Download Print this page

Specifications; Performance Data - Sanyo MDF-5386SC Specifications

-86ºc ultra low temperature freezer

Advertisement

Performance Data

evOluTIONARY DeSIgN
■ The newly developed refrigeration system and freezer
Pull-down characteristics
stem
structure minimizes intrusive operating noise.
40
Central air temp. in cabinet
■ Achieves 10% higher energy conservation than competing
20
Ambient temp. 30°C, No-load
0
ation
models.
–20
■ The newly developed back cover combined with new
V, 50Hz)
–40
mbined
aerodynamically designed and placed components in the
–60
and
refrigeration compartment provide superior air flow, making
–80
on
it possible to drastically reduce the stress to the freezer and
w,
–100
0
2
contributing to excellent durability.
rs
■ Two independent and insulated inner doors ensure maxi-
Back
cover
Pull-up characteristics
mum interior chamber uniformity.
during power failure
ACCeSSIBIlITY
40
■ A new rugged inner door latch tightly closes the inner door
Air
20
doors
and makes accessing chamber contents easier.
0
■ The condenser filter is located at the bottom right side of
–20
the front panel to make filter removal and cleaning easier.
–40
■ Alarm lamp and buzzer offer secure warning of power
–60
–80
failure or abnormal temperature increase.
–100
■ New rugged, one-handed outer door latch has a hasp to
0
2
allow a padlock for additional security
SAFeTY
CO
backup system operation
2
her.
■ high and low temperature warning provides an audible and
characteristics
visual alarm when the temperature deviates more than ±5°C
14
o make
12
to ±20°C (adjustable) from the set point.
10
■ Alarm ring back function ensures that buzzer will resume
8
operation, should alarm conditions continue after it is
6
warning
silenced.
Ambient temp. and maintenance
4
ature
■ Microprocessor-controlled filter-clog check function
of –70°C by injection of LCO
MDF-U5386S
2
protects the refrigeration circuit.
0
rovides
0
5
■ Control panel with digital display for easy operation.
Features
Ambient temperature (°C)
to
• The newly developed refrigeration system
Dimensions
and the freezer structure offer a quiet
t buzzer
experimental environment.
• Achieves 10% higher energy conservation
956 (37.6)
48
d.
(1.9)
890 (35)
than previous model.
630 (24.8)
check
• The newly developed back cover combined
cuit.
with new aerodynamically designed and
placed components in the refrigeration
compartment provide superior air flow,
making it possible
to drastically
reduce the stress
to the freezer and
contributing to
excellent durability.
ith
• Two independent and insulated inner doors
r easy
prevent cold air leakage.
• A new beak style inner
*Appearance and specifications are subject to change without notice.
door latch tightly closes
*Cooling performance is indicated by the temperature reached at the center of the freezer (at ambient temperature of 30°C with no load). In order to use the freezer at a
stable temperature for a long time, it is recommended that the temperature be set to at least 5°C higher than the indicated lowest temperature. In addition, depending on the
the inner door against
usage conditions, it may not be possible to reach the indicated lowest temperature.
the freezer frame. It
also helps make
opening and closing the door smoother.
• The condenser filter is situated at the
bottom right side of the front panel to make
filter removal and cleaning easier.
• Alarm lamp and buzzer offer secure warning
of power failure or abnormal temperature
4
6
8
10
12
14
Time (hour)
4
6
8
10
12
14
Time (hour)
(30kg)
2
10
15
20
25
30
35
EVOLUTIONARY DESIGN
[Unit : mm (inch)]
867 (34.1)
105
(4.1)
795 (31.3)
18 (0.7)
(For model rated 220V, 50Hz)
605.5 (23.8)
140
(5.5)
Compressors
11
(0.4)
Front
Rear
ACCESSIBILITY
SAFETY
w w w . s a n y o b i o m e d i c a l . c o m
• Achieves 10% higher energy conservation
than previous model.
(For model rated 220V, 50Hz)
• The newly developed back cover combined
MDF-U5386S
with new aerodynamically designed and

Specifications

placed components in the refrigeration
Features
MDF-U5386SC
compartment provide superior air flow,
Model
making it possible
–50°C to –86°C (1°C increment)
Temperature range
EVOLUTIONARY DESIGN
to drastically
–86°C
Maximum cooling
Front
• The newly developed refrigeration system
reduce the stress
(Ambient temp. 30°C)
performance
and the freezer structure offer a quiet ex-
to the freezer and
perimental environment.
890 x 875 x 1990mm
Exterior dimensions
contributing to
• Achieves 10% higher energy conservation
(35.0" x 34.4" x 78.3")
(W x D x H)
excellent durability.
than previous model.
630 x 600 x 1280mm
Interior dimensions
• The newly developed back cover combined
• Two independent and insulated inner doors
(24.8" x 23.6" x 50.4")
(W x D x H)
with new aerodynamically designed and
prevent cold air leakage.
Approx. 305kg (672 lbs.)
Net weight
placed components in the refrigeration
483 L (17.1 cu.ft.)
Effective capacity
compartment provide superior air flow,
ACCESSIBILITY
making it possible
Stainless steel, Adjustable,
Shelf
• A new beak style inner
to drastically
Front
3 shelves, W608 x D533mm,
door latch tightly closes
reduce the stress
Max. load: 50kg (110 lbs.)/shelf
the inner door against
to the freezer and
17mm diameter, 3 locations
Access port
contributing to
the freezer frame. It
(back, bottom left/right corner)
excellent durability.
also helps make
Hermetic type, Output: 1100 W
Compressor
• Two independent and insulated inner doors
opening and closing the door smoother.
(high stage side), 1100 W (low
prevent cold air leakage.
• The condenser filter is situated at the
stage side)
bottom right side of the front panel to make
ACCESSIBILITY
HFC refrigerant
Refrigerant
filter removal and cleaning easier.
• A new beak style inner
High/low temperature,
Alarm
door latch tightly closes
Power failure, Filter check,
SAFETY
the inner door against
Battery check
the freezer frame. It
• Alarm lamp and buzzer offer secure warning
Allowable contact capacity:
Remote alarm contact
also helps make open-
of power failure or abnormal temperature
DC 30V, 2A
ing and closing the door smoother.
increase.
• The condenser filter is situated at the
1 set of keys, 1 scraper
Accessories
• High and low temperature warning provides
bottom right side of the front panel to make
Power Requirement
220V / 60 Hz
an audible and visual alarm when the
filter removal and cleaning easier.
Built-in voltage booster
temperature deviates more than 5°C to
20°C (adjustable) from the set point.
SAFETY
• Alarm ring back function ensures that buzzer
• Alarm lamp and buzzer offer secure warning
*Appearance and specifications are subject to change without notice.
will resume operation, should alarm
of power failure or abnormal temperature
*Cooling performance is indicated by the temperature reached at
conditions continue after it is silenced.
increase.
the center of the freezer (at ambient temperature of 30C with no
• High and low temperature warning provides
• Microprocessor-controlled filter-clog check
load). In order to use the freezer at a stable temperature for a long
time, it is recommended that the temperature be set to at least 5C
an audible and visual alarm when the tem-
function protects the refrigeration circuit.
higher than the indicated lowest temperature.
perature deviates more than ±5˚C to ±20˚C
• New rugged, one-handed
In addition, depending on the usage conditions, it may not be
Performance Data
(adjustable) from the set point.
outer door latch has a hole to
possible to reach the indicated lowest temperature.
• Alarm ring back function ensures that buzzer
allow a padlock to securely
Pull-down characteristics
will resume operation, should alarm condi-
protect valuable samples.
Options
tions continue after it is silenced.
40
Central air temp. in cabinet
• Microprocessor-controlled filter-clog check
20
Ambient temp. 30°C, No-load
function protects the refrigeration circuit.
0
• New rugged, one-handed
–20
• Control panel with
outer door latch has a hole to
–40
digital display for easy
allow a padlock to securely
–60
protect valuable samples.
operation.
–80
–100
0
2
4
Inventory rack
Circular recorder
Time (hour)
SUR-624-I
MTR-C954
Distributed by:
Back
Rack Capacity: 24
MTR-85H
• Control panel with
cover
Pull-up characteristics
Recorder mounting
during power failure
bracket
40
MDF-S3085
Air
20
(for MTR-85H)
Distributed by:
0
–20
–40
ISO 9001 & 14001 certified
–60
–80
–100
0
2
4
Time (hour)
1062 Thorndale Avenue, Bensenville, IL 60106 USA
CO
backup system operation
2
Toll Free USA 800-858-8442 • Fax 630-238-0074
characteristics
SANYO Electric Co., Ltd.
14
1- 300 Applewood Crescent, Concord, Ontario L4K 5C7
SANYO Sales & Marketing Corporation
12
5-15, Hiyoshi-cho 2-chome, Moriguchi City, Osaka, 570-8634, Japan
Telephone: +81-6-6992-0322 Fax: +81-6-6992-9291
10
URL http://www.sanyo.co.jp/cmg/biomedical
8
©2006 SANYO Printed in Japan 2006.3. MA. SHR102
6
Ambient temp. and maintenance
4
of –70°C by injection of LCO
0
–20
–40
–60
–80
Performance Data
–100
0
Compressors
Back
cover
Pull-up characteristics
during power failure
40
(For model rated 220V, 50Hz)
Air
20
Rear
0
–20
–40
–60
Compressors
Back
–80
cover
–100
0
Air
Rear
CO
characteristics
14
12
10
8
6
4
2
0
0
Dimensions
48
956 (37.6)
(1.9)
890 (35)
630 (24.8)
Specifications
Dimensions
Model
Temperature range
48
(1.9)
Maximum cooling
performance
Exterior dimensions
(W x D x H)
Interior dimensions
(W x D x H)
Net weight
Effective capacity
6
8
10
12
14
Shelf
digital display for easy
Access port
operation.
LCO
backup system
2
CVK-UB2
LN
backup system
2
Compressor
CVK-UBN2
Refrigerant
Alarm
6
8
10
12
14
Remote alarm contact
SANYO E&E America Company
Accessories
www.sanyobiomedical.com
Power Requirement
SANYO Canada, Inc.
905-760-4025 • Fax 905-760-9945
*Appearance and specifications are subject to change without notice.
*Cooling performance is indicated by the temperature reached at
the center of the freezer (at ambient temperature of 30C with no
(30kg)
load). In order to use the freezer at a stable temperature for a long
2
2
4
6
8
10
12
14
Time (hour)
Pull-down characteristics
40
Central air temp. in cabinet
20
Ambient temp. 30˚C, No-load
0
–20
–40
–60
–80
–100
0
2
4
6
8
10
12
14
Time (hour)
Pull-up characteristics
during power failure
2
4
6
8
10
12
14
Time (hour)
40
20
backup system operation
2
0
–20
–40
–60
–80
–100
0
2
4
6
8
10
12
14
Time (hour)
Ambient temp. and maintenance
of –70°C by injection of LCO
(30kg)
CO
backup system operation
2
2
characteristics
14
5
10
15
20
25
30
12
Ambient temperature (°C)
10
8
[Unit : mm (inch)]
6
Ambient temp. and maintenance
4
of –70˚C by injection of LCO
(30kg)
2
867 (34.1)
2
795 (31.3)
18 (0.7)
0
605.5 (23.8)
0
5
10
15
20
25
30
Ambient temperature (˚C)
[Unit : mm (inch)]
MDF-U5386SC
867 (34.1)
–50°C to –86°C (1°C increment)
956 (37.6)
795 (31.3)
890 (35)
–86°C
18 (0.7)
605.5 (23.8)
630 (24.8)
(Ambient temp. 30°C)
890 x 875 x 1990mm
(35.0" x 34.4" x 78.3")
630 x 600 x 1280mm
(24.8" x 23.6" x 50.4")
Approx. 305kg (672 lbs.)
483 L (17.1 cu.ft.)
Stainless steel, Adjustable,
3 shelves, W608 x D533mm,
Max. load: 50kg (110 lbs.)/shelf
17mm diameter, 3 locations
(back, bottom left/right corner)
Hermetic type, Output: 1100 W
(high stage side), 1100 W (low
stage side)
HFC refrigerant
High/low temperature,
Power failure, Filter check,
Battery check
Allowable contact capacity:
DC 30V, 2A
1 set of keys, 1 scraper
220V / 60 Hz
Built-in voltage booster
35
105
(4.1)
140
35
(5.5)
11
105
(0.4)
(4.1)
140
(5.5)
11
(0.4)
DMREV606.1

Advertisement

loading

This manual is also suitable for: