Blood Glucose Meter Meter LCD Window Please read this User Guide before using your Blood Glucose Test Strip PRECICHEK Blood Glucose Meter. If you have any GL™ Control Solution questions or enquiries, please contact us or your Check Strip local distributor.
The PRECICHEK System is an electrochemical PRECICHEK Blood Glucose Meter, PRECICHEK Blood biosensor system that measures the amount of Glucose Test Strips (Model No. : ACH), PRECICHEK electric current produced and displays the result Check Strip, GL™ Control Solution, Lancing as a blood glucose level.
4. For ACH Strip, A sample with large amount 10. Do not use PRECICHEK system to test on of reducing substances such as triglycerides...
Page 5
critically ill patient. NOTE: PLEASE MAKE SURE THAT ALL PRODUCTS LISTED ON THE “CONTENTS” OF THE BOX ARE CONTAINED IN THE PACKAGE BEFORE USING THIS SYSTEM. IF YOU FIND ANY IMPERFECTION IN OUR PRODUCTS, PLEASE RETURN THE WHOLE SYSTEM TO THE PLACE OF PURCHASE.
STRIP SLOT: Holds a Blood Glucose Test Strip or Check meter serial number. Strip in place when you perform blood glucose test or perform check test. DISPLAY: The large, easy Blood Glucose Cloudia Cloudia Monitoring System For Self-testing Battery:AAA,6V 8901204043780370392...
Meter LCD Window Symbol:The test c a n b e a d j u s t e d a n d Alarm Symbol: stored into 3 types of test Appears when modes - Before Meal Test Symbol: Ap - alarm is set. ), After Meal Test( pears when data Strip Symbol: Appears when the strip is inserted and meter is...
Blood Glucose Test Strip automatically drawn to the test strip through capillary action. 2. PRECICHEK Blood Glucose Test Strips are in- Aperture: ap- Blood Reaction Black Electrode End: tended for in vitro diagnostic use with capil- ply the blood Zone: It must be Insert this end of test lary whole blood or GL™...
It can be used in two ways: carry test strips in their original vial. 1. To practice the test procedure 2. To make sure that the PRECICHEK Do not place in direct heat or sunlight. Blood Glucose Meter and Test 10.
Page 10
30ºC (86ºF). NOTE: REPEAT TEST IF THE RESULT FALLS OUTSIDE THE 4. Please carefully read the label before use. CONTROL RANGE. IF YOU CONTINUE TO GET THE RESULT FALLING OUTSIDE THE CONTROL RANGE, YOUR METER Perform Control Test AND STRIP MAY NOT BE WORKING PROPERLY. DO NOT 1.
Check Strip GVAH Check Strip RESULT PERSISTS, CONSULT YOUR LOCAL DISTRIBUTOR FOR HELP. The Check Strip can be used in CHECK How to delete memory 2 ways: 1. Insert check strip into strip slot with label side up. 1. To test only the function of 2.
2. Press Mem button to confirm and shift to the 12h/24h period. next setting. 5. The PRECICHEK has 5 alarms to keep your 3. Repeat above steps to set the month, day, hour blood glucose monitoring on time. Press Set and minute by Set and Mem buttons.
Prepare for Blood Sampling and press Mem button to enter next setting. Adjustable Lancing Device • When “OFF“ is chosen, press Mem button to enter next setting. Depth Selector Lancet 6. To configure the other 4 alarms, please repeat Depth Adjustment Cap step 5.
Performing Blood Test Set Lancing Device 1. Wash your hands in warm, 1. Remove the depth adjust- soap water. Rinse and dry ment cap. Insert a lancet into completely. Warm your fin- the lancet holder and push gers to increase blood flow. down until it is fully seated.
Page 15
6. Hold the prepared lancing 4. When the symbol flashes, device firmly against the side you are ready to perform a of your finger. Press the re- test. lease button. Gently massage 5. If the test is done within 2 your finger to obtain the hours of meal, press Set but- required blood volume.
Page 16
11. After removing the strip, 9. Before removing the strip from the result will be transmit- the meter, you can press Set ted to database through GSM button to set the test as BE- automatically. If OK appears FORE MEAL TEST , or AFTER on screen, the transmission MEAL TEST...
Understanding Your Test Result 3. Was the size of the blood sample sufficient? Normal Blood Glucose Range 4. Was the test strip vial cap tightly sealed? According to ADA current information, glucose values are listed for an healthy adult. 5. Was the test strip used before the expiration date? Time of Day Expected Range, Non-Diabetes...
Memory Recall The PRECICHEK Blood Glucose Meter can automati- 3. Press the Set button to view the cally store up to 999 results and auto-calculate average results. “19” represents your average results of 7-, 14-, 21-, 28- days on within the past 7 days there were both before and after meal.
The test result could be transmitted to database Replacing the battery for further management. The PRECICHEK Blood Glucose Meter comes with 1. When the meter is set “auto transmission” (refer 4 AAA battery (not yet be installed when new to section “Setting auto GSM Transmission”), purchasing).
The following disinfactant product have 3. Slide battery compartment door open. been shown to be safe for use with Cloudia 4. Remove the old batteries and insert 4 new AAA meter, but any disinfectant product with the batteries into the battery compartment. Be...
Page 21
Differences between Cleaning and Disinfection 3. Take a piece of premoistened wipe out According to the Merriam-Webster Medical of canister. Remember dictionary, the definitions of both terms are to recap the canister. below: 4. Wipe entire device and • Clean - to free from the dirt, or pollution lancing device until to free from disease or infectious agents.
Disinfect Procedures 6. Repeat step 4 for left 1. Prepare wipes and sides surface. meter. 2. Take a piece of pre 7. Repeat step 4 for right moistened wipe out of sides surface. canister. Remember to recap the canister. 3. Put the moistened wipe on a smooth surface.
Page 23
disinfection. 10. Make sure the meter 6. Please do not cloud up LCD by over-moist stays moisted for 1 min. the meter. 7. Please call us if a tear or a crack appeared 11. Discard the used wipes. on the meter. For Lancing Device 1.
Safety/AlertsandNotices/ucm224025.htm 4. Discard the used wipe according to the local 2. “CDC Clinical Reminder: Use of Fingerstick regulation. D e v i c e s o n M o r e t h a n O n e Pe r s o n Poses Risk for Transmitting Bloodborne Pathogens”...
Please follow all the instructions on the User This warranty is limited to replacement due to Guide while operating PRECICHEK Blood Glucose defects in parts or workmanship. HMD shall not be Meter. HMD BioMedical Inc. will not be responsible...
Blood glu- Review proper testing When any of the following messages appears, cose result procedure and perform there is a problem with your PRECICHEK Blood may be a quality check with Glucose Meter or perform wrong steps during a lower than control solution.
Page 27
N o r e - Maybe: You have to: Te s t s t r i p Perform the test with sponse s u n k n o w n new test strip. 1. B l o o d 1. Repeat test with suf- w h e n or ma y be sample is...
Page 28
Device can Please switch to man- not be syn- ual transmission and c h r o n o u s change location then w i t h l o c a l try again. date/time. Device can Please switch to man- n o t c o n - ual transmission and nect to the...
System Specification 1. Assay Method: Electrochemical biosensor 10. Result Setting: Before meal, After meal and Control solution test setting 2. Test Sample: Fresh Capillary Whole Blood 11. Average Display: 7, 14, 21, 28-days AC/PC 3. Sample Size: 0.5 μL average 4.
Performance Evaluations the PRECICHEK System; plasma samples from the Precision subjects were tested with YSI Model 2300 Glucose Tests were carried by trained technicians in the Analyzer as reference. The results are shown in laboratory setting. The venous whole blood from the following table.
Page 31
15.105(b): Batch code Please consult instructions for use NOTE: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. Temperature limitation/ In vitro diagnostic medical device Store at These limits are designed to provide reasonable protection against harmful interference in a residential installation.
Page 32
SAR is determined at the highest certified power level, the actual SAR level of the device while operating can be well below the maximum value. This is because the device is designed to operate at multiple power levels so as to use only the poser required to reach the network.
Need help?
Do you have a question about the Cloudia and is the answer not in the manual?
Questions and answers
How to remove depth adjustment cap
To remove the depth adjustment cap on a PreciChek Cloudia, pull it off from the lancing device.
This answer is automatically generated