The readings of the additional sensors are shown at the third page. Those values represent
the average derived from all "one minute shots" of the whole integration interval.
Ambient
12:20
21.5°C
987mbar
46%rH
12.3V
TSC-00256
The next page shows the average values of the Radon and Thoron concentration from the
very begin of the actual measurement series. The total sampling time is given in the first row.
Average
68.0Hrs
Rn:
314Bq/m³
Tn:
141Bq/m³
TSC-00256
The last page contains the status information, beginning with the date and time of the start of
the measurement series followed by the actual alert settings in the second line. The lower
line shows the mode for the acoustic signal (buzzer).
>>15/06/19
ALM:
10000Bq/m³
INTVL.
BUZZ.OFF
TSC-00256
To finish a measurement series keep the Toggle button down and wait for at least four beeps
from the buzzer. If the button is released, the sample will be stopped, and the device will be
brought into the stand-by mode.
Thoron-Scout
SN:00256
If the button has been locked by software, the button has to be unlocked before.

Adjustment of the sampling interval

The adjustment can be carried out as long the sampling is stopped. The Toggle button must
be hold down for at least 6 seconds (beeper). On the display appears:
INTERVAL:
1min
Now, the interval can be toggled by the button between 1, 5, 10, 15, 30, 60 and 120 minutes.
To accept the new setting, the button must be pressed again for at least 6 seconds.

Operation Modes

All the settings for the operation mode of the device can be performed with usage of the
software application delivered with the unit. Please refer to the software manual for further
instructions.