• REVIEW AND UNDERSTAND ALL SAFETY PROCEDURES IN THE "HANDLE WITH CARE"
BOOKLET BEFORE ATTEMPTING TO INSTALL, OPERATE OR PERFORM MAINTENANCE
ON THIS LN
CONTROLLER.
2
• DO NOT ATTEMPT TO USE OR MAINTAIN ANY LIQUID NITROGEN FREEZER UNTIL YOU
READ AND UNDERSTAND THESE INSTRUCTIONS.
• DO NOT PERMIT UNTRAINED PERSONS TO USE OR MAINTAIN THIS UNIT.
• IF YOU DO NOT FULLY UNDERSTAND THESE INSTRUCTIONS, CONTACT YOUR
SUPPLIER FOR FURTHER INFORMATION.
• BEFORE ATTEMPTING TO OPERATE THIS CONTROLLER WITH ANY WORTHINGTON
INDUSTRIES LABS OR K SERIES UNITS, YOU MUST READ THE SEPARATE OPERATING
AND SAFETY MANUAL PROVIDED WITH THAT UNIT.
CAUTION - SAFETY FIRST!
CS200 Control Systems
LABS and K Series Cryostorage Systems
Operation & Maintenance Instructions

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CS200 and is the answer not in the manual?

Questions and answers

Summary of Contents for Worthington CS200

  • Page 1 • IF YOU DO NOT FULLY UNDERSTAND THESE INSTRUCTIONS, CONTACT YOUR SUPPLIER FOR FURTHER INFORMATION. • BEFORE ATTEMPTING TO OPERATE THIS CONTROLLER WITH ANY WORTHINGTON INDUSTRIES LABS OR K SERIES UNITS, YOU MUST READ THE SEPARATE OPERATING AND SAFETY MANUAL PROVIDED WITH THAT UNIT.
  • Page 3: Table Of Contents

    Keep Equipment Area Well Ventilated ..............5 Liquid Nitrogen System ..................5 Electrical ......................5 GENERAL INFORMATION ..............6 CS200 Series Control System Specifications ............7 DELIVERY AND RETURNS ..............10 Unpacking and Inspection ..................10 Freight Damage Procedures ................. 10 Repackaging for Shipment ..................
  • Page 4 Test Alarms ......................37 Logging ......................37 Display Brightness ....................38 Making Adjustments to the CS200 Series Control System Sensor Assembly ....38 Removing/Installing the Solenoid Valve ..............39 Removing/Installing the Controller 10K/24K Units ..........39 Removing/Installing the Thermocouple ..............40 Removing/Installing the Sensor Probes ..............
  • Page 5: Safety Precautions

    • This product is intended to be used in hospitals and clinics. any service on these units without first • This product has no Radio Transmitter (Intentional Radiator) functions. disconnecting the electrical power cord. • This product is not intended for electromagnetic shielded rooms only. CS200 Control Systems...
  • Page 6: General Information

    General Information The CS200 SERIES Control System can monitor and control both the liquid nitrogen level and the vapor temperature range in the cryostorage unit you have selected. CS200 SERIES Control Systems are designed to work with Worthington Industries Cryostorage Systems. The features are designed to provide a safe environment for samples while at the same time tracking all relevant information associated with the freezer.
  • Page 7: Cs200 Series Control System Specifications

    General Information CS200 SERIES Control System Specifications Specifications Control Type Level Control & Temperature Control Level Measurement Sensor Type 8-Thermistor Fixed (standard) Range 8 inch range (8-Thermistor) Redundancy Multiple discrete points Temperature Measurement Sensor Type Type T Thermocouple or 1.5% of reading...
  • Page 8 General Information CS200 SERIES Control System Specifications (cont’d) User Interface Display Type Vacuum Fluorescent Display (VFD) Buttons Level, Temperature and Alarm Information “At a Glance” status LED status wheel Filling 1 LED Menu Access 1 LED Language English and German...
  • Page 9 1.0 lbs (.45 kg) Battery Width 7.25” (184.2 mm) Battery Height 6.375” (161.9 mm) Battery Depth 3.25” (82.6 mm) Battery Weight 12.4 lbs (5.6 kg) NOTE: Above measurements are for LABS; CS200 controllers are integrated into all K Series models. CS200 Control Systems...
  • Page 10: Delivery And Returns

    Worthington Industries for shipping and packing instructions. It is the responsibility of the customer to assure that the goods are adequately packaged for return to the factory. All freezers returned to Worthington Industries must be clean and decontaminated before return.
  • Page 11: Installation

    Liquid Nitrogen source. The CS200 SERIES Control System is designed to be operated at normal room temperatures 15° C to 27° C (60° F to 80° F) at a relative humidity level below 50%.
  • Page 12: Power Supply Connection

    Installation power supplies and electrical accessories for use with the CS200 SERIES Control WARNING: System and they are not listed or referenced in this manual, it is the responsibility of This equipment that third-party supplier to determine compliance with the standards and tests in the is intended for EN Compliance tables found on pages 50–54.
  • Page 13: Operation

    Quick Start Guide for freezer set-up. Start operation of the control system by plugging the power cord into the wall outlet. Press and hold the POWER button for 2 seconds. The CS200 Series Control System will go through a short startup routine and then start operation. Refer to Figure 2.0 CS200 Controller Display for a description of the controller’s navigation buttons.
  • Page 14: Cs200 Controller Display Guide

    Figure 1.0 CS200 Controller Display Operational Theory The CS200 SERIES Control System automatically maintains the Liquid Nitrogen (LN ) level and monitors temperature in the Cryostorage freezer. Operational conditions are monitored High Level Alarm and any alarm is triggered if necessary.
  • Page 15: Alarm Conditions

    NOTE - Offset value for LABS94K only = 3.0 in (126 mm). Note: Please see Alarm Conditions description of The CS200 SERIES Control System monitors a number of conditions and provides an the offset. Level alarm if a problem is detected. The alarms are listed below: displayed may vary depending on the offset setting.
  • Page 16: Lid Switch

    The Lid Switch (Figure 4.0 and 4.1) is attached to the hinge and determines whether or not the lid is open on the freezer. This also allows the control to determine whether to activate the Auto Defog, Quick Chill or Lid Alarm features. Lid Switch Figure 3.0 LABS Lid Switch Figure 3.1 K Series Lid Switch CS200 Control Systems...
  • Page 17: Solenoid Valve

    Power Supply A 12 VDC power supply is supplied for the CS200 SERIES Control System. This system is supplied with a universal power supply that accepts 100/240 VAC (50/60 Hz). UL approval for the system as a whole is not required since the control operates on low voltage.
  • Page 18: Temperature Monitoring

    6 in. (152 mm). Temperature Monitoring The CS200 SERIES Control System uses a Type T thermocouple to monitor the temperature in the vessel. The thermocouple is factory-installed near the top of the Worthington Industries Cryostorage freezer vapor chamber and the temperature is measured at that point.
  • Page 19 Operation Table 3.0 Annual Maintenance Schedule CS200 SERIES Control Examine for exposure to moisture, wear and tear, connector problems, and damage to the faceplate or buttons. In addition, periodic firmware updates may be important Harness Assembly Examine for damage to the cable and damage to connectors, Battery Examine connection cable and connector for damage.
  • Page 20: Controller Operation

    Controller Operation This section of the operating manual is for Worthington Industries approved equipment that uses the CS200 SERIES Control System. Introduction The CS200 Control System, temperature and LN level controller is designed for easy operation and reliable uninterrupted service. This controller will maintain the selected liquid level range of LN in your freezer as well as providing audible and visual alarms for any non-conforming conditions that may occur.
  • Page 21 The control will continue to monitor all sensors and conditions while the user accesses the menu system. If no activity is detected for 30 seconds, the control will automatically return to the main operational screen. CS200 Control Systems...
  • Page 22 Low Temperature Alarm: Allows the user to set the low temperature alarm for both thermocouples. Settings [0 to -190°C] • Time with no Temp alarm: Displays the time since the last high or low temperature alarm • Temperature Units: Allows the user to choose the units of measure for temperature. CS200 Control Systems...
  • Page 23 U.S. and International format. • Date & Time Sync: This feature can be used to Figure 5.0 Set Date/Time Menu automatically synchronize the date and time between the CS200 Control and the ciSmart Com Communications Hub. • Lid Functions •...
  • Page 24 • Maintenance Reminder: Allows the user or service company to acknowledge a reminder that preventive maintenance should be performed on the freezer. Contact Worthington Industries Customer Service for instructions how to reset the reminders. • Acknowledge Reminder: Acknowledges the maintenance reminder and resets it 12 months into the future.
  • Page 25 Displays the control serial number, the firmware version, the freezer ID and any options that are enabled. DATA MANAGEMENT: Allows the user the capability to download operational data. USAGE: Provides easy access to view the usage data. CS200 Control Systems...
  • Page 26: Temperature

    Temperature Control The CS200 SERIES Control System has the capability to control the vapor temperature at the top of the freezer. The temperature is controlled at the location of thermocouple 1.
  • Page 27: Menu System

    SET DAY SET TIME SET TIME Daily- Daily-Off-Day Off-Day 00:00-24:00 00:00-24:00 ACKNOWLEDGE REMINDER Enter Level Measured 0:1250 Enter Level Measured 0:1250 (in .5” increments) (in .5" increments) NEXT REMINDER 00-00-00 00:00 Figure 6.0 Menu System - View Only CS200 Control Systems...
  • Page 28 SELECT LANGUAGE ENGLISH-GERMAN SYSTEM INFORMATION SERIAL # 12345 VERSION 1. Options FREEZER ID FRZ01 DATA MANAGEMENT 00-00-00 All Logs-Temp Logs- Alarm DOWNLOAD LOGS Logs- Temp&Alarm Logs USAGE 0.00 IN PER DAY Figure 6.1 Menu System - Change Settings CS200 Control Systems...
  • Page 29: Factory Defaults

    30 minutes Lid Open Alarm Delay 10 minutes Valve Stuck Alarm Audible Retrigger 30 minutes Display Brightness Screen Saver Time 30 minutes Freezer ID Not set Scheduled Fills Supply Reminder Settings Password Power Password Table 5.0 Factory Defaults CS200 Control Systems...
  • Page 30: Troubleshooting

    The CS200 SERIES Control System has incorporated state of the art diagnostic tools to assist in the identification and correction of any issues that may arise.
  • Page 31: Wiring Diagram

    Red + Black P2P1 White/Yellow + Yellow P2P1 White/Orange + Orange P2P1 Remote Alarm Connector White/Brown + Brown P2P1 Black P2P2 Brown P2P2 P2P2 Sensor Assembly Orange P2P2 Yellow P2P2 Figure 8.0 CS200 Series Control System Harness Wiring Diagram CS200 Control Systems...
  • Page 32: Installation & Setup

    Controller Operation Intallation & Set-up The CS200 CONTROL SYSTEM consists of the following components: • Main Control • VFD Display Module • Wiring Harness Assembly • Power Supply • Thermocouple Assembly • Lid Switch Assembly • Sensor Assembly • Cryogenic Solenoid Valve •...
  • Page 33 Each is keyed so they can only be Remote Battery plugged in one way. Alarm Lid Switch Harness Connector to Main Control Connector Valve and Freeze Guard Sensor Connector Figure 12.0 Harness Assembly & Back Panel CS200 Control Systems...
  • Page 34: External Connector Ratings

    2.5mm barrel connector CryoWire Secure™ Modular shielded jack Remote Alarm 300 volts 5mm terminal block Solenoid Valve 600 volts 4.2mm header Thermocouple 2-pin thermocouple Level Sensor 5 amps / contact Sealed circular connector Table 6.0 External Connector Ratings CS200 Control Systems...
  • Page 35: Led Status Wheel Flash Patterns

    The CS200 CONTROL SYSTEM uses a Type T thermocouple to measure temperature within the Cryostorage freezer. The temperature curve for a thermocouple is nonlinear so it is important that the CS200 CONTROL SYSTEM have a good calibration to provide accurate temperature readings. The temperature is traceable to the National Institute of Standards and Technology (NIST) ITS-90 Thermocouple Database.
  • Page 36: Test Temperature System

    This is an easy means to tell if sensors are in or out of liquid or if a new sensor assembly is needed (open or defective sensors). If the sensor assembly is unplugged and the main control is still on, the display will indicate that a sensor error CS200 Control Systems...
  • Page 37: Alarms And Error Conditions

    ) and an infinite resistance (when an open circuit appears in the level sensing circuitry). Alarms and Error Conditions The CS200 SERIES Control System control monitors many different conditions in the freezer and has a full complement of alarms associated with these different conditions. As alarms occur, they trigger an audible tone as well as a flashing red light on the status wheel.
  • Page 38: Display Brightness

    • Return to the CS200 SERIES Control System main screen and the level indicated should match the physically measured reading. The Sensor Offset, the START FILL and the STOP FILL can be set through the CS200 SERIES Control System menu system.
  • Page 39: Removing/Installing The Solenoid Valve

    (2) two supplied screws. Be sure that all of the necessary installation procedures have been completed before you start to fill the refrigerator. To start filling, refer to Filling the Refrigerator (Initial Fill) section of this manual. Figure 15.0 K Series Controller Removal CS200 Control Systems...
  • Page 40: Removing/Installing The Thermocouple

    Carefully remove the sensor tube plug from the sensor tube and remove the sensor leads from the plug. Making Adjustments to the Level Settings The factory settings for the CS200 control system are as follows: • Low Level Alarm = 2 (Always 1” below the start fill) •...
  • Page 41: Removing/Installing The Solenoid Valve

    (2) mounting screws. Attach the solenoid valve lead connection to the controller board. At the completion of maintenance or repairs, install the controller using the procedure outlined for your refrigerator model in the Removing/Installing the Controller section. CS200 Control Systems...
  • Page 42: Replacement Parts

    Figure 16.2 K Series 10K/24K 367264 Elbow, 90 Deg,. 1/4" MNPT x 3/9" ODT Brass 367170 Union 3/8" ODT x 3/8" ODT, SS 367174 Connector, Male, 3/8" ODT x 1/4" MNPT, Brass 367259 Elbow, 90 Deg, Street, 3/8" x 1/4" Brass CS200 Control Systems...
  • Page 43: Controller Assembly

    Junction Box, CS LABS 367791 Valve/Freezegaurd Harness LABS20K/38K/40K 367792 Valve/Freezegaurd Harness LABS80K/94K 367782 Harness Assembly, CS LABS 367788 Level Sensor, 8-Thermistor CS200 367785 Thermocouple Assembly, CS LABS 367786 Thermocouple Wire, Short LABS20K/38K/40K 367787 Thermocouple Wire, Short LABS80K/94K 367795 Lid Switch Assembly...
  • Page 44 Solenoid Valve Harness, K-CS-Series, 36” 367823 12 VDC Transformer No Plug 367824 Transformer Cover, LABS CS 367913 Junction Box, With No T Plate, CS200 Figure 16.4 K Series Controller 367908 Controller Module, 10K/24K CS200 Assembly Parts 367911 Harness Assembly 10K-CS, 74” Cable 367912 Harness Assembly 24K-CS, 104”...
  • Page 45: Troubleshooting

    Troubleshooting Symptoms The key to troubleshooting your CS200 SERIES Control System and your Cryostorage system is to determine which component in the system is the source of the problem. Determine if the problem is occurring in any of the following subsystems: Supply Vessel, Transfer Line, Power Source, Temperature, Level Sensing, Security, Lid Switch, Solenoid Valve, Control Display, Alarm System, Communications.
  • Page 46: Indicates Low Ln Supply

    • Prepare an ice water slurry with crushed ice and tap water. Dip or pour LN into a styrofoam cup to prepare an LN bath. Calibrate the controller. Select MENU, TEMPERATURE, CALIBRATE TEMPERATURE. Follow the on-screen instructions. Hold the thermocouple in each bath until the control completes its self- calibration. CS200 Control Systems...
  • Page 47: Fill Solenoid Cycles On And Off

    Fill Solenoid Cycles On and Off CS200 SERIES Control System features a timer function whereby the valve open duration is limited, allowed to shut for a short warm-up and then opened again. To confirm or change the valve open duration: •...
  • Page 48: Push Buttons Will Not Respond

    If power is interrupted, a log of the time and duration will be recorded. If the unit is not connected to a battery back-up or an external alarm, no local alarm will sound unless a high temperature is detected after power is restored. A Remote Alarm will be triggered anytime the power is interrupted. CS200 Control Systems...
  • Page 49: Service And Maintenance History Log

    Date: ___________________________________________________________________________________________ NOTE: All Worthington Industries Cryostorage Systems must be cleaned and decontaminated prior to return to Worthington Industries for repair or maintenance and must be accompanied by a written statement to this effect. Any Cryostorage System received without this statement will be returned to the sender, freight collect.
  • Page 50: Appendix

    Harmonic Emissions all establishments, including domestic establish- Class A ments and those directly connected to the public EN/IEC 61000-3-2 low-voltage power supply network that supplies buildings used for domestic purposes. Voltage fluctuations/ Flicker Emissions Complies EN/IEC 61000-3-3 CS200 Control Systems...
  • Page 51 Power frequency magnetic fields (50/60Hz) magnetic field should be at levels characteristic of a typical location in a typical commer- EN/IEC 61000-4-8 cial or hospital environment. Note UT is the a.c. mains voltage prior to application of the test level. CS200 Control Systems...
  • Page 52 CS CONTROL SYSTEM should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the CS CONTROL SYSTEM. B. Over the frequency range 150kHz to 80MHz, field strengths should be less than 3V/m. CS200 Control Systems...
  • Page 53 NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. CS200 Control Systems...
  • Page 54: Ec Declaration Of Conformity

    • IEC/CISPR 11:2009+A1:2010 Radiated & Conducted Emissions. • IEC61000-3-2:2005+A1:2008+A2:2009. Harmonics • IEC 61000-3-3:2008. Flicker Year of CE Marking: 2012 Manufacturer: Pacer Digital Systems, Inc. Signature: Kevin Oeff (Digitally signed by Kevin Oeff) Position: President Date: 24 May 2012 Place: Indianapolis, IN USA CS200 Control Systems...
  • Page 55: Warranty

    (2) years from the original date of purchase. Worthington Industries at its option will either repair or replace any item covered under this warranty.
  • Page 56 CONTACT US 844-273-7517 US Toll Free: +43 7485 606 286 / +1 614-438-7968 Global: Sales Support/Order Placement: customerservice@worthingtonindustries.com Technical Services: techservices@worthingtonindustries.com ©2017 Worthington Industries. All rights reserved. CS200 Controller Operation Manual P/N 372409 - 01/17 7950-8346...