Phenix PH50 series Manual

Monoculaire microscop

Advertisement

Quick Links

Phenix PH50 monoculaire microscoop
4503000

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PH50 series and is the answer not in the manual?

Questions and answers

Summary of Contents for Phenix PH50 series

  • Page 1 Phenix PH50 monoculaire microscoop 4503000...
  • Page 3 Although the design of our product provided the best safety for your usage.  But, inaccuracy of operation method or neglect the usage elucidation probably cause body hurt and the prop-erty lose. For your safety. Before using this instrument ,please carefulness reading the instructions. Don’t lose the instruction manual, please saving it at near the instrument, in ...
  • Page 4: Table Of Contents

    Contents Main Features ..........................5 Usage and Coverage Area ....................... 5 Electrical Capability ......................... 5 Working Circumstance of Microscope .................... 6 Structure and Nomenclature ......................7 Instrument’s Installation ......................... 8 Operation Procedure ........................10 Digital Part, Installation and Useage of Parts ................16 Care and maintenance ........................
  • Page 5: Main Features

    Main Features PH50 series biological microscope have Mono Viewing and Bino Viewing for choice. And can be built-in CMOS or CCD camera system equipped with USB output ports, making use of computer image processing software and computer multimedia technology, so that you have observed show for variety of video figures , can capture images of static and dynamic video, with Objective measurement and a variety of image processing functions.
  • Page 6: Working Circumstance Of Microscope

    Working Circumstance of Microscope This microscope is a precise optical instrument, if the usage or safekeeping isn’t appropriate, it will cause the instrument damaged or influence precision. While choose to use the place, please consider the following condition The place for this microscope should not be too bright, and direct sun shine on ...
  • Page 7: Structure And Nomenclature

    Structure and Nomenclature Fig.1 (1) Viewing head:Viewing components. (2) Stand: The microscope’s main body. (3) Revolving nosepiece:Used to select objectives needed(Locate with sound). (4) Objective:Magnify specimen for the first time. (5) Slide clip:Clip specimen. (6) Mechanical stage:Lay specimen (7) Condenser moving knob:Adjust the height of condenser. (8) Power plug:Connect power supply.
  • Page 8: Instrument's Installation

    Instrument’s Installation 6.1 Place the microscope on a flat and stable working table, while moving the instrument, especially the optical parts, make sure to avoid to contact the lenses surface with hand or artic with grease.6.2 On the way to transport carry the objective table, it's surface will be covered with the protection film, before use please tear up the prot-...
  • Page 9 6.4 Turn on the power: after all above procedures, please insert the Power plug on the wall outlet, and turn on the Power switch , make sure you are using the right voltage. 6.5 Insert the Eyepiece: slide the eyepiece into the Eyepiece tube. (Fig.4) Fig.4...
  • Page 10: Operation Procedure

    Operation Procedure 7.1 Turn on the power 8.1.1 Turn the power switch(turn switch to "-" ), make the bulb give out light。 8.1.2 Then revolve the handwheel of brightness regulates to regulate the brightness of filed . 7.2 Restoration to adjust diopter of tube To adjust diopter of tube on the R/L eyepiece tube, Make its bottom edge with engrave the line to align, use the same method, adjust the left diopter .
  • Page 11 7.4 Mounting on the specimen 7.4.1 Turn apart the Specimen Clips, put into the Specimen, and use slide clip to hold it. Release the fingers;make sure that the viewing area right in the middle of stage.(Fig.6) 7.4.2 Adjust the stage hand wheel make sure that viewed field locates objec -tive under.
  • Page 12 7.5.2 Rotating Coarse Focusing wheel, moving the stage to the tiptop (Fig.9)。 7.5.3 Adjust Coarse focusing knob to find sharp image. Adjust the diopter ring on the diopter eyepiece until the image is clear.(Fig.10) 7.6 Adjust fine focusing knob to find clear image. (Fig.11)
  • Page 13 7.7 Rotating the Condenser moving knob, make condenser move highest and a little lower. If you obverse dispersion image in view field , please adjust condenser(Fig.13).
  • Page 14 7.8 Rotating Revolving nosepiece,choose magnification that you need. 7.9 Adjust the Viewing head setscrew, in order to objective easy. (Fig.14) 7.10 If objective selected is 100X oil (spring), use oil as observing medium. By this the quality of image can be improved and the specimen is observed clear.
  • Page 15 7.11 When the microscope is not needed, turn off the power switch.
  • Page 16: Digital Part, Installation And Useage Of Parts

    Digital Part, Installation and Useage of Parts 8.1 Take out the digital observation, Place the pin at locating hole of the observation, Load the observation lightly, Press the upper observation softly by one hand , Fixing the screw on both sides of the body. At the time of installation to ensuret digital head tight.
  • Page 17: Care And Maintenance

    Care and maintenance 9.1 The power switch at the back side of Base is main power control. If the microscope is no longer in use, please switch off the power to avoid the electronic elements in working status. If the microscope will not be used for a long period , please unplug the power wire from socket.
  • Page 18: Replacing Light Bulb And Fuse

    10. Replacing light bulb and fuse 10.1 Replacing light bulb 10.1.1 Turn off the microscope ,unplug the power cord from socket. 10.1.2 Waiting for 30min, until it cools down. 10.1.3 Screw off the light base, and pull out the light base board, and rotate at some angel until the light base comes out.
  • Page 20: Troubleshooting And Handling

    11. Troubleshooting and Handling Symptom cause Handling Optical Parts nosepiece Revolving the nosepiece, make it positioning(Different stalk into right position light road) Condenser position too low Adjust the height of Not centering with Condenser Readjust the condenser Uneven brightness Wrong installation of light bulb Check if the light bulb has been inside Field of view mounted correctly...
  • Page 21: Configuration Table

    12. Configuration table PH50 Biological Microscope:Specification PH50 Series Product sequence number 1A31 1A32 2A42 Name Specification 2L-A 3L-A 3L-A 3L-A 40X-640X ● ● Total Magnification 40X-1600X ● ● ● ● ● Wide Field WF10X/18mm ● ● ● ● ● ●...
  • Page 22 ● ● ● ● ● Package Aluminium Metal box TV viewing head with simple CCD Others adapter PH50 Digital Microscope Specification for parts PH50 Series Name Specifiaction DM048U DB048U DB130U DB200U Main optical machine part Total 40X-1600X ● ● ●...
  • Page 23 brightness adjustable Power transformer ● ● ● ● Package Aluminium Metal box ● ● ● ● Dividing ruler 0.01mm ● ● ● ● Slides Plant slides ● ● ● ● CCD & CMOS CCD Camera 470K Pixels,480 TV Line CMOS Camera 480K Pixels,800×600 ●...