Chemetron 560 Series Operation And Maintenance Manual

Chemetron 560 Series Operation And Maintenance Manual

Medical gas outlet stations

Advertisement

OPERATION, MAINTENANCE & REPAIR MANUAL
MEDICAL GAS
OUTLET STATIONS
500/560 SERIES
Table of Contents

Introduction...................................................1

General Mechanical Principles..........................2
Materials Of Construction.................................3
Component Parts For Quick Connect
And DISS Outlet Stations...............................3
Inspection, Maintenance And Repair Procedures
(Quick Connect).............................................4
Inspection, Maintenance And Repair Procedures
(DISS).........................................................4
Troubleshooting Chart For Quick-Connect Outlet
Stations......................................................5
Operating Instructions For Chemetron )-Ring Service
Tool..........................................................7
O-Ring Service Replacement Tool Troubleshooting
Chart....................................................... 7
Recommendations....................................... 8
Additional Product Information........................8
____________________________________________________________________________________
Page 1
Andersen Medical Gas
12 Place Lafi fi tte
Madisonville, LA 70447
http://www.TheMedicalGas.com
1-866-288-3783
CHEMETRON MEDICAL GAS OUTLET STATIONS are designed
to provide immediate supply of medical gases and vacuum at such
points of use as hospital operating rooms, emergency rooms,
intensive care and patient rooms, as well as non-hospital based
facilities where a convenient source of medical gases and
additional services are needed.
The medical gases include, but are not limited to, oxygen, nitrus
oxide, compressed air, nitrogen, carbon dioxide, helium and
mixtures of such gases used for medical purposes.
Medical Gas pipeline equipment must be kept in dependable
operating condition at all times, and demand an effective
conscientiously followed program of periodic inspection, cleaning,
maintenance and repair.
For such a program, this manual is a concise, thorough, step-by-
step guide for easy use by responsible maintenance personnel in
hospitals or other installations where Chemetron medical gas
outlet stations are installed.
Read this manual thoroughly, and follow it closely at each
stage of cleaning, maintenance and repair procedures.
It is the responsibility of the user to maintain efficient performance
of Chemetron medical gas outlet stations. Allied Healthcare
Products, Inc. assumes no liability for accidents or other
irregularities arising from failure to observe recommended
maintenance procedures or follow cleaning, maintenance and
repair practices as presented in this manual.
Instructions are applicable to the following series of quick connect
stations; 500 and 560 DISS (Diameter Index Safety System) unit.
Allied Healthcare Products, Inc.
Form Number 64-00-2008
(S168-154-001)
APRIL 1993
Introduction

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 560 Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Chemetron 560 Series

  • Page 1 For such a program, this manual is a concise, thorough, step-by- Troubleshooting Chart For Quick-Connect Outlet step guide for easy use by responsible maintenance personnel in Stations……………………………………….…..5 hospitals or other installations where Chemetron medical gas outlet stations are installed. Operating Instructions For Chemetron )-Ring Service Tool……………………………………………..…..7 Read this manual thoroughly, and follow it closely at each stage of cleaning, maintenance and repair procedures.
  • Page 2 Though not illustrated, the Chemetron DISS* outlet stations (C). O-ring (B) seals the valve body within the outlet base are designed with most if the same features as the Chemetron housing - Figure 2. quick connect outlet stations with the exception of the primary valve body assembly (see component parts p.3).
  • Page 3 64-90-0034 13. Vacuum spring kit (6 each). ______________________________________________ 64-90-0652 14. Test plug kit (12 each) 64-90-0652 15. Back box mounting screw (24 each) Inspection Maintenance and Repair Procedures For Chemetron Quick-Connect Outlet Stations ____________________________________________________________________________________ Page 3 Allied Healthcare Products, Inc.
  • Page 4 Following is a step-by-step method of inspection, maintenance replace the Poppet O-ring (item 7), first lubricating it lightly with and repair for the Chemetron medical gas outlet station, series Chemetron O-ring lubricant (Part # 64-90-2111). 500, 400, 378, 349 and 348 quick-connect units. The purpose is to determine if parts are broken, missing, worn, distorted or 8.
  • Page 5 Time prescribes by local or national standards Replace them if necessary. Otherwise, replace the poppet O-ring and codes. Refer to NFPA 56F and/or CSA (item 23) lubricating it lightly with Chemetron O-ring lubricant. (Part #64-90-2111). Z305.1 standards for non-flammable medical gas systems.
  • Page 6 ____________________________________________________________________________________ Page 6 Allied Healthcare Products, Inc.
  • Page 7 Operating Instruction For The Chemetron O-Ring Service Tool The Chemetron outlet station O-ring service tool aids in snug against the face plate. CAUTION: DO NOT OVER- rapid removal and replacement of the valve body O-ring TIGHTEN OR DAMAGE TO THE LATCH MECHANISM (item 9) without disassembly or removal of the face plate.
  • Page 8 Series 238, 248 and 148 quick connect outlet stations Series 236, 246 and 146 diss outlet stations See modernization kits form #49-00-0001. The Chemetron Medical division urges the installation of modernization kits in all of the outlet station series indicated on previous page. The benefits of modernization are: Improved appearance.