Download Print this page

Care & Maintenance - BCF Solutions Easi-Scan:Go Instruction Manual

Advertisement

8 Care & maintenance
8.1 Easi-Scan: Go & battery
Please note the following important points:
The unit is not sealed against jets of water. Water will penetrate your scanner if it is hosed
down or rinsed under a fast flowing tap.
The unit is NOT corrosion proof and thus it should not be left wet.Take care to detach the
external battery and ensure interconnect areas are dry also.
The battery charger and charging carry case are not waterproof – avoid getting them wet.
8.2 Easi-Scan: Go probe
While every attempt has been made to make the probe as rugged as possible the crystal array
remains vulnerable and should be protected from knocks. The probe has a protective sheath
covering the length of cable subject to wear and flexing. Any damage to this sheath or the rest
of the cable should be repaired before further use to prevent moisture ingress that will rapidly
damage the probe. The probe should be checked regularly for such damage.
Gels
Use only gels recommended by BCF Technology Ltd. Do not store the probe in gel.
Cleaning
The probe may be cleaned by removing all residue and disinfecting using an IPA wipe.
Storage
Clean and dry the probe head prior to storage in dry location. Do not store the transducer in gels
or cleaning/disinfecting solutions. This will lead to premature deterioration of the probe.
Do not expose the transducer to materials containing the following agents:
Acetone
Methanol
Mineral oil
Iodine
Freon
Industrial cleaners
Materials containing perfumes (lotions, gels, etc.)
9. Specifications
Warning:
Never sterilize the transducer with autoclave, ultraviolet, gamma
radiation, gas, steam or heat sterilization techniques. Severe
damage will result.