Table of Contents Safety, Storage and Handling Data General Installation The following symbols apply to this product and are used 3 Description in these instructions and on the product in question. The meanings of these symbols are as specified below: - 3 Operation 3 Safety information 3 Installation procedure - Wall mounted, surface and concealed...
Description and Operation Multi-gas terminal unit installations when vertically mounted shall be spaced at 135 ± 2.5mm between centres. BEACONMEDÆS Anaesthetic Gas Scavenging (AGS) terminal units are designed to fully comply with BS6834:1987 and The centre of a vertically mounted terminal unit shall be not accept probes to the same standard, their installation fully less than 200mm from any possible obstruction on either satisfies the United Kingdom Health Technical...
Page 4
nuts (2 - off) and fully tighten. layout and mark copper stub pipe for length required. Re- move terminal unit assembly. Stub pipe. Braze to downdrop pipe. Copper stub pipe. Cut to length. Using the approved jointing technique as detailed in the contract specification, braze stub pipe to downdrop pipe.