Site Requirements; Space Requirements; Telephone; Environmental Requirements - HP 5973 Hardware Installation Manual

Mass selective detector
Hide thumbs Also See for 5973:
Table of Contents

Advertisement

 General Information
6LWH UHTXLUHPHQWV

Site requirements

Space requirements

Adequate space for the MSD, GC, data system, and other peripheral devices must
be provided. The sizes and weights of the MSD system components are listed in
Table 1. A minimum of 30 cm of unobstructed distance between the rear of the MSD
and the nearest wall must be provided for service access and air circulation.

Telephone

A telephone with a cord long enough to be used at the computer will allow the MSD
operator to communicate with HP support personnel. In addition, a second, analog
phone line is strongly recommended to allow remote control and diagnosis of the
MSD through a modem. This is required for some service contracts.

Environmental requirements

Indoor use only.
Thermal outputs of the MSD system components are listed in Table 1.
Temperature
Operating: 15°C – 35°C (constant temperature = ±2°C per hour)
Storage: -20°C – 70°C
Humidity
Operating: 15% - 95% relative humidity
Storage: 65°C at 90% relative humidity for 24 hours
Altitude
Operating: 4.6km (15,000 ft)
Storage: 4.6km (15,000 ft)
9