Summary of Contents for Bevco IECAX480V125A

  • Page 1 Technical Publications Installation, Operation, and Service Manual SIEMENS HEALTHCARE INTEGRATED ELECTRICAL CABINET ARTIS Q / Q.ZEN / ONE ARTIS ZEEGO / ZEEGO.ZEN ARTIS ZEE MULTIPURPOSE SINGLE PLANE IECAX480V125A Main Disconnect Panel...
  • Page 2: Table Of Contents

    • (3) sets of Emergency Power Off pushbuttons with 2NC contacts on each EPO • Drawings and Electrical Schematics Copyright 2016: Bevco Engineering Company, Inc., W222N5739 Miller Way, Sussex, WI 53089 USA IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems...
  • Page 3: Description Of Operation

    Contacts or 24 Volts AC or DC. Note the maximum connected load to the X-Ray On lights is 20 Amps. Copyright 2016: Bevco Engineering Company, Inc., W222N5739 Miller Way, Sussex, WI 53089 USA IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems...
  • Page 4: System Configuration

    POWER PANEL CIRCUITS 120/277V EPO1 EPO3 EPO2 ROOM LIGHTS GENERATOR 1 20A MAX 100A MAX Copyright 2016: Bevco Engineering Company, Inc., W222N5739 Miller Way, Sussex, WI 53089 USA IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 4 of 14...
  • Page 5: Specifications

    CONTACT BEVCO ENGINEERING, SUSSEX WI (262) 820-2400 FOR ISSUES FOLLOWING INSTALLATION, CONTACT SIEMENS UPTIME CENTER, (800) 888-7436 Copyright 2016: Bevco Engineering Company, Inc., W222N5739 Miller Way, Sussex, WI 53089 USA IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 5 of 14...
  • Page 6: Panel Dimensions

    18.37 inches 467 mm Mounting Hole Height 28.5 inches 740 mm Weight 67 pounds 30.4 kg Copyright 2016: Bevco Engineering Company, Inc., W222N5739 Miller Way, Sussex, WI 53089 USA IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 6 of 14...
  • Page 7: Center Of Gravity

    INTEGRATED ELECTRICAL CABINET Installation, Operation, and Service Manual • CENTER OF GRAVITY Copyright 2016: Bevco Engineering Company, Inc., W222N5739 Miller Way, Sussex, WI 53089 USA IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 7 of 14...
  • Page 8: Mounting Hole Dimensions

    INTEGRATED ELECTRICAL CABINET Installation, Operation, and Service Manual • MOUNTING HOLE DIMENSIONS Copyright 2016: Bevco Engineering Company, Inc., W222N5739 Miller Way, Sussex, WI 53089 USA IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 8 of 14...
  • Page 9: Installation

    12. Connect the Door Switch Connections of the Siemens System Cabinet SC1 to the terminals 3 and 4 of X07. See Door Switch Wiring Diagram. Copyright 2016: Bevco Engineering Company, Inc., W222N5739 Miller Way, Sussex, WI 53089 USA IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems...
  • Page 10 INSTALLATION (continued) TERMINAL LAYOUT (Located in the bottom left of the Disconnect Panel) EPO WIRING Copyright 2016: Bevco Engineering Company, Inc., W222N5739 Miller Way, Sussex, WI 53089 USA IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 10 of 14...
  • Page 11 INTEGRATED ELECTRICAL CABINET Installation, Operation, and Service Manual LIGHTING WIRING DOOR SWITCH WIRING Copyright 2016: Bevco Engineering Company, Inc., W222N5739 Miller Way, Sussex, WI 53089 USA IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 11 of 14...
  • Page 12: Testing And Troubleshooting

    • Verify Continuity between terminals 3 and 4 of terminal bloc X05 if dry contact provided by Siemens System Cabinet. • Verify K4 is on. Copyright 2016: Bevco Engineering Company, Inc., W222N5739 Miller Way, Sussex, WI 53089 USA IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 12 of 14...
  • Page 13 K10 Safety Relay configurations The Safety Relay DIP switches are set as follows from the factory. Copyright 2016: Bevco Engineering Company, Inc., W222N5739 Miller Way, Sussex, WI 53089 USA IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 13 of 14...
  • Page 14: Maintenance

    Update the manual based on schematic 03/30/16 IEC Redesign 07/10/18 Update of system config data Copyright 2016: Bevco Engineering Company, Inc., W222N5739 Miller Way, Sussex, WI 53089 USA IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 14 of 14...