Alert switch

Both, Radon Scout and Radon Scout PLUS are equipped with a switch output which can be
used for ventilation control or alert indication. The potential-free contacts of the switch will be
connected at the end of the sampling interval if the measured Radon concentration exceeds
the threshold. The output will be deactivated one minute before the subsequent interval
ends. Therefore, the sampling interval must be set at least to two minutes. The alert
threshold is fixed at 300Bq/m³ for the Radon Scout while it is user adjustable in the Radon
Scout PLUS. The connector is located on the rear panel of the instrument. In case of pending
alert, the red light at the front panel will blink instead of the green one.
As an accessory, we offer a wireless power switch which can be used for direct ventilation
control. The switch includes a transmitter unit to be connected to the alert switch output of
the instrument, and a plug adaptor (receiver) which can be placed between the wall outlet
and the power cable of the ventilation unit. Optionally, a flush-mounting power switch is also
available. The plug adaptor offers an additional timer function to define the duration of the
ventilation period independently on the sampling interval of the instrument (for example 15
minutes on while the sampling interval is 60 minutes). The wireless switch is coded - that
means, several transmission lines can be established within one building. The assignment
between transmitter and receiver is easily done by a push button.
The sampling interval should be set to 60 or 120 minutes if the threshold level is lower than
300Bq/m³.

Technical Data

Alert Switch
Max. switchable voltage
Max. switchable current
Max. pulse current
connector
Wireless switch
Transmitter
Frequency
Range
Power supply
Plug adapter
Max. switchable power
Timer settings
Manual_Radon-Scout_EN_24-03-17.doc
24V
0,4A
1,5A
2-pins
Manufacturer: Binder, series 710
2 inputs
Additional input can be used for manual
switch
433MHz
With coding for secure transmission
approx. 30m
Battery CR2032
Good for 20000 switch actions
220V/50Hz
6 code memories for group switching
(e.g. for more than one Radon monitor)
3500W
7s, 5min, 30min,
If the timer function is not used the
1h, 2h, 4h, 8h
power switch follows the alert switch
6

Advertisement

loading