Download Print this page
Gast 0211 Series Operation & Maintenance Manual

Gast 0211 Series Operation & Maintenance Manual

Lubricated laboratory vacuum pumps & compressors

Advertisement

Quick Links

70-180 G325PL (Rev. C)
0211 SERIES LUBRICATED LABORATORY
VACUUM PUMPS & COMPRESSORS
OPERATION & MAINTENANCE MANUAL
CONTENTS:
General Information and Installation ....................................................................................................... 2
Operation and Maintenance ..................................................................................................................... 3
Exploded View and Parts Ordering Information ..................................................................................... 4
Shutdown Procedures, Service Kit Installation, and Troubleshooting Guide ......................................... 5
Warranty and Authorized Service Facilities ............................................................................................ 6
Visit us at our website
www .gastmfg.com
KEEP THIS DOCUMENT FOR FUTURE REFERENCE

Advertisement

loading

Summary of Contents for Gast 0211 Series

  • Page 1 70-180 G325PL (Rev. C) 0211 SERIES LUBRICATED LABORATORY VACUUM PUMPS & COMPRESSORS OPERATION & MAINTENANCE MANUAL CONTENTS: General Information and Installation ....................... 2 Operation and Maintenance ........................3 Exploded View and Parts Ordering Information ..................4 Shutdown Procedures, Service Kit Installation, and Troubleshooting Guide ......... 5 Warranty and Authorized Service Facilities ....................
  • Page 2: Installation

    Proper lubrication is necessary for normal service life. Vanes occur if hazard is ignored. ride on a film of oil that fills clearances and provides a seal. Gast lubricating oil is available in convenient quart containers CAUTION Minor injury or property damage can from the factory or Gast representatives (order part #AD220).
  • Page 3: Maintenance

    2. Remove all plumbing and accessories connected to This unit requires lubrication. Failure to maintain inlet and exhaust ports. proper level of oil (10 wt. H.D. Gast Part # AD220) in 3. Remove filter/muffler assembly. siphon oiler will cause unit to malfunction.
  • Page 4 EXPLODED VIEW & PARTS ORDERING INFORMATION t l i & & l i a • t l i • • t l i • • • c t i & • . t i l a i...
  • Page 5: Shutdown Procedures

    2. Remove filter jar and filter assembly. This unit requires lubrication. Failure to maintain proper 3. Remove six (6) Phillips-head screws from end plate. oil level (10 wt. H.D./Gast part #AD220) in reservoir will 4. Remove end plate. cause unit to malfunction.
  • Page 6: Your Warranty

    NO WARRANTIES WHICH EXTEND BEYOND THE DESCRIPTIONS SET FORTH IN THIS WARRANTY, notwithstanding any knowledge of Gast regarding the use or uses intended to be made of goods, proposed changes or additions to goods, or any assistance or suggestions that may have been made by Gast personnel.

This manual is also suitable for: