R I W I T N E S S ™ Embryology Heated Plate +44 (0) 1737 243869 | customerservice@origio.com | origio.com only Research Instruments Ltd, Bickland Industrial Park, Falmouth, Cornwall TR11 4TA, UK 0120 Document 6-70-807IM | DRF 3728 | Issue 4 | 13 December 2017...
Contents Research Instruments Ltd CONTENTS SECTION 1 - PREFACE SECTION 2 - INTRODUCTION TO RI WITNESS Indications for Use for RI Witness Embryology Heated Plate Contraindications Applicable Part Numbers Related Documents Compatibility Installation SECTION 3 - SAFETY WARNINGS Safety/Information Symbols...
Page 4
Research Instruments Ltd Contents RFID Tag Read Range Checklist Temperature Calibration Checklist SECTION 6 - REPAIRS AND RETURNS Reuse Statement RI Repairs System RI Returns System Product Disposal (European Union) Contact Details Obligation to Inform...
Instruments or an appointed representative before attempting to use this equipment. In no event does Research Instruments Ltd (RI) assume the liability for any technical or editorial errors of commission, or omission; nor is RI liable for direct, indirect, incidental, or consequential damages arising out of the use or inability to use this manual.
RI authorised personnel. Incorrect installation could result in overall poor performance. All relevant sections of this manual should be read and understood fully before any use of RI Witness takes place. If the operator is unsure of any of the information contained in this manual, they should contact...
The system should be operated by qualified and trained personnel only. DO NOT disassemble or modify any part of the RI Witness Embryology Heated plate, or substitute any component for any other. Doing so may result in damage to samples. This voids the warranty and/or service contract.
Section 3 Safety Warnings Research Instruments Ltd Safety/Information Symbols Symbol Meaning Do not dispose of product with normal waste. In accordance with Annex II of the European Medical Device Directive 93/42/EEC, as amended by Directive 2007/47/EC under the supervision of notified body No.0120, SGS, UK Ltd.
RFID Reader Environment An RI Witness system uses Radio Frequency Identification (RFID) readers to monitor a work area. Readers detect RFID tagged containers that are placed in the work area. The performance of RFID tag detection may be compromised by proximity of metal objects or electrical equipment Do not place metal objects near reader.
Guidance and Manufacturer’s Declaration (IEC 60601-1-2) — Electromagnetic Emissions RI Witness is intended for use in the electromagnetic environment specified below. The customer or the user of RI Witness should ensure that it is used in such an environment. Emissions test...
40 % UT 40 % UT hospital environment. (60 % dip in UT) for 5 (60 % dip in UT) for 5 If the user of RI Witness IEC 61000-4-11 cycles cycles requires continued 70 % UT...
Page 12
Section 3 Safety Warnings Research Instruments Ltd Guidance and Manufacturer’s Declaration — Electromagnetic Immunity IMMUNITY Test IEC 60601 Compliance Electromagnetic environment - guidance Test level level Portable RF communications equipment Conducted RF IEC 3 Vrms 3 Vrms (including peripherals such as antenna...
Check for worn or damaged cables 2. Install software and perform database set up. Refer to 6-70-121IT RI Witness IT Requirements. 3. Position the device in the desired location and adjust the movable aperture as shown below for flush fitted configurations.
Page 14
Section 4 Installation Checklist Research Instruments Ltd Legs are also available enabling the Embryology Heated Plate to be supported when placed on top of a stereo zoom microscope light base. Figure 4-3 ii. Positioning of Flush Embryology Heated Plate (6-70-807-A/B) The Flush Embryology Heated Plate should be mounted in a flow hood tray or other work surface prior to being installed.
Page 15
USB socket on the PC or Tablet. Verify the connection to the software as described in ‘Connecting to the Software’ section in 6-70-807UM RI Witness Embrology Heated Plate User Manual. 6. Go into the RFID Tuning Screen and check that all connected antennas have a green tick next to them.
Service Checklist and Procedures Research Instruments Ltd SECTION 5 – SERVICE CHECKS AND PROCEDURES The following service activities should be performed to service an RI Witness work area. Component and Connections Damage Checklists 1. Check data and power connections for security and damage.
Verify that the temperature indicated by built in user interface and software matches a reading taken by thermometer for all five heating channels. Refer to the ‘Temperature Calibration’ section of section in 6-70-807UM RI Witness Embrology Heated Plate User Manual.
7 years continual use, after which time we recommend you consider its replacement. Should you notice impaired performance and/or any issues where safety is compromised, or have any other concerns during the use of RI Witness, seek the advice of RI or their authorised representative promptly. RI Repairs System In the event that you have a problem with a RI instrument, please follow the procedure below to ensure prompt attention.