EI-1000 & 2000: Setup, Calibration and Troubleshooting. EI-1000 & EI-2000 DIGITAL WEIGH INDICATOR/TRANSMITTER EAGLE MICROSYSTEMS, INC. 366 CIRCLE OF PROGRESS POTTSTOWN, PA 19464 Page 1 of 19...
Page 2
EI-1000 & 2000: Setup, Calibration and Troubleshooting. Page intentionally left blank… ( Move along, nothing to see here! ) Page 2 of 19...
EI-1000 & 2000: Setup, Calibration and Troubleshooting. Table of Contents: Page: Installation and Wiring. Calibration. Operation in automatic modes. Scale Wiring color codes. Troubleshooting. Changing Configuration Parameters, turning off auto-zeroing. 4-20 mA Outputs - Parameter Setup, Testing, and Calibration. 4-20 mA Troubleshooting.
3. Wire scale base. Use TB4 on EI1000 units. See Scale Wiring section for details. The EI-1000 has a shorting jumper on TB5 (+S,-S,-E). Do not remove this jumper. 4. Check S2 and set for the proper power input voltage 230 or 115 V.A.C. S2 is directly above the power terminal TB1.
6. On the EI-2000, repeat the calibration on the other side. 7. Press the right arrow twice on the EI-1000 to exit the menus and return to the gross weight reading. On the EI-2000 use the left SELECT.
The reading will jump to TARE after several seconds. Use the UP/DOWN arrows to set the tare weight. Press SELECT when done. Please copy this page and post it by the EI-1000 / 2000 for quick reference. Page 6 of 19...
Code C works with the LP4310, scales with a hinge and cell mounted inside of the deck, and our HC1000 hydraulic conversion kit. If your scale or load cell was not obtained from Eagle Microsystems, try code C or A first. You may want to contact the manufacturer for the color code.
If it powers up now there is a short in the scale wiring. • The jumper on J3 should be below the 1 (right side) for an EI-1000 and below the 2 (left side) for the EI-2000.
Page 9
• On the EI-1000 with a single scale the scale should be wired to TB4 (seven position terminal). • The EI-1000 comes with a jumper shorting the +S,-S, and –E terminals on TB5.
Page 10
Open the switch and you will be able to select the tare. Note: On old style EI-1000’s hold down SELECT until “nOnE” flashes. Press UP to get to “CFG” and press SELECT to get to “P001”. Hold down “SELECT” until “DONE”...
Page 11
• Enter the configuration as shown above. • Use the UP arrow to get to P113. • EI-1000 - Press SELECT and use the down arrow to change the value to OFF. Press SELECT again. • EI-2000 - Use the right DOWN arrow to change the value to OFF.
Page 12
- remaining total. “Total” modes transmit the sum of both sides of the display on channel 1. • EI-1000 - Press SELECT and use the arrows to change the value if needed. Press SELECT again. • EI-2000 - Use the right arrows to change the value if needed.
EI-1000 & 2000: Setup, Calibration and Troubleshooting. 8. Use the arrows to set the output to 20.00 mA on your meter. 9. Press select on the EI-1000 to stop the flashing and save the value. 10. If needed, now hook your meter to CH2 output.
Page 14
EI-1000 & 2000: Setup, Calibration and Troubleshooting. Remote readout shows proper values in test mode but doesn’t track the weight shown by on the indicator’s display: • Check the P600 and P610 settings as shown in the “4-20 mA parameter setup”...
EI-1000 & 2000: Setup, Calibration and Troubleshooting. Day tank use – Instrument setup: If you are using a day tank you must set parameter 113 to off. This is the auto zero. If this parameter is not turned off the instrument will zero out the reading as the chemical is pumped in.
7. To verify the calibration, add another known amount of product. The weight reading should increase to match. 8. Press the right arrow twice on the EI-1000 to exit the menus and return to the gross weight reading. On the EI-2000 use the left SELECT.
Page 17
7. To verify the calibration, add another known amount of product. The weight reading should increase to match. 8. Press the right arrow twice on the EI-1000 to exit the menus and return to the gross weight reading. On the EI-2000 use the left SELECT.
Page 18
2.o (Channel 2 zero) 2.c (Channel 2 span) See the EI-1000 and EI-2000 Operating manual for more in depth information on these instruments. We’ll be glad to email it to you if needed. Contact Eagle Microsystems with any questions at 610-323-2250.
Page 19
EI-1000 & 2000: Setup, Calibration and Troubleshooting. Notes: 9/10/14 6:59 AM Page 19 of 19...
Page 20
EI-1000 FEATURES • Up to four scale bases • User selects scale base via front panel keypad. • Up to four continuous 4-20mA outputs • Six-digit display for up to 999,000 lb or kg scale displays. • Up to ten relay outputs EI-2000 FEATURES •...
Page 21
PUSHBUTTONS The EI-2000 has a set of pushbuttons for each of the two channels, and the EI-1000 has one set of pushbuttons that control up to four channels (scale bases). The pushbutton functions listed below describe their typical uses. However, when applicable, the pushbutton(s) may be used for other functions, which is documented in this manual.
Page 22
The descriptions below will reference the set of pushbuttons for the left or right (CH1 or CH2) channel on the EI-2000 or the set of pushbuttons on the EI-1000 that control all, up to four, scales. Configuration parameter P001 selects which operating mode, 1 thru 5, is enabled for all scales.
Page 23
MODE 2: ADJUSTABLE REMAINING Typical application: drum scales - Remaining & gross weights are displayable. An empty scale should read zero. If it does not, press the ZERO pushbutton. If “no 0” is displayed, access the calibration mode to zero the scale. Place container with product on the scale.
Page 24
The user has the option of accessing ALL configuration, calibration and diagnostic functions without opening the enclosure. Parameter P999 selects whether or not this is enabled (it is set to no or yes). EI-1000: Press and hold [SELECT] + [RIGHT ARROW] for greater than three seconds.
Page 25
When the FPA menu is first displayed, the CFG option is shown. The selections are configuration (CFG), calibration (EI-2000 has CAL and the EI-1000 has CAL1 through 4), analog output (AOUt), and diagnostic (dIAG). Press the UP or DOWN ARROW pushbutton below to select another mode, and press the “enter”...
Page 26
P400-P499 Scale 4 parameters (EI-2000-EXP must be installed) P100 Scale selection (auto-configuration) This parameter selects the Eagle Microsystems scale connected to the scale channel input Px00 Function This turns the channel off if it’s not needed (2, 3 and 4).
Page 27
This sets the weight division for the unit selected by the lb/kg switch. The choices for the EI- 2000 are 0.1, 0.2, 0.5, 1, 2, and 5. The choices for the EI-1000, since it has a six digit display, are extended to the range of 0.0001 to 500.
Page 28
P116 Threshold (0 to full-scale) This applies to operating modes 2 through 5. This sets the point (in lbs or kgs) that the meter will enable, depending on the operating mode, switching to remaining or enabling adjustment of the gross weight calibration or remaining weight. P117 Start (0 to full-scale) This parameter applies to operating modes 3 &...
Page 29
METER K1 (CH1) K2 (CH2) K3 (CH1) K4 (CH2) SP-A (P510) SP-A (P520) SP-B (P511) SP-B (P521) EI-1000 (1 or 2-ch) EI-2000 METER K5 (CH3) K6 (CH4) K7 (CH3) K8 (CH4) EI-1000 (3 or 4-ch) SP-A (P530) SP-A (P540) SP-B (P531) SP-B (P541) Note: Both the EI-2000-EXP and the EI-2000-IO are both installed in an expanded enclosure.
Page 30
SERIAL INTERFACE PARAMETERS RS232 and RS485 serial interface control parameter(s): P700 serial mode (basic/RTU) BAS: BASIC serial protocol: RS-232 only RTU: MODBUS-RTU serial protocol: RS-485 or RS-232 (see end of manual for register list) P701 Baud Rate Select the serial port baud rate (default = 9600). The selections are 300, 600, 1200, 2400, 4800, 9600 &...
Page 31
<LF> is a line-feed control character (0aH). • The numbers shown in the text boxes (in blue if printed in color) are character indexes for programming purposes. RS232 Wiring to PC DB9 Connector EI-1000/2000 DB9 (PC) TB6-3 (C) 5 (SG) Signal ground...
Page 32
WEIGHT SUMMATION FUNCTION (FIRMWARE VERSION 6.40 AND ABOVE) The EI-2000 has the ability to sum all of the scales connected to the meter (up to two) as well as two additional scales, which are connected to another EI-2000 (secondary) meter. P990 SUM CH (sum all channels) YES: Sum all active scale inputs.
Page 33
To alter the value of a parameter, press the SELECT pushbutton. Use the UP/DOWN ARROWS pushbuttons to set the desired value and then press the SELECT pushbutton when finished. PARAMETER EI-1000: Ao.tESt PARAMETER EI-2000: Ao t xx.xx This parameter tests both Digital to Analog Converters using ten, linear 1.6mA steps from 4.00 mA to 20.00 mA.
Page 34
Each analog output is calibrated to 4mA and 20mA points, so each of the up to four analog outputs has a 4mA and 20mA point setting – see the parameters below: EI-1000 ANALOG OUTPUT 4.00 mA (ZERO) ADJUSTMENT: EI-1000 (x = 1,2,3, or 4) Press PARAMETER EI-1000: Ao.x...
Page 35
Use the formula below to calculate an approximate calibration factor that’s entered into this parameter. To shift the decimal point, press the RIGHT ARROW pushbutton on the EI-1000 or the RIGHT SELECT pushbutton on the EI- 2000.
Page 36
The 4-20mA analog outputs, scale inputs and ALL parameters must be checked and changed to proper values to match the scale. Recalibration of the scale is required. EI-1000: [DOWN ARROW] + [SELECT] + [UP ARROW] EI-2000: [LEFT UP ARROW] + [LEFT SELECT] + [RIGHT SELECT] Load typical settings for the selected scales.
Page 37
HARDWARE TB1 – POWER - A.C. power connections Ground Earth ground line Neutral Non-fused neutral line Fused line TB2 – RELAYS – Set point relay contacts The relays, which are mechanical, can switch 3A, 250V A/C or D/C loads. The contacts are “dry;”...
Page 38
EI-1000 & EI-2000 METER SELECTION HEADER SHUNT (J3) Connect J3-3 and J3-4 to select an EI-2000 meter; the display is the EI-2000-DSP. Connect J3-1 and J3-2 to select an EI-1000 meter; the display is the DSP1000. Invalid shunt positions: REMOTE SCALE SENSE...
Page 39
A OL Scale input negative overload -A OL -A OL *Channel 1 & 2 A/D hardware failure. FAIL.AD FAIL.AD *IIC bus error IIC.Err IIC.Err *EEPROM device was not found EE.Err EE.Err * Hardware failure: Return the EI-1000/EI-2000 to factory for service.
Page 40
3 3/16" Enclosure conforms to: NEMA 4X SPARE PARTS SPARE PARTS EI-2000-DSP P/N = 110266 1. INDICATOR CPU BOARD: EAGLE P/N 110267 DSP1000 (EI-1000) = 110271 2. INDICATOR DISPLAY BOARD: P/N 110266 CPU P/N = 110394 EI-2000-IO = 110396 EI2000 - O&M-v3.0...
Page 41
LOAD FULL-SCALE FULL-SCALE FULL-SCALE # LOAD SCALE EX Ω SI Ω NOTE CELL HINGED mV/V CELLS IHPN WT-3600 4000 1820 RED GRN WHT 500680 RED GRN WHT 500317 TB5 (RIGHT CHANNEL) DCS-302 1/side TB4 (LEFT CHANNEL) EDS-400 RED GRN WHT 500317 LP-4310 LP-4320...
Page 42
MODBUS REGISTER LIST (FIRMWARE VERSION 6.xx) SEE Modbus_Application_Protocol_V1_1b.pdf FOR PROTOCOL FORMAT REVISED: 2015-09-18 ADDRESS NAME UNITS TYPE #REG CHANNEL 0x0000 TEST VALUE uint16 This reads a 66.6ms/tick counter, which has the range from 0 to 0xffff. 0x0900-0x0903 Device name char x 8 0x0904-0x090b firmware name char x 16...