User Manual AWR Nebulizer MN-300/2

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MN-300/2 and is the answer not in the manual?

Questions and answers

Summary of Contents for AWR MN-300/2

  • Page 1 User Manual AWR Nebulizer MN-300/2...
  • Page 3 Tel.: +49 33708 355020 Online: http://www.awr-inhalation.de Document version MN-300/2-EN-V3 NOTE: Please carefully read the user manual before first use of the AWR Nebulizer MN- 300/2. Please store the user manual in a safe place. (Version MN-300/2-EN-V3) Page 3 of 32...
  • Page 4: Table Of Contents

    User Manual AWR Nebulizer MN-300/2 Table of Contents Safety Information ......................5 Scope of Delivery, Components of the AWR Nebulizer MN-300/2 ......8 LED Displays and Graphical Symbols ................10 Intended Use ......................12 Cleaning ........................17 Power Supply ......................18 Environmental Conditions ..................
  • Page 5: Safety Information

    2. The device is unsuitable for use in proximity to combustible gases. 3. The device must not come into contact with heated surfaces. 4. The AWR Nebulizer MN-300/2 is not an officially certified medicine product and therefore not approved for nebulizing medications. Usage and choice of possible medications for nebulizing can only be determined by the treating physician.
  • Page 6 2. Make sure before every use that the medication has been filled into the nebulizer unit. 3. Please make sure to always hold the AWR Nebulizer MN-300/2 during nebulizing in a way that the liquid always directly touches the silver membrane of the nebulizer unit.
  • Page 7 (Elektro- and Elektronikgesetz – ElektroG). For this reason, the electronic components of the AWR Nebulizer MN-300/2 (control unit and power adapter) must not be disposed as domestic waste. Disposal of these electronic components has to meet local disposal policies.
  • Page 8: Scope Of Delivery, Components Of The Awr Nebulizer Mn-300/2

    AWR Nebulizer MN-300/2 for the first time. NOTE Please inspect the delivery for completeness before first operation of the AWR Nebulizer MN-300/2. Also inspect the delivered components for visual defects such as contamination, stress cracks or damage. The AWR Nebulizer MN-300/2 or its components must not be used if they display visual defects or damage.
  • Page 9 User Manual AWR Nebulizer MN-300/2 Article # Description Image 201 102 05 Power adapter AWR Nebulizer MN-300/2 201 102 02 Optionally available: USB charging cable AWR Nebulizer MN-300/2 201 102 06 Optionally available: Nebulizer unit AWR Nebulizer MN-300/2 MMD: 6.0 µm...
  • Page 10: Led Displays And Graphical Symbols

    User Manual AWR Nebulizer MN-300/2 3. LED Displays and Graphical Symbols This chapter lists the graphical symbols and LED displays you find on your AWR Nebulizer MN-300/2, the accessories, this user manual and the packaging. Graphical symbols on the packaging:...
  • Page 11 User Manual AWR Nebulizer MN-300/2 Graphical symbols on the control unit: Image Definition On/Off button Start Start/Stop button Stop Distributing company: Shows distributing company of the product Serial number: Shows the serial number designated by the distributing company in order to make identification of a single...
  • Page 12: Intended Use

    User Manual AWR Nebulizer MN-300/2 4. Intended Use The AWR Nebulizer MN-300/2 is a nebulizer designed to introduce exact doses of liquids in the form of aerosol into the airways of the lungs. WARNING The AWR Nebulizer MN-300/2 is not an officially certified medicine product and therefore not approved for nebulizing medications.
  • Page 13 User Manual AWR Nebulizer MN-300/2 Step 2: Connecting the nebulizer unit to the control unit: Connect the filled nebulizer unit to the control unit by connecting the female USB connector of the control unit with the male USB connector of the nebulizer unit.
  • Page 14 NOTE The standard setting for the AWR Nebulizer MN-300/2 is the manual mode. When you switch on the AWR Nebulizer MN-300/2 and it’s ready for operation, the AWR Nebulizer MN-300/2 will be in manual mode. You can perform the switch from manual to continuous mode within a timeframe of 60 seconds after startup or last manual nebulizing.
  • Page 15 ATTENTION Please make sure to always hold the AWR Nebulizer MN-300/2 during nebulizing in a way that the liquid always directly touches the silver membrane of the nebulizer unit. If an air bubble forms behind the metal membrane or if no liquid touches the membrane, the control unit will switch off automatically.
  • Page 16 User Manual AWR Nebulizer MN-300/2 Step 6: Switching off the control unit: To abort or pause nebulizing, the AWR Nebulizer MN-300/2 has to be switched off. To do this, press the on/off button of the control unit for a prolonged period (>...
  • Page 17: Cleaning

    NOTE The cleaning process for the nebulizer unit described here refers to both the AWR Nebulizer MN-300/2 nebulizer unit with an MMD of 4.0 µm and the optionally available AWR Nebulizer MN-300/2 nebulizer unit with an MMD of 6.0 µm.
  • Page 18: Power Supply

    User Manual AWR Nebulizer MN-300/2 6. Power Supply The control unit of the AWR Nebulizer MN-300/2 is powered by the internal rechargeable battery (lithium-ion battery). The battery status indicator of the control unit shows you the charging status of the...
  • Page 19 NOTE Depending on the initial state, the charging time for the internal, rechargeable battery (lithium-ion battery) of the control unit for the AWR Nebulizer MN-300/2 is about 2 hours for 80% charging capacity and about 3 hours for 100% charging capacity.
  • Page 20: Environmental Conditions

    7. Environmental Conditions Described below are the environmental conditions which have to be maintained to ensure proper functioning of the AWR Nebulizer MN-300/2 in its intended use. Operating conditions: Reliable operation of the device is guaranteed under the following environmental...
  • Page 21: Disposal

    (Elektro- and Elektronikgesetz – ElektroG). For this reason, the electronic components of the AWR Nebulizer MN-300/2 (control unit and power adaptor) must not be disposed as domestic waste. Disposal of these electronic components has to meet local disposal policies.
  • Page 22: Consumables

    User Manual AWR Nebulizer MN-300/2 9. Consumables Die nebulizer units with an MMD of 4.0 and 6.0 µm as well as the mouth piece of the AWR Nebulizer MN-300/2 are consumables. The consumables can be supplied by the distributing company.
  • Page 23: Maintenance, Service And Order Information

    User Manual AWR Nebulizer MN-300/2 10. Maintenance, Service and Order information If the AWR Nebulizer MN-300/2 has been damaged or stops working properly, it should be examined and a technical safety inspection should be performed. WARNING Do not open the case. Any repairs or maintenance must only be performed by the distributing company.
  • Page 24: Technical Specifications

    Power adapter AWR Nebulizer MN-300/2: Size 74 x 44 x 73 mm Cable length 1.8 m Weight 165 g Nebulizer unit MMD 4.0 µm AWR Nebulizer MN-300/2: Size 57 x 56 x 30 mm Maximum filling capacity 8 ml Weight 20 g Nebulizer unit MMD 6.0 µm AWR Nebulizer MN-300/2:...
  • Page 25: Performance Characteristics And Aerosol Spectrum

    12. Performance Characteristics and Aerosol Spectrum Aerosol delivery as well as particle size have been evaluated and determined for the AWR Nebulizer MN-300/2 in accordance with DIN EN 13544-1, appendix CC. Performance characteristics and aerosol spectrum for the AWR Nebulizer MN-300/2 when using the nebulizer unit with an MMD of 4.0 µm:...
  • Page 26: Information On Electromagnetic Compatibility

    IEC 61000-3-3 WARNING The AWR Nebulizer MN-300/2 must not be used in immediate proximity to other devices. If this is necessary, the device should be monitored to ensure proper function in this operational setup. Page 26 of 32...
  • Page 27 User Manual AWR Nebulizer MN-300/2 Guidance and manufacturer's declaration – electromagnetic immunity: The nebulizer is intended for use in the ELECTROMAGNETIC ENVIRONMENT specified below. The customer or user should make sure that it is being operated in such an environment.
  • Page 28 User Manual AWR Nebulizer MN-300/2 Portable and mobile RF communications equipment should be used no closer to the nebulizer including its cable than the recommended safety distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance:...
  • Page 29 User Manual AWR Nebulizer MN-300/2 Recommended safety distances between portable and mobile HF telecommunications devices and the nebulizer: Der nebulizer is intended for operation in an electromagnetic environment in which high-frequency interference variables are controlled. The customer or user of the nebulizer can help avoid electromagnetic interference by maintaining the following minimum safety distances between portable and mobile high-frequency telecommunication devices (transmitters) and the nebulizer –...
  • Page 30: Ec Declaration Of Conformity

    73/23/EWG (electrical equipment designed for use within certain voltage limits. 15. Warranty We grant you a 24-months warranty starting on the date of the purchase for your AWR Nebulizer MN-300/2. This warranty applies to the control unit of the AWR Nebulizer MN-300/2 and power adapter of the AWR Nebulizer MN-300/2.
  • Page 32 Are you happy with the AWR Nebulizer MN-300/2? Tell your friends about it (we’d be delighted)! If not: Tell us and we will try to help.