CPM device after the same type of surgery. About the Manufacturer As a US based medical manufacturer, The Furniss Corporation builds Continuous Passive Motion (CPM) devices for the upper and lower extremities. The durable, user friendly devices are built to withstand the rigorous hospital and home care environment.
Handle the unit only when the unit is dry and hands are dry to prevent electrical shock. Unconscious patients or patients under heavy influence of medication must be constantly attended and monitored while the device is in use. The Furniss Corporation 800 273 5233 KneeCPM.com...
Page 4
Loosen the large knobs on the Foot Cradle Assembly and position in the upright position. Tighten knobs. Foot Cradle Assembly Thigh Length Adjustment Knobs (2) Control Box with Circuitry Analog Control Foot Cradle Pendant Adjustment Knobs (2) Bed Stabilizer Rods (2) Digital Control Pendant The Furniss Corporation 800 273 5233 KneeCPM.com...
Page 5
Furniss Softgoods are designed specifically for use with the Knee CPM Series. Use of any other brand of CPM Softgoods may adversely affect performance of the Knee CPM device and result in discomfort or injury to the patient. Reorder number for the Furniss Softgoods: SKIT 1828-1 The Furniss Corporation 800 273 5233 KneeCPM.com...
Remember: The goal is to align the knee axis of the patient with the knee pivot axis of the CPM Foot assembly adjustment knobs (2) Calf adjustment knobs (2) Knee pivot axis of unit Thigh adjustment knobs (2) Hip pivot assembly The Furniss Corporation 800 273 5233 KneeCPM.com...
Page 7
A strip of adhesive-backed loop Velcro may be applied to this side of the footplate for added stability of the foot boot softgood. The calf pad is omitted with patient set up in the pediatric conversion application. The Furniss Corporation 800 273 5233 KneeCPM.com...
Page 8
Power indicator light Pause option: Phoenix Models are set to pause at the end of flexion and extension cycles. This setting can be changed to pause the CPM device in extension only. The Furniss Corporation 800 273 5233 KneeCPM.com...
Page 9
- The meter is read by depressing the FLX and SPD buttons simultaneously - Clear meter by depressing the EXT, FLX, SPD and Pause buttons simultaneously. Note: The Start/Stop button must be pressed in order to lock in setting changes. The Furniss Corporation 800 273 5233 KneeCPM.com...
Page 10
The stop/start button functionality remains the same. Pause option: Phoenix Models are set to pause at the end of flexion and extension cycles. This setting can be changed to pause the CPM device in extension only. The Furniss Corporation 800 273 5233 KneeCPM.com...
A suggested Locktite product assists to secure the fasteners. Contact the Technical Service Department for recommendations on which type of Locktite to use on certain fasteners. Toll free 800.273.5233. The Furniss Corporation 800 273 5233 KneeCPM.com...
Page 12
Should the CPM unit require service, please do not hesitate to contact us. As always, we are here to answer any questions and help with any issues regarding the CPM units. The Furniss Corporation 800 273 5233...
Phillips Head Screws Allen Screws Plug Plug Tools Required: -9/64 Allen Wrench -Phillips Head Screwdriver -Compressed Air The Furniss Corporation 800 273 5233 KneeCPM.com Instructions: 1. Turn the power off to the unit. 2. Unplug the motor from the circuit board...
Page 14
10. Plug in the motor to the Circuit Board. 11. Replace the top cover. Rubber Bushing Allen Screws (4) Motor Mount Bracket Allen Screws (2) Rubber Coupler Power Motor J1 J2 Plug 1903 LM Knee Potentiometer To:______________________________________________ The Furniss Corporation 800 273 5233 KneeCPM.com Company:________________________________________ Fax:_____________________________________________...
Attach the green ground wire. Replacing the Power Supply Test the device for operation. Reinstall top cover. ON/OFF SWITCH WHITE WIRE TO CIRCUIT GROUND BOARD WIRE (GREEN) BLACK WIRE FROM POWER SUPPLY The Furniss Corporation 800 273 5233 KneeCPM.com Tools Required:...
Page 16
Run the unit for approximately one (1) hour after application to ensure proper alignment of track seals. View tight positioning of the track seals against the strut. Most adhesives require 24 hours for a complete curing process. The Furniss Corporation 800 273 5233 KneeCPM.com...
FROM BOARD POWER SUPPLY BLACK WIRES Tools Required: Pliers The Furniss Corporation 800 273 5233 KneeCPM.com Caution: To avoid electric shock unplug the unit before beginning this procedure. 1. Turn the Power Switch OFF and unplug unit from wall outlet.
Page 18
Using the non-magnetic screwdriver, remove the 4 screws holding the Circuit Board in place. Replace with the new board and secure with the 4 screws. Plug in the motor, power, pendant and potentiometer connectors. Follow the calibration procedure. Replace the top cover. The Furniss Corporation 800 273 5233 KneeCPM.com...
Page 19
Reinstall 2 position jumper on its original settings pins 5 and 6 or pins 6 and 7 2 position jumper on pins 5 and 6 (Pause in Extension only) 2 position jumper on pins 6 and 7 (Pause in Extension and Flexion) The Furniss Corporation 800 273 5233 KneeCPM.com...
Page 20
Depress start/stop switch on the hand control and allow unit to travel through full range of motion. If the unit travels the full range of motion without binding then the unit has been calibrated successfully. The Furniss Corporation 800 273 5233...
Page 21
Depress start/stop switch on the hand control and allow unit to travel through full range of motion. If the unit travels the full range of motion without binding then the unit has been calibrated successfully. The Furniss Corporation 800 273 5233...
If the Unit does not conform to this warranty, The Furniss Corporation will at its option repair or replace the defective Unit.. The remedy of repair or replacement is purchasers’ sole and exclusive remedy and will satisfy all of The Furniss Corporation’s liabilities, whether based on contract, tort, product liability, strict liability, or otherwise.