Fluke P5523 User Manual

Fluke P5523 User Manual

Liquid to gas separator

Advertisement

Quick Links

P5523
Liquid to Gas Separator
Users Manual
PN 3952355
November 2010
© 2010 Fluke Corporation. All rights reserved. Printed in USA. Specifications are subject to change without notice.
All product names are trademarks of their respective companies.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the P5523 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Fluke P5523

  • Page 1 P5523 Liquid to Gas Separator Users Manual PN 3952355 November 2010 © 2010 Fluke Corporation. All rights reserved. Printed in USA. Specifications are subject to change without notice. All product names are trademarks of their respective companies.
  • Page 2 Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available only if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price.
  • Page 3: Table Of Contents

    Table of Contents Title Page General Information................... 1 How to Contact Fluke ..................2 Safety Summary....................2 Compressed Liquid & Gas ................2 Personal Protective Equipment..............2 Symbols Used in this Manual ................2 Preparations ....................... 3 Connect the Gauge ..................4 Connect the Test Port Insert ................
  • Page 4 P5523 Users Manual...
  • Page 5 List of Tables Table Title Page Symbols........................2 Test Port Insert - Parts List..................5...
  • Page 6 P5523 Users Manual...
  • Page 7 List of Figures Figure Title Page Hydraulic/Pneumatic Circuit Schematic ..............1 Connecting the Gauge .................... 4 Test Port Insert ....................... 5 Angle Adapter ......................7 Pointer Remover/Punch..................7...
  • Page 8 P5523 Users Manual...
  • Page 9: General Information

    General Information The P5523 is used for calibrating high-pressure gas measuring instruments against a hydraulic Deadweight Tester. Note The terms ”Master Test Gauge” and “Gauge” in this document refer to any pressure-measuring instrument such as Transfer Standards, Digital Calibrators and Transducers.
  • Page 10: How To Contact Fluke

    P5523 Users Manual How to Contact Fluke To order accessories, receive operating assistance, or get the location of the nearest Fluke distributor or Service Center, call: • Technical Support USA: 1-800-99-FLUKE (1-800-993-5853) • Calibration/Repair USA: 1-888-99-FLUKE (1-888-993-5853) • Canada: 1-800-36-FLUKE (1-800-363-5853) •...
  • Page 11: Preparations

    The external gas supply is connected to the ¼ NPT port in the rear of the P5523 case. A compressed gas cylinder fitted with a suitable pressure regulator is recommended.
  • Page 12: Connect The Gauge

    P5523 Users Manual gle11.eps Figure 2. Connecting the Gauge Connect the Gauge Refer to Figure 2 for steps 1-4. 1. Screw the appropriate gauge adapter fully on to the instrument to be tested. 2. Screw assembly down COUNTER-CLOCKWISE onto test port.
  • Page 13: Priming

    Liquid to Gas Separator Priming gle012.eps Figure 3. Test Port Insert Table 2. Test Port Insert - Parts List Item Description Part Number O Ring 3865142 Test Port Insert 3919892 O Ring 3883397 Test Port — To calibrate panel-mounted gauges with pressure connections in the rear, use an Angle Adapter.
  • Page 14: Operation

    1. With the system fully depressurized, remove the reservoir cap from the left-hand reservoir. 2. Ensure reservoir valve is closed on the deadweight tester, and turn the screw press out to draw fluid back into the deadweight tester reservoir. (Observe fluid level in P5523 reservoir.)
  • Page 15: Ancillary Equipment

    Liquid to Gas Separator Ancillary Equipment 3. When the fluid has been withdrawn from the separator, the connection hose can be disconnected from the deadweight tester. Periodically, the right-hand reservoir body should be inspected and cleaned, to remove any fluid that may be present. Ancillary Equipment Angle Adapter, P5543 To calibrate gauges with the pressure connection on the rear (e.g.
  • Page 16 P5523 Users Manual...

Table of Contents