Cleaning and disinfection

Advertisement

CLEANING AND DISINFECTION OF DELTASTREAM HC
Disinfection
• Empty the system as described.
• Insert the adapter between the Hansen couplings.
• Fill the device again with the chosen H2O2 solution until the
tank is completely filled.
• Leave the fuel cap open during the cleaning process, as over-
pressure may occur.
• Switch on the device and allow the disinfectant solution to
circulate at 35 °C. Ventilate the hose set.
• Complete this process after a total of 30 minutes and drain
the used disinfectant solution from the water circuit as de-
scribed.
• Refill with sterile water
• Take the device for 5 min. in operation and then drain the wa-
ter from the water cycle. Repeat this procedure several times
with fresh sterile water.
• The water cycle must be flushed until a suitable detection me-
dium (e.g. test strips "Sanosil and pH test") no longer detects
disinfectant residues.
• If the device continues to operate, refill it with sterile water
until it is slightly below the MAX mark on the level gauge.
• Stored devices must be disinfected every 30 days to prevent
the formation of a biofilm.
• If a biofilm has formed, all water-bearing parts must be re-
placed.
• Liquids should never penetrate the inside of the device
outside of water-bearing parts. Medos Medizintechnik AG
recommends removing devices whose internal parts have
come into contact with foreign fluids from the sur-gical area
and to disinfect them afterwards. Contami-nated devices may
only be used again once disinfection and maintenance have
taken place. For more infor-mation, please take a look at the
deltastream HC IFU in chapter 7 starting on page 25.
BRIEF INSTRUCTIONS
DISTRIBUTOR
Xenios AG
Im Zukunftspark 1
74076 Heilbronn
Germany
Phone +49 7131 2706-0
Fax +49 7131 2706-299
info@xenios-ag.com
www.xenios-ag.com
Xenios_HC_EN_003_2018_02
For the disinfection of the water cycle a H2O2 -
containing solution is recommended. Concentra-
tion of H2O2 should be between 3% - 7.5%.
 For a 3% H2O2 solution of Sanosil™
S015 mix 400 ml of Sanosil™ S015
with 600 ml of sterile water.
Sanosil™ S015 is not available in every region!
Please make sure that a suitable product is ap-
proved in your region! (e.g., Amity™ 75).