1. INTRODUCTION OVERVIEW 1. INTRODUCTION 7. TRANSDUCER SETUP Overview Pulse Frequency Congratulations on your purchase of a CorDEX Instruments UT5000 Intrinsically Safe Package Contents Transducer Select Thickness Gauge with CorDEX CONNECT. Multi Echo Probe Zero Built In The UT5000 measures material thickness for Non-Destructive Testing (NDT) and 2.
2. GETTING TO KNOW YOUR UT5000 FRONT VIEW DISPLAY RFID detect/ID Shock-resistant skin Battery level meter RFID Reader Measurement E004010002BCBFCA Measurements Keypad buttons 3.8mm 5.2mm 4.2mm Key function ALARMS 3.2mm 6.0mm Probe 04/07/2011 10:23:50 Settings Pulse rate: 8Hz Material: From List Alumin...
TURNING THE POWER ON/OFF to open the USB and probe connector covers. To switch the UT5000 on or off, press the Power button and hold it down for about 5 seconds. 2. Half turn the two screws on the compartment cover and then carefully lift off the cover.
CONNECTING THE UT5000 REGISTRATION Using the supplied USB cable, connect the UT5000 to a PC with a working internet connection. The UT5000 Mini App starts automatically and prompts you to enter your registration Windows will detect the connection and display the Found New Hardware prompt. Allow details.
100mm square area and layers of paint or rust can interfere with the transmission of the ultrasound. For these situations, the UT5000's EchoEcho technology allows you to measure thicknesses through protective paints monitor the maximum and without having to remove the coating.
To carry out an accurate measurement, it is important that you provide the UT5000 with Before you take a measurement check that you have: the correct velocity of sound in the medium to be measured. The UT5000 contains a table Fitted the probe (see page 5) ...
DETECTING A TAG PULSE FREQUENCY Position the UT5000 so that the cutout area on the top of the unit is within 5cm of the The pulse frequency determines the rate of repeated measurements. To change the RFID tag. Press the Scan button. The RFID tag reference number is displayed across the pulse frequency: top of the screen.
HANDEDNESS Select this option to restore the factory default gauge settings, for example, Pulse Choose whether the main functions of the UT5000 use the buttons on the left or right Frequency. Tag and material information is not affected. side. The “handedness” setting refers to the hand used by the operator to hold the device.
10. USING THE MINI-APP STARTING THE MINI APP Connect your PC to the gauge using the supplied cable. Run the UT5000 miniapp by choosing: Start > All Programs > CorDEX Instruments > UT5000 The following functions are accessed in the USB mode: Adding/editing material velocity information ...
Detection range up to 5cm Supports ISO/IEC 15693-2, ISO/IEC 18000-3 tag formats Battery Details 3.6V Lithium Thionyl Chloride Cell (19.0Ah, SL-2780/S, CorDEX Part No: CDX5000-327) 100 hours continuous operation with back light restrictions Low battery warning Operating Temperature -10 to +50°C...