Af Measur Ements - Quer Y Af Level Measur Ement - Aeroflex 3900 Series User Programming Manual

3900 series digital radio test set; tetra programming manual
Hide thumbs Also See for 3900 Series:
Table of Contents

Advertisement

2 .2 .1 2
A F M e a s u re m e n t s - Q u e r y A F L e v e l M e a s u r e m e n t
:F E T C h : A F : A N A L y z e r :L E V e l ?
De scri pt ion:
Quer y D ata:
st atu sbyt e ( NR 1):
fai lbyt e ( NR 1):
avg count ( NR 1):
av g ( NR 2):
u nit s ( NR 1):
Quer y R esp onse:
TET RA B S R emot e C omman ds
Co mm an d re tur ns AF Level measur ement d ata.
<stat usbyte> ,<f ailbyte >,<avg co unt>, <avg> ,<unit s>
0 = Val id
1 = Inval id
2 = Set tling
4 = Inaccur ate
6 = Set tling and Inaccur ate
7 = Set tling , I naccura te an d In va lid
0 = All limit checks passe d
1 = Aver age upper fai led limi t
2 = Aver age lowe r f ailed limit
value
mV ( U nbalanced )
dBm (B alanced)
6 = dBm
7 = V
11 = W
12 = mW
13 = μW
16 = dB r
17 = dB V
20 = nW
:FE TC h:A F:A NA Ly z er: LEV el?
0,0, 1,0. 002
St atusbyt e and Failb yte may ret urn mo re than one con dition as a bitmask.
S u b j e ct t o E x p o r t C o n tr o l , s ee C o ve r P a g e f o r d e t ai l s .
2 - 7