Summary of Contents for Peak Scientific Corona

  • Page 1 Corona Air Compressor Installation Guide Corona Air Compressor Installation Guide Rev 2 PN IG-0017 RSID 271 EN 22/08/18...
  • Page 2: Change History

    User Manual for your Air Compressor. Thank you for selecting Peak Scientific to meet your laboaratory gas generation needs, and should you require any further assistance or support please do not hesitate to contact Peak Scientific or Peak Partner from which you purchased your air compressor.
  • Page 3: Safety Notices

    If the user employs an operating procedure, item of equipment or a method of working which is not specifically recommended by Peak Scientific, the user must ensure that the equipment will not be damaged or become hazardous to persons or property.
  • Page 4: Fittings Kit

    Fittings Kit 1/4” Teflon Tubing 00-1266 US Mains Cable C13 6-15 04-1029 US Mains Cable C13 5-15 04-1039 1/4” x 1/4” Compression Fitting 02-4421 UK Mains Cable C13 04-1028 2.5mm Hex Key 00-5483 EU Mains Cable C13 04-1038 Page 4...
  • Page 5 WARNING Unpack the air compressor from the shipping crate and position on a flat surface, in desired area. WARNING the weight of the unit constitutes a two man lift. Safe lifting practices are advised. Remove the fascia from the air compressor, by first pulling towards you. Page 5...
  • Page 6 Next pull the fascia upwards and away from the air compressor. CAUTION 00-5483 Now take the left side panel off, by removing the two screws shown in the image above, using the hex key (00-5843) provided Page 6...
  • Page 7 The side panel can now be removed by pulling the panel towards you. With the side panel removed you now have access to the transit bracket. This is located under the compressor and is marked with a sticker reading Remove Before Use.
  • Page 8 The bracket can now be slid out from underneath the compressor tray. Re-attach the side panel and replace the screws. Page 8...
  • Page 9 Attach some of the supplied ¼” Teflon Tubing (00-1266) to the Air Output port and connect the other end to the Corona Nitrogen Generator. Attach the rest of the supplied tubing to the Drain port and connect it to a suitable unsealed drain container.
  • Page 10 CAUTION 04-1028/04-1029/04-1038/04-1039 Select the appropriate mains cable from the fittings kit and plug the IEC 60320/C13 cable into the mains input at the rear of the air compressor. Please note both 110V and 230V US cables have been supplied, select the correct mains cable for your supply voltage.
  • Page 11 Press the POWER switch located at the rear of the air compressor. CONGRATULATIONS Your Peak Scientific Corona air compressor is now fully installed. operational and ready to supply air on demand to your generator. Page 11...
  • Page 12: General Notes

    General Notes * Electrical requirements are 110 or 230VAC nominal +/- 10% depending on chosen model. However, running continuously at voltages outwith this is not recommended. Extended periods at extremes can have a detrimental effect on the operation and life of the air compressor.
  • Page 13 See our enclosed Peak [Protected] leaflet for further information. Important! You have 1 month to register your Peak Scientific product from the date of installation. Once registered the warranty will be honoured for a period of 12 months. If you wish to defer the installation of your generator, you must notify Peak Scientific immediately by emailing warranty@peakscientific.com.
  • Page 14 Page 14...
  • Page 15 Page 15...
  • Page 16 User Manual for your air compressor. Peak Scientific has highly trained, fully certified Field Service Engineers located in over 20 countries across every continent around the world. This allows us to provide an industry-leading rapid response service to our customers.