Table of Contents Product Manual - APPS 160 Revision Log Revision Level Date Description 07/25/2014 Convert to new layout and combine with manual PN 13428250 for 220V...
The APPS 160 Product Manual is designed to be used in is used to lower the saturation point of liquid nitrogen conjunction with APPS 160 provided by Chart If there are Liquid nitrogen saturated at a high pressure flows into the any questions regarding the operation of this unit, contact Chart’s Technical Service division at 1-800-400-4683.
Preface Product Manual - APPS 160 Acronyms / Abbreviations The following acronyms / abbreviations are used throughout this manual: APPS 160 Adjustable Pressure Phase Separator Decibel Gallons per minute Hertz MAWP Maximum Allowable Working Pressure Outer Dimension Part Number Pounds per Square Inch PSIG Pounds per Square Inch (Gauge)
Product Manual - APPS 160 Safety General Strict compliance with proper safety and handling practices One aspect of this possible hazard is the response of humans when exposed to an atmosphere containing only 8 to 12% is necessary when using an APPS 160 unit We recommend that all our customers re-emphasize safety and safe handling oxygen In this environment, unconsciousness can be practices to all their employees and customers While every...
Safety Product Manual - APPS 160 Warning! PRESSURIZED EQUIPMENT: Nitrogen and argon vapors in air dilute the concentration In normal operation the APPS of oxygen necessary to support or sustain life Inhalation 160 and the associated piping of high concentrations of these gases can cause anoxia, will be pressurized.
Product Manual - APPS 160 Installation / Operation Installation Start-Up • Remove the APPS 160 from the shipping crate Visually • Ensure APPS 160 is completely and properly installed inspect the unit to ensure there is no visible shipping • The APPS 160 unit should be purged with warm, dry damage nitrogen gas prior to introducing liquid to ensure that the •...
Installation / Operation Product Manual - APPS 160 Adjustments Liquid Nitrogen Level Operating Pressure • To change the operating pressure of the APPS 160, the set Caution! The liquid nitrogen level point of the vent regulator and pressure switch must be should only be changed by adjusted an experienced technician.
The Troubleshooting table is arranged in a “Trouble/Possible Cause/Remedy” format. Note that possible causes for specific symptoms are listed in descending order of significance. That is, check out the first cause listed before proceeding to the next. If you need further assistance please contact Chart’s Technical Service Team at 1-800-400-4683. Trouble...
11529090 11529090 Inlet check valve 11208931 11208931 11208931 11208931 See pictures on following pages for the identification of components. All parts are available from Chart at www chartparts com Vent Fill Valve Regulator Inlet Piping Vent Valve Liquid Level Liquid Level...
Page 15
Product Manual - APPS 160 Specifications Inlet Check Valve Inlet Strainer Inlet Connection Control Box Fill Valve Indicator Light On / Off Switch Vent Valve Indicator Light...
Page 16
Specifications Product Manual - APPS 160 Control Box hot - black wire connection Control Box neutral - white wire connection Control Box ground - green wire connection Liquid Level switch connections Liquid Level switch adjustment...