“Operating the Pure-Vu Workstation”. • The Pure-Vu System should never be stored in areas where the unit could get wet or be exposed to high temperature, humidity, direct sunlight, dust, salt, which could adversely affect the equipment.
1.1. Symbols Used The Pure-Vu System has a safety warning and compliance label located at the rear of the WS. For a defi nition of individual symbols, see table below: SYMBOL DESCRIPTION FDA (USA) law restricts this device to sale by or on order...
PRODUCT DESCRIPTION The Pure-Vu System enables colon cleansing during colonoscopy using a standard or slim colonoscope with a length of 1630mm – 1710mm and a slim colonoscope outer diameter range of 11.7mm – 13.3mm and a standard colonoscope outer diameter range of 12.8mm – 13.7mm. The Oversleeve, which fi ts over the colonoscope and is connected to an external Workstation, generates fl uid to break up feces.
[PC]. The Pure-Vu System undergoes a system test when the System Test button [ST] is activated. The system test ensures The Pure-Vu System is functioning properly prior to the start of the procedure. The system can be operated in three cleansing modes; low, medium (default mode) and high.
Cleansing Mode Selector – activated by the right button [CMS]. User can switch between three cleansing modes; low, medium (default mode) and high. Manual Purging Cleansing Mode [MP] Selector [CMS] Evacuation Cleansing [EP] [CP] FIGURE 3: Pure-Vu Foot Pedal Unit MOTUS GI PURE-VU SYSTEM INSTRUCTIONS FOR USE...
Oversleeve’s conduits which branch off and attach to the WS Connector [WSC]. The WS Connector [WSC] connects the Pure-Vu Oversleeve to the Workstation. 3.3. Accessories for Use with Pure-Vu The following waste receptacle accessories have been validated be used with The Pure-Vu System. MANUFACTURER REFERENCE CODE...
(D), distal channel cover (B) and distal end clamp (E). Place the Oversleeve (F) into the loading fi xture channel with the Oversleeve’s infl ation hub luer connector facing up. Close and lock the Proximal clamp (D). MOTUS GI PURE-VU SYSTEM INSTRUCTIONS FOR USE...
Page 11
Then place the colonoscope connector (C) and control section on the loading fi xture. Rotate the colonoscope holders to lock. Ensure all of the colonoscope valves (D) are in place and secure. MOTUS GI PURE-VU SYSTEM INSTRUCTIONS FOR USE...
Page 12
Pure-Vu Oversleeve (C). Continue inserting the colonoscope until it reaches the distal end and will not advance with a gentle push. Place the colonoscope control section onto the scope support bracket (D). MOTUS GI PURE-VU SYSTEM INSTRUCTIONS FOR USE...
Turn on the Workstation by pressing the power button (A). Place the Workstation connector on front of the Workstation and slide down into place (B). Push the Workstation connector lock button (C). MOTUS GI PURE-VU SYSTEM INSTRUCTIONS FOR USE...
Page 14
FIGURE 11: Close pump heads (A). Snap evacuation funnels into connector (B). Connect the luer fi ttings on the umbilical tubing to the luer fi ttings (C) on the Workstation connector by matching the corresponding colors (Clear, Blue and Red). MOTUS GI PURE-VU SYSTEM INSTRUCTIONS FOR USE...
Page 15
FIGURE 13: Confi rm presence of green lights. Point the tip of the colonoscope (A) and Pure-Vu Oversleeve (B) toward a container (C) and press the system test button (D) for 2 seconds to verify irrigation. The Pure-Vu System is now ready to use. MOTUS GI PURE-VU SYSTEM INSTRUCTIONS FOR USE...
[PW], disconnect and discard the Oversleeve. Assemble new Pure-Vu Oversleeve and repeat steps outlined in sections 4 & 5. After determining The Pure-Vu System is ready for use apply a water based gel (as used in standard colonoscopy procedures) to distal 30cm of the Oversleeve.
FIGURE 15: Cut the umbilical tubing (A) approximately 15 cm from the proximal end of the Oversleeve. Open Workstation pumps (B) and remove all tubing (C) then discard the Workstation connector (D) and umbilical tubing as per facility protocol. MOTUS GI PURE-VU SYSTEM INSTRUCTIONS FOR USE...
Connect the stopcock and syringe assembly to the Oversleeve luer hub. Flush the 60 cc volume of diluted detergent through the Oversleeve and look for the solution to exit the distal end. MOTUS GI PURE-VU SYSTEM INSTRUCTIONS FOR USE...
Page 19
(do not clamp the endoscope). Turn the stopcock off valve toward the Oversleeve, and draw up 60 cc of air. Turn the off valve to the side port position and infuse the air. Close by turning the off valve towards the Oversleeve. MOTUS GI PURE-VU SYSTEM INSTRUCTIONS FOR USE...
Page 20
Oversleeve by pulling the Oversleeve and endoscope away from each other) (B). Point the endoscope tip and Oversleeve hub downward just before the endoscope leaves the Oversleeve to avoid spray. Discard per facility protocol. MOTUS GI PURE-VU SYSTEM INSTRUCTIONS FOR USE...
The Pure-Vu Workstation and Oversleeve should never be installed, used or transported in an inclined position nor should it be subjected to impact or vibrations. MOTUS GI PURE-VU SYSTEM INSTRUCTIONS FOR USE...
• While maintaining pressure on the plunger, turn the off valve towards the Oversleeve • Proceed to separate the endoscope from the Oversleeve by pulling the Oversleeve and endoscope away from each other MOTUS GI PURE-VU SYSTEM INSTRUCTIONS FOR USE...
Following are the mechanical properties and size for Pure-Vu Workstation: Workstation Outer Dimensions W460 \ D480 (520 with pump head \ H340 m Workstation Weight 25kg Power Supply 100 - 240 VAC / 50 - 60 Hz MOTUS GI PURE-VU SYSTEM INSTRUCTIONS FOR USE...
Page 24