Testing Procedure

Running a Test

For more information, see the test cassette package inserts. The
Quick Reference Guide gives a brief summary of the procedure.
NOTE: Gloves should be worn whenever working with samples
that are potentially biohazardous.
1. If the cassettes have been refrigerated, allow them to
come to room temperature (at least 10 minutes) before
opening.
2. Make sure the analyzer is plugged in.
3. Remove the cassette from its pouch. Hold the cassette
by the short sides only. Do not touch the black bar or
the brown magnetic stripe. Place the cassette on a flat
surface.
4. Press RUN. The analyzer will do a selftest, and the
screen will display:
5. The cassette drawer will open, and the screen will display:
X indicates current sample setting.
W= Whole Blood, S= Serum
6. Place the sample into the cassette well. Use the Alere
Cholestech LDX
Use an appropriate MiniPet
samples and quality control, calibration verification, and
proficiency testing materials. Check the package insert for
the test cassette you are using to determine the volume
for the sample type you are using. See "Performing a
Fingerstick" and "Using Your MiniPet
22–25, for additional information.
18
Alere Cholestech LDX
Selftest
running.
Selftest OK
Load cassette
and press RUN....(X)
X = W or S
®
Capillary Tube for fingerstick samples.
®
System User Manual
®
Pipette for venous blood
®
Pipettes", pages