Table of Contents

Advertisement

EN
Nursing Anne
Simulator
User Guide
www.laerdal.com

Advertisement

Table of Contents
loading

Summary of Contents for Laerdal Medical Nursing Anne

  • Page 1 Nursing Anne Simulator User Guide www.laerdal.com...
  • Page 2 Intended Use Nursing Anne Simulator is a realistic, interactive training simulator to educate healthcare professionals to initiate and maintain care for patients, within the fundamentals of nursing curricula. It is tetherless, WiFi operated, with a flexible operating system depending on training needs.
  • Page 3: Table Of Contents

    Use - Airway Management and Resuscitation Resuscitation Preparing for Simulation - Blood Pressure (BP) Arm Spontaneous Breathing Enema Defibrillation – Nursing Anne Simulator and ShockLink Suppository Rectal Suppositories Use - Cardiac Related Skills Vaginal Suppositories Heart, Lung, Bowel and Vocal Sounds...
  • Page 4 Contents Use - Sounds Use - Pulses Use - Ear Ear Irrigation Use - Recommended Sizes of Clinical Equipment Cleaning - After Each Use General Care External Skin Indwelling Reservoir Blood, Urine and Stomach Reservoirs Genitalia After Enema Administration After Suppository Administration Stomas IM Injection Pads IV Arm...
  • Page 5: Operating Software

    • Voice Conference Application (VCA) transmits all vocal sounds used during from here. Laerdal Simulation Home is located in the Laerdal Medical folder under the simulation. It enables the instructor to communicate through the simulator during the Windows start menu (Windows 7) and can be launched using the desktop shortcut on session.
  • Page 6: Overview - Nursing Anne Simulator

    Overview - Nursing Anne Simulator Overview - Features Airway and Breathing • Spontaneous Breathing • Head Tilt, Chin Lift • Tracheostomy care and suctioning Microphone Microphone • Oxygen delivery via BVM ventilation, Nasal Canula, Non-Rebreather Mask • Oral and Nasal Intubation •...
  • Page 7: Overview - Reservoirs

    Overview - Reservoirs Overview - Reservoirs Abdominal Plate Overview Urine Port Air Port G-Tube Port Stomach Port Reservoirs Urine Reservoir Central Line Reservoir Stomach Reservoir Esophagus Tube Connection Port Reservoir Cap 1. Central Line Port 5. Air Reservoir Tubing 2. G-Tube 6.
  • Page 8: Preparing For Simulation - Reservoirs

    Preparing for Simulation - Reservoirs Preparing for Simulation - Reservoirs To prepare for simulation, fill Urine (yellow), Central Line (red) and Stomach (green) Cautions Reservoirs with fluid as required. • Do not perform chest compressions with fluid in the stomach reservoir bag. Filling Reservoir - Quick Fill •...
  • Page 9: Central Line Reservoir

    Central Line Reservoir Stoma Placement Nursing Anne Simulator features an opening for preplacement of a central line and Unzip torso skin on manikin’s right side. Insert luer lock connection on back of selected a 500 ml reservoir for practice of infusion and dressing change.
  • Page 10: Iv Drain Bag

    Preparing for Simulation - IV Arm Preparing for Simulation - Blood Pressure (BP) Arm Nursing Anne Simulator’s arms provide radial IV access through female luer fittings, and Nursing Anne Simulator features bilateral BP Arms for non-invasive measurement of support training for IV drug or fluid administration.
  • Page 11: Enema

    Preparing for Simulation - Enema and Suppository Administration Preparing for Simulation - Trach Tube Insertion Enema A concealed tracheostomy opening in the neck of the simulator allows for the care and maintenance of a placed Trach Tube. To prepare, remove Trach Plug. To prepare for enema administration, connect the enema connector to the bypass 1.
  • Page 12: Preparing For Simulation - Eyes

    Eye lid position and blink settings are adjusted and controlled in the instructor software application. See SimPad or LLEAP User Guide for instructions. Changing the Upper Dentures Nursing Anne Simulator comes with a set of upper teeth that may be removed for care Caution and cleaning practice.
  • Page 13: Preparing For Simulation - Skin

    ) as an adhesion promoter (primer), applied prior to simulation ® • Apply dressing activities undertaken on the Nursing Anne Simulator. • Gently remove dressing as to not remove too much Mastisol • Clean area gently as per local protocol (Mastisol will last longer by using less...
  • Page 14: Use - Patient Handling

    Nursing Anne Simulator has articulation for: NG Tube Insertion and Care • Realistic patient handling procedures Nursing Anne Simulator features an esophagus and 500 ml stomach reservoir for practice • Protective positioning of NG tube insertion and skills such as lavage and gavage.
  • Page 15: Gastrostomy Tubes (Gastric/Jejunal)

    Connect to Bypass System for large volume infusions. See Large Volume Infusions (Bypass Rectal Suppository Insertion system) section. Nursing Anne Simulator will allow the insertion of real or simulated rectal suppositories. Urinary Catherization Note Nursing Anne Simulator is fitted with realistic female genitalia. The pressurized urinary Only 1 suppositories should be used at any time.
  • Page 16: Use - Intramuscular (Im) Injections

    Use - Intramuscular (IM) Injections Use - IV Arm Simulated medications can be administered via intramuscular injections in several sites. Connect syringe or IV tubing (for medical administration and/or infusion) with luer locks Clean IM foam pads after each use. See ‘Cleaning’ section. to the pre-ported IV in the arm.
  • Page 17: Airway

    Use - Airway Management and Resuscitation Use - Airway Management and Resuscitation Airway Nursing Anne Simulator accepts a range of airway devices and airway techniques, and is CPR capable. The following can be practiced: • Head tilt/Chin lift • Jaw thrust •...
  • Page 18: Defibrillation - Nursing Anne Simulator And Shocklink

    Heart, Lung, Bowel and Vocal Sounds features an extensive library of ECG variations. See SimPad PLUS User Guide or LLEAP Nursing Anne Simulator has heart, breath, bowel and vocal sounds which are determined Help Files for operating instructions. by the scenario used and controlled by the instructor.
  • Page 19: Use - Pulses

    Use - Pulses Use - Pulses Nursing Anne Simulator has palpable pulses in various locations. Pulses are synchronized Nursing Anne Simulator features automatic pulses in the carotid (bilateral), radial and with the ECG and adjustable with three different strengths: weak, normal and strong.
  • Page 20: Use - Ear

    Use - Ear Use - Recommended Sizes of Clinical Equipment Ear Irrigation Device Recommended Size Irrigation, cleaning and administration of eardrops can be practised in the ear canal. Urinary Catheter Size 14Fr Nasogastric Tube Size 16Fr Nasogastric Feeding Tube Size 12Fr Gastrostomy Tube Size 16Fr Central Line...
  • Page 21: Cleaning - After Each Use

    Cleaning - After Each Use Cleaning - After Each Use Clean individual components following a training session where heavy usage occurs, or Indwelling Reservoir when liquids have been introduced into the system. 1. Unzip and remove chest skin. Cautions 2. Gently remove the abdominal plate. •...
  • Page 22: Genitalia

    Cleaning - After Each Use Cleaning - After Each Use Stomas Genitalia Clean with mild soap and water. 1. Remove genitalia from pelvis manifold. Detach from urine reservoir. 2. Clean with mild soap and water. IM Injection Pads 3. To reconnect genitalia, connect genitalia to valves on urine reservoir and anus and Immediately after use, remove IM injection pads from the simulator.
  • Page 23: Torso Liner

    Cleaning - Every Six-Months Cleaning - Every Six Months Inspect for worn internal parts and install the replacement consumables available for sale Wig Care and Upkeep on www.laerdal.com. Inspect the internal chest cavity and replace any consumable parts as When brushing the wig, use combs and brushes that are specially designed for wigs. To needed.
  • Page 24: Cleaning And Maintenance

    Cleaning and Maintenance Preventative Maintenance Preventative maintenance (PM) service can be purchased for this system. This service will allow you to maintain the product in optimal operating condition. Inspections Checklist Checkpoints Daily/Weekly Six Month Annual Preventative Contact your local Laerdal representative for more information. Maintenance (Depending on usage, expect that these will...
  • Page 25: Maintenance

    Maintenance Maintenance Inspect Compressor Air Filter Change Compressor Air Filter Air Filter must have weekly inspections. Use the Air Filter color guide to determine if filter needs to be changed. Air Filter Color Guide Change 1. Remove torso skin 2. Remove Reservoirs from Torso Liner 3.
  • Page 26: Replacing Spontaneous Breathing Bladders

    Maintenance Transport and Storage Replacing Spontaneous Breathing Bladders Cautions • Do not store this product outside the storage conditions specified in the Important Product 1. Unzip chest skin to expose rib plate. Information. 2. Gently lift away rib plate to expose the two spontaneous breathing bladders. •...
  • Page 27 © 2018 Laerdal Medical AS. All rights reserved. Manufactured in USA for: Laerdal Medical AS P.O. Box 377 Tanke Svilandsgate 30, 4002 Stavanger, Norway T: (+47) 51 51 17 00 www.laerdal.com...

Table of Contents