Operation Set-up Tubing assembly-Internal irrigation Tubing assembly–External irrigation Tubing assembly-combined Operation Starting the MD 10 Keyboard Setting the revolutions- Setting the output of the irrigation pump Operation with vario-pedal Setting the torque limiter AL (Automatic Limiter) Setting the torque limiter AS (Automatic Stopper)
MD 10 NOUVAG AG 1 Description 1.1 Operations and functions The MD10 is used in Implantology and microsurgery. The unit is designed for drilling, screwing and sawing. A built-in irrigation system minimizes the heat generation of the rotating instruments in order to reduce infections that can cause tissue damage.
Repairs are to be carried out by authorized NOUVAG service technicians only! • Nouvag AG cannot be held liable for any malfunction of the MD 10, or performance failure and/or its designed or desired utility, nor can Nouvag AG be held liable for any injuries to persons or animals, in any case when the MD 10 is miss-used or not operated, applied or maintained in strict accordance with the user/owner instructions set out in the operating manual.
NOUVAG AG MD 10 Insure that „pressure“ tube runs to irrigation needle, and „suction“ tube runs to solution bottle! 10) Insert tube needle and hang cooling flask on hanger rod. 11) Open tube clamp. 12) Open air valve of the drip chamber.
NOUVAG AG MD 10 Press keys “UP” or “DOWN” to select the desired speed. With the keys “SAW“ (for Nouvag-saws) and “MUCO“ (for Nouvag-Mucotome) the corresponding fixed speed and torque values can be adjusted. The reduction or multiplication on the keyboard panel has to be according to the reduction or multiplication of the contra angle, otherwise the device will indicate an incorrect revolution.
MD 10 NOUVAG AG 5.6 Setting the torque limiter AL (Automatic Limiter) The automatic limiter (AL) limits the torque level applied to the instrument. The AL-modus is used for drilling into the bone, in order to allow the electronicmotor to develop enough power, so the instrument can have a good drive in hard bones, as well.
NOUVAG AG MD 10 6 Disinfection, Cleaning and Sterilization Please pay attention to the following important points for the maintenance of the material: • Clean, disinfect and sterilize the device and components after each use. • Do not use dissolving agents for cleaning.
MD 10 NOUVAG AG 7 Troubleshooting Refer:operating Manual: Problem: Cause: Solution: Device does not Turn main switch System Assembly Control unit not turned on “ON/OFF“ON operate at all Check mains voltage Technical Data Wrong operating voltage output Connect mains to control...