Set-Up menu parameters of the DLR3110. LED Display Alpha-Numeric Display Key Pressed Remarks SET-UP TIME&D ? YES key Enters Time and Date menu STANDARD UP or DOWN...
5.13.9 Print and Zero Wiring Configration Example Table 5-21 illustrates a typical wiring configuration to operate the PRINT and ZERO commands remotely. LED Display Alpha-Numeric Display Remarks INPUT.1 ZERO INPUT 1 assigned as a ZERO input INPUT.2 PRINT INPUT 2 assigned as a PRINT input Table 5-21.
Serial Input/Output The DLR3110 offers easy access and configuration of two serial communication ports. Both communication ports are ASCII-compatible and thus the serial data output formats are compatible with most printers, scoreboards, or other remote serial devices. Port One may be configured for bi-directional full duplex or half duplex (duplex = two way communication) mode of operation while Port Two is restricted to a simplex output mode only (simplex = one way communication).
A more extensive document detailing the full operational parameters for serial duplex protocol is available upon request from CONDEC. Ask for Duplex Serial Protocol Manual, PN 74845. •...
Weight calibration (Zero Cal, etc.) Data request code = < R > character (3FH). “R” commands are requests for data from the host to the Examples: • Weight/pressure data • Status • Setup data • Calibration data DLR3110 Rack Mount Digital Pressure Calibrator Manual...
Basic RS485 Commands Unit address 01 is used in Table 6-2 for illustration. NOTE: A more extensive document detailing the full RS485 command codes is available upon request from CONDEC. Ask for Duplex Serial Protocol Manual, PN 74845. Polling Command Function <*><01><00><PDR><cr>...
Page 40
The DLR3110 comes equipped with a 25 pin “D” female serial I/O connector (J6) mounted at the rear of the RS485 hook up. enclosure which provides access to Port One - RS232 or JUMPER PLUGS SHOWN ON S6 & S7 CONFIGURED AS RS232...
Page 41
JUMPER PLUGS SHOWN ON S6 & S7 CONFIGURED AS RS232 SIG COM SIG COM TXD (OUT) RXD (IN) REMOTE DEVICE Figure 6-5. Port One RS232 (Full Duplex) - Direct CPU Terminal Block (TB3) Wiring JUMPER PLUGS SHOWN ON S8 & S9 CONFIGURED AS PORT 1 RS485 -RX (-IN) -OUT +RX (+IN)
Page 42
JUMPER PLUGS SHOWN ON S8 & S9 CONFIGURED AS PORT 1 RS485 -RX (-IN) -OUT +RX (+IN) +OUT REMOTE DEVICE -OUT +OUT Figure 6-9. Port Two RS232 Active 20mA Current Loop (Simplex) - Direct CPU Terminal Block (TB3) Wiring DLR3110 Rack Mount Digital Pressure Calibrator Manual...
Only those persons who are formally trained as skilled technicians should attempt to repair these units. All relevant safety precautions should be observed due to the presence of electrical components and high-pressure components. NOTE: Based on model, the following may give information on components that are not part of your DLR3110. 7.2.1 Panel/Chassis Removal and Installation...
8. Remove the valve stem seats (8) and valve needle seats (9) using the needle-nose pliers. 9. Remove the inner and outer O-rings (28, 27) and back-up rings (31, 30) from the valve stem seats and wash all parts in solvent (denatured alcohol). DLR3110 Rack Mount Digital Pressure Calibrator Manual...
10. To remove valve seats (7) from either the coarse (pressure), vent or isolation valves, try blowing compressed air through the inlet and outlet fittings. Otherwise, the center holes have to be drilled and a tap used to extract the seat (Steps 11-14). 11.
5. For input (pressure) and vent valves (two outer valves), disassemble the valve needle (11) from its housing (10) and check for any burrs or dirt on the threads which might interfere with smooth operation. DLR3110 Rack Mount Digital Pressure Calibrator Manual...
6. Clean both the needle (11) and housing (10) in solvent, dry the parts and apply a small amount of fluorinated Krytox grease to the needle threads before reassembly. 7. Assemble the Valve Needle (11) into the Valve Needle Housing (10) and turn it until it stops. 8.
1/2" from the panel surface, then tighten the cap screw on the mounting collar. 4. Using a pressure source set input pressure to approximately 100% of full scale rating of DLR3110 and check all fittings for leaks.
Hold knob in position and install the locknut. 5. Close Pressure Limit Monitor, by turning regulator knob counter-clockwise. 6. Using a pressure source set input pressure to approximately 100% of full scale rating of DLR3110 and check all fittings for leaks.
Connector on to the gauge. 4. Attach the tubing section that connects to the gauge fitting. 5. Using a pressure source set input pressure to approximately 100% of full scale rating of DLR3110 and check all fittings for leaks. 6. Install panel/chassis assembly in its enclosure as described in Section 7.2.1 on page 39.
Installation 1. To reinstall, reverse the order of steps 2 and 3 of Section 7.2.17 above. 2. Using a pressure source set input pressure to approximately 100% of full scale rating of DLR3110 and check all fittings for leaks. 3. Install panel/chassis assembly in its enclosure as described in Section 7.2.1 on page 39.
10. Using a pressure source set input pressure to approximately 100% of full scale rating of DLR3110 and check all fittings for leaks. 11. Install panel/chassis assembly in its enclosure as described in Section 7.2.1 on page 39. 7.2.19 Test Port (Output) Hose Quick-Connect Fitting and Filter Removal and Installation (PN 56991)- 5000 and 10000...
1. To reinstall, reverse the order of steps 2, 3 and 4 of the “Input Port Filter Removal” procedure above. 2. Using a pressure source set input pressure to approximately 100% of full scale rating of DLR3110 and check all fittings for leaks.
3. Tighten the switch mounting nut and lock washer from the rear of panel. Do not over tighten if using wrench. Doing so could result in damage to the switch. Caution 4. Install panel/chassis assembly in its enclosure as described in Section 7.2.1 on page 39. DLR3110 Rack Mount Digital Pressure Calibrator Manual...
Page 58
DLR3110 Rack Mount Digital Pressure Calibrator Manual...
Page 59
O-RING, ITEM 27, .239 ID O-RING, ITEM 28, .042 ID O-RING, ITEM 29, .799 ID O-RING, ITEM 32, .424 ID THESE COMPONENTS TO BE INSTALLED AT LOCATIONS B & C Figure 7-1. ORION-2C Exploded View DLR3110 Rack Mount Digital Pressure Calibrator Manual...
Page 60
DLR3110 Rack Mount Digital Pressure Calibrator Manual...
Packing backup retainer 55615 O-ring, Fluorocarbon (Viton) color black w/white dot 58314 Ball, chrome, steel 59731 Male connector, 1/8 tube x 1/8 NPT, stainless steel Table 7-4. ORION 3-A Valve Assembly Parts List DLR3110 Rack Mount Digital Pressure Calibrator Manual...
Page 63
O-RING, ITEM 28, .042 ID O-RING, ITEM 29, .799 ID O-RING, ITEM 32, .424 ID O-RING, ITEM 40, .145 ID THESE COMPONENTS TO BE INSTALLED AT LOCATIONS B & C Figure 7-2. ORION-3A Exploded View DLR3110 Rack Mount Digital Pressure Calibrator Manual...
Page 64
DLR3110 Rack Mount Digital Pressure Calibrator Manual...
Page 65
Figure 7-3. Tescom and Standard Pneumatic Regulator Mounting Maintenance and Service...
C - Standard unit plus selectable scalable analog output 0-5VDC, 1-6VDC, 0-10VDC, 0-20mA, and 4-20mA D - Standard unit plus parallel BCD output NOTE: DLR3110 is available with test port on back of chassis. DLR3110 Rack Mount Digital Pressure Calibrator Manual...
11.0 Options, Replacement Kits There are numerous replacement part numbers mentioned throughout manual that may be ordered. ORION-2C and ORION-3A O-Ring Replacement Kit (data sheet # 65308) • Fluorocarbon “Viton” (standard for ORION-3A)...........PN 55277 • Nitrile Buna-N (standard for ORION-2C)............PN 58499 •...
All other parts 300 series stainless steel *Over-pressure Rupture Disk: Rating: 3000 PSIG, nominal Type: Stainless steel outer case *Relief Valve: Type: Adjustable, atmospheric bleed Setting: Adjustable to 10% above highest calibrated pressure Material: 300 series stainless steel DLR3110 Rack Mount Digital Pressure Calibrator Manual...
DLR3110 Warranty and Return Policy If possible, please save original packing material, which is specifically designed for the unit. Should it be necessary to ship the unit back to the factory, a suitable shipping container must be used along with sufficient packing material.