4.1. Assembly

A front panel cut-out of 45
depth is ≥ 165 mm. The screw clips provided with the DA 6000-N are for fastening it after it has been in-
serted into the opening.
The electrical connections should be made as shown in the schematic below.

4.2. Electrical connections on the rear panel

4.2.1. Power supply connection

The DA 6000-N can be powered by a mains
voltage of 85 to 265 V AC (connections 1, 2 and
3) or, alternatively with a voltage of 18 to 30 V
DC (connections 4 and 5).
Please be sure that the mains cable connec-
tions are sufficiently protected from contact!
Note:
Only one of the two power supply possibilities can be connected!
The mains supply connection must meet the requirements of DIN/VDE regulation 57 411!

4.2.2. Interface connection

The digital interface is for the connection to a
digital. The serial interface is galvanically iso-
lated from other units connected to the display
and can be ordered as either an RS232 or
RS485. This contact arrangement is as fol-
lows:
Contact
RS232
8
Rx
9
Gnd
10
Tx

4.2.3. Connecting the control input (switching bridge)

Partial keyboard lock
Parameters can only be set using the operating
buttons, when the connection between 6 and 7
is open (the LED above the enter button will be
illuminated)
12
+0.6
+0.8
x 92
mm is necessary to install the DA 6000-N. The required installation
RS485
A
S
B
(Both RS232 and RS485 can just be use for one Pyrometer)
1
2
3
4
5
L
N PE
+
-
85...265V~
18...30V=
Power Input
1
2
3
4
5
violett
TxD
Digital -
rot
Gnd
pyrometer
schwarz
RxD
1
2
3
4
5
6
Disable
Enter
6
7
8
9 10
6
7
8
9 10
Rx Gnd Tx
(RS232)
A
S
B
(RS485)
7
8
9 10