Important!
DO NOT
invert the instrument with the sparge bottle attached.
DO NOT
transport or store the instrument with liquid in the sparge bottle.
DO NOT
handle or carry system when sample is being analyzed.
Powering the FROG
The FROG may be run on battery or an AC power source.
Battery:
1.
The battery is already installed in the FROG.
2.
There is a battery indicator on the back of the FROG. Five bars show that the FROG
is fully charged.
AC Power Source:
1.
The FROG can be run on an AC Power Source. To run on power, plug power supply
(Figure 3) into power source and then into 12 VDC Power port on side of FROG
(Figure 2).
To turn the FROG on, slide the power switch located on the side of the instrument into
the up position. A green light should appear (Figure 4).
Installing the Ellvin GC Software
The Ellvin GC Software is located on a USB Drive in the zipper pouch of the FROG's
case. To install the software run the setup file located in the folder labeled Defiant Tech
Ellvin.
Recommended Computer Specifications:
more current, and 2 GB memory.
Connecting The FROG to Ellvin
1.
Open the Ellvin GC Software
2.
Plug serial data cable (Figure 5) into data port on side of FROG (Figure 2).
3.
Turn FROG on
4.
In Ellvin select the dropdown menu to the right of the port icon (Figure 6) and
select a COM port (the FROG is usually the last port on the list).
5.
If connection is successful in the lower left corner of the software it will display:
CONNECTED PARAMS LOADED
6.
If a connection was unsuccessful it may be necessary to install the proper drivers.
These can be installed manually by opening the CDM application on the software
USB drive under Defiant Tech Ellvin\USB to Serial Drivers\ CDM 2.04.06.exe.
Setting up the FROG for Sampling
Water/Soil Sampling will require:
Sparge Tube and 5 mL Syringe. For blank water samples use distilled or de-ionized
water.
Air Sampling will require:
FROG Air Sampling Kit (pg 6). For blank air samples the carbon filter tube in the air
sampling kit can be used to filter the air or any other source of clean air may be used.
Refer to the following pages for further details: Water Analysis page 4, Soil Analysis
page 5, Air Analysis pages 6 and 7.
Before sampling:
1.
Make sure instrument settings are correct for the application
2.
Calibrate or have the Frog calibrated for the application
3.
Run a blank before attempting sample analysis. A blank run should produce a clean
baseline (Figure 7)
4.
(Optional) Adjust Retention Times if using the Frog at a different location than the
one where it was calibrated. Elevation changes will affect retention times.
Home
TM
FROG-5000
Quick Start Guide
Pentium dual core or faster, Windows 7 or
www.defiant-tech.com
Figure 1: Do not invert FROG with sparge
bottle attached.
Battery
Indicator
Here
Figure 2: FROG-5000 and close up of data and power ports
Figure 4: Power switch (on position)
Figure 5: Serial Data Cable
Figure 6: Selecting a COM port to connect FROG to software
Figure 7: Clean baseline in Ellvin, free of peaks (Note: there
is a slight peak in the beginning, this is normal and accepta-
ble)
Figure 3: Power Cable

Advertisement

loading
Need help?

Need help?

Do you have a question about the FROG-5000 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Defiant FROG-5000

  • Page 1 Installing the Ellvin GC Software The Ellvin GC Software is located on a USB Drive in the zipper pouch of the FROG’s case. To install the software run the setup file located in the folder labeled Defiant Tech Ellvin. Figure 2: FROG-5000 and close up of data and power ports...
  • Page 2 Setting the Frog up for an Application It is important to make sure the Frog is using the correct settings for an application. If the Frog was calibrated by Defiant or another company the settings have already been established and set for you.
  • Page 3 Figure 12: Opening a Calibration File. Calibration files can be created (with analysis data from chemical standards) or loaded (if someone else calibrated the FROG) with the software. Defiant recommends re-calibrating the FROG at least every three weeks for the most Figure 13: Analytes List and accurate analyses.
  • Page 4 Figure 25: Play Button to start run in Software Step 6 Step 7 Figure 24: Changing Valve position Figure 23: Lifting Sparge Bottle up and Home to “Run Water” tightening sparge nut to hold it in place www.defiant-tech.com...
  • Page 5 Extract is diluted to a 5mL sample volume and then run Figure 29: Lifting Sparge Bottle into the Raised (Up) the same way a water sample is. For details on Methanol Position on the Frog Extraction please see the application note at: www.defiant-tech.com. Home www.defiant-tech.com www.defiant-tech.com...
  • Page 6 This sequence will run until you pause the unit by clicking on the stop button in the Ellvin software. Figure 38: Play Button to start run in Software Step 4 Figure 37: FROG Ready for Air Sampling Home Figure 36: Valve set to “Run Air” www.defiant-tech.com...
  • Page 7 This keeps particulates from damaging the air sampler. Figure 45: Attaching threaded Figure 46: Wand attached to Transfer Line, fitting on Wand to Transfer Line ready for sampling. Figure 47: Sampling with the Air Sampler Wand. Home www.defiant-tech.com www.defiant-tech.com...
  • Page 8 Right axis: GC Temperature Left (y) axis: Sensor Data (clicks from PID) GC Temperature Data Chromatogram PID Data Notes Section Bottom (x) axis: time in seconds Connection Indicator GC Temperature Indicator Figure 55: Chromatogram in Live Data Tab with Features labeled Home www.defiant-tech.com...
  • Page 9 Once the report is generated it can be saved as a pdf, word document or excel file (Figure 65). Figure 64: View Report Figure 63: Chromatogram Integrated, Data shown in Results Home Figure 65: Report with Save Icon Circled. Table. www.defiant-tech.com...
  • Page 10 Frog-5000™ Packing Instructions: Item List 1. Syringe 2. Sparge Bottle 3. Frog (With Sparge Tube Protector) 4. Serial Data Cable 5. Battery 6. Air Sampler a. Air Sampler Coupler Line b. Transfer Line c. 1/16-3/16 Adapter d. Activated Carbon Scrubber e.
  • Page 11 Frog-5000™ Packing Instructions: Item Locations 1. Syringe 2. Sparge Bottle 3. Frog (With Sparge Tube Protector) 4. Serial Data Cable 5. Battery (In Frog) 6. Air Sampler a. Air Sampler Coupler Line b. Transfer Line c. 1/16-3/16 Adapter d. Activated Carbon Scrubber e.