CLEANING AND DISINFECTION OF DELTASTREAM HC
Valid for:
ME DPHC 0001 deltastream HC, 230V
ME DPHC 0002 deltastream HC, 115V

Safety instructions:

Attention
Read the safety instructions
below before starting to clean
the unit:
• This system is a closed circuit. The water line should not
be disconnected in the operating area. Do not perform
water changes and disinfection in the operating area.
Carry out water changes and disinfection only in hygien-
ically clean surroundings.
• Pay
attention to hygienically clean work processes. (Dis-
posable gloves, protective mask, hood).
Make sure you and your staff are aware of mycobacteria
and their effects.
Regularly check the system for microbiological contami-
nation. We recommend to test the water quality once a
month, nevertheless the system is in use or stored.
• Disconnect the mains plug before starting to clean and
disinfect the unit.
For all surfaces and parts use only disinfectants based on
aldehydes, ammonium components or alcohols that do
not attack ABS plastic, PVC or PU. As far as possible, do
not use disinfectants based on phenol derivatives be-
cause they shorten the life of plastics. A list of suitable
disinfectants can be found on Page 2 of these Brief
Instructions as well as on Page 26 of the IFU.
• Make sure no liquid enters the unit (e.g. through the ven-
tilation slots).
• Only switch on the device if neither the surface is wetted
nor if liquid is inside the device outside the tank.
• Only use sterile water for disinfecting the device!
• Fill sterile water into the tank only with auxiliaries that
have been previously disinfected.
BRIEF INSTRUCTIONS
Xenios_HC_EN_003_2018_02
Surface
Surface cleaning and disinfection
After each water change, carry out a surface disinfection
(device housing, hose set and connectors).
If the device is contaminated by contact with body flu-
ids, disinfect the surface and, if necessary, the water-
carrying parts.
Make sure that after surface disinfection no residues of
the cleaning agent are left on the surface and the hous-
ing resp. interior of the device is dry before use.
For disinfecting the surface of the device, please, note
the list of approved surface disinfectants on Page 2 of
this Quick Guide or Page 26 of the IFU. When using the
disinfectant, observe the manufacturer's instructions.
Also clean and disinfect the screw thread of the filler
neck.
Do not switch the unit on again until the disinfectant has
evaporated fully.
For more information, please see the deltastream HC
IFU in chapter 7 from page 25 and following.
Ventilat ion slots
Check the ventilation slots on the sides of the unit for dirt
regularly, but at least every six months. Do not open the
device yourself.
Dust deposits in the unit will reduce the efficiency of the
system. Have any inside dirt removed by an authorized
service technician.
These cleaning instruction only apply in connec-
tion with the currently valid instruction for use of
the above mentioned devices and does not re-
place it!
CAUTION
During the water change or the
disinfection, no oxygenator may
be connected to the water circuit.
Pay attention to hygienic handling
when cleaning the device!

Advertisement

loading

Summary of Contents for Xenios AG Deltastream ME DPHC 0001

  • Page 1: Safety Instructions

    BRIEF INSTRUCTIONS CLEANING AND DISINFECTION OF DELTASTREAM HC Surface Valid for: Surface cleaning and disinfection ME DPHC 0001 deltastream HC, 230V • After each water change, carry out a surface disinfection ME DPHC 0002 deltastream HC, 115V (device housing, hose set and connectors). •...
  • Page 2 BRIEF INSTRUCTIONS CLEANING AND DISINFECTION OF DELTASTREAM HC Filling the device during initial operation and recommis- Company Product name Actve substance sioning: Isopropanol Isopropanol (70%) Bode Chemie KohrsolinTM Aldehyde • Fill approximately 1 liter of sterile water, e.g. with GmbH SteriliumTM Propanol the help of a disinfected funnel into the water tank...
  • Page 3 For more infor-mation, please take a look at the deltastream HC IFU in chapter 7 starting on page 25. DISTRIBUTOR MANUFACTURER Xenios AG medos Medizintechnik AG Obere Steinfurt 8-10 Im Zukunftspark 1 74076 Heilbronn...

This manual is also suitable for:

Deltastream me dphc 0002