CryoSecure™ Compressor Power Controller Hardware Manual ® Printed in U.S.A. ORTEC Part No. 774890 1209 Manual Revision D...
Page 2
If it becomes necessary to return this instrument for repair, it is essential that Customer Services be contacted in advance of its return so that a Return Authorization Number can be assigned to the unit. Also, ORTEC must be informed, either in writing, by telephone [(865) 482-4411] or by facsimile transmission [(865) 483-2133], of the nature of the fault of the instrument being returned and of the model, serial, and revision ("Rev"...
SAFETY INSTRUCTIONS AND SYMBOLS This manual contains up to three levels of safety instructions that must be observed in order to avoid personal injury and/or damage to equipment or other property. These are: DANGER Indicates a hazard that could result in death or serious bodily harm if the safety instruction is not observed.
BIAS OFF indicates the detector bias voltage has cooling is restarted. . The CryoSecure is intended to be been interrupted or shutdown. A front-panel, two- used with a cryogenic cooling system such as the position rocker switch turns power on and off, and a ORTEC EC-III™.
10.2 cm. (4.0 in.) high. Aluminum enclosure. BIAS OFF TIME Internal PWB-mounted rotary switch allows the user to set amount of time the CryoSecure waits after restarting the compressor, before allowing the detector bias voltage to be applied.
3.3. INSTALLATION IN A “WARM” SYSTEM The battery is connected to the control logic by a self- If the CryoSecure is installed in a system with the holding relay which is kept in the OFF state during detector at room temperature, it is inconvenient to have storage and shipping (see Section 5.2 for storage and...
(BIAS-OFF) TIME The cooling, or bias-off, time is the time the CryoSecure If the mains supply fails during the cooling period, the waits after restarting the compressor before allowing the CryoSecure will return to the start of the warming state.
3 hours will actually elapse in 3 minutes. This is useful for checking system operation. With the CryoSecure disconnected from the mains supply, remove the top cover and locate the sealed, WARNING Remember to reinstall W3 after rechargeable, lead-acid battery (to the left of the unit as testing.