INSTRUCTIONS FOR
USE
ECONOX
USA
600 Mogadore Road
Kent, OH 44240
T: +1 330 678-3889
www.econox.us
info@econox.us
The details in this document are provided for information purposes only. This manual may not, in any
event, be reproduced, dissociated or distributed to third parties without the consent of ECONOX SA.
1/7
X:\Fiches tech. - Normalisation\ECONOX\Normalisation\Mode d'emploi\PDF\sonde

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CarboProbe CP and is the answer not in the manual?

Questions and answers

Summary of Contents for Econox CarboProbe CP

  • Page 1 T: +1 330 678-3889 www.econox.us info@econox.us The details in this document are provided for information purposes only. This manual may not, in any event, be reproduced, dissociated or distributed to third parties without the consent of ECONOX SA. X:\Fiches tech. - Normalisation\ECONOX\Normalisation\Mode d'emploi\PDF\sonde...
  • Page 2: Table Of Contents

    The benefits of an oxygen probe ..........5 How to install the oxygen probe ..........5 Understanding an oxygen probe ..........6 2.Checking temperature using the CarboProbe CP plus ......6 3.Repairing the sensor ............... 7 X:\Fiches tech. - Normalisation\ECONOX\Normalisation\Mode d'emploi\PDF\sonde...
  • Page 3: Operational Principles

    1. Operational principles The purpose of ECONOX CarboProbe oxygen sensors is to measure and regulate atmospheres. ECONOX uses two different types of electrolyte made of ZrO (zirconium oxide) for its oxygen sensors: 1. A ball made of ZrO , an ECONOX-patented system, which may only be obtained from ECONOX.
  • Page 4: Introduction

    ECONOX supplies 2 different kinds of CP probes For the measure of oxygen concentration only For measure of oxygen concentration AND temperature (includes an R-type thermocouple for temperature up to 1700°C)
  • Page 5: How Does The Air Supply Affect The Firing

    When fitting or removing a probe into or out of a hot furnace, move the probe slowly to prevent thermal shock. Once the kiln is over 700 °C and the reference air is available, the probe is ready to use. X:\Fiches tech. - Normalisation\ECONOX\Normalisation\Mode d'emploi\PDF\sonde...
  • Page 6: Understanding An Oxygen Probe

    There is no sharp distinction between oxidizing, neutral and reducing. There is a smooth variation from one to the next, so the above figures are only guidelines. Checking temperature using the CarboProbe CP plus When the probe is ready to use, connect the probe to the meter. Select the "temperature"...
  • Page 7: Repairing The Sensor

    When sending a sensor for repair, pack it carefully in its original packaging, mark it "Fragile Instrument", and return it to: ECONOX USA 600 Mogadore Road Kent, Ohio 44240 T: +1 330-678-3889 www.econox.us info@econox.us X:\Fiches tech. - Normalisation\ECONOX\Normalisation\Mode d'emploi\PDF\sonde...

This manual is also suitable for:

Carboprobe cp plus