Page 5
Warranty Tektronix warrants that this product will be free from defects in materials and workmanship for a period of one (1) year from the date of shipment. If any such product proves defective during this warranty period, Tektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product.
Do not operate this product with covers or panels removed. Do not operate with suspected failures. If you suspect that there is damage to this product, have it inspected by qualified service personnel. TDP3500 and TDP4000 Quick Start User Manual...
Page 10
DANGER indicates an injury hazard immediately accessible as you read the marking. WARNING indicates an injury hazard not immediately accessible as you read the marking. CAUTION indicates a hazard to property including the product. TDP3500 and TDP4000 Quick Start User Manual...
Page 11
General safety summary The following symbol(s) may appear on the product: TDP3500 and TDP4000 Quick Start User Manual...
This symbol indicates that this product complies with the applicable European Union requirements according to Directives 2012/19/EU and 2006/66/EC on waste electrical and electronic equipment (WEEE) and batteries. For information about recycling options, check the Tektronix Web site (www.tek.com/productrecycling). TDP3500 and TDP4000 Quick Start User Manual...
Preface This manual describes the installation and operation of the TDP3500 and TDP4000 Differential probes. Basic probe operations and concepts are presented in this manual. You can also check the Tektronix Web site for this document and other related information.
3. Place the probe into the box and stabilize it with light packing material. 4. Seal the carton with shipping tape. 5. Refer to Contacting Tektronix at the beginning of this manual for the shipping address. viii TDP3500 and TDP4000 Quick Start User Manual...
Key features Key features The TDP3500 and TDP4000 Differential probes enables you to make accurate differential measurements from DC to 3.5 GHz and 4 GHz using oscilloscopes featuring the Tektronix TekVPI oscilloscope interface. Key features include: DC to 3.5 Ghz bandwidth (TDP3500) DC to 4 Ghz bandwidth (TDP4000) 100 kΩ...
Operating: up to +40 °C (+104 °F) 5-90% RH Nonoperating: up to +60 °C (+140 °F) 5-90% RH Altitude Operating: Up to 3000 meters (10,000 feet) Nonoperating: Up to 15230 meters (50,000 feet) Pollution Degree 2, Indoor use only TDP3500 and TDP4000 Quick Start User Manual...
Connecting to the host instrument NOTE. Your TekVPI instrument may require a firmware upgrade to support full functionality of the TDP3500 and TDP4000 probes. Before you connect the probe, check the version requirements. (See page 36, Host instrument firmware.) 1. Slide the probe into the TekVPI receptacle.
If the Status LED glows red, a fault may exist. To clear the fault: 1. Disconnect the probe from the oscilloscope. 2. Reconnect the probe to the oscilloscope. If the Status LED continues to glow red, return the probe to Tektronix. TDP3500 and TDP4000 Quick Start User Manual...
Page 19
Offset. 2. Use the touch-screen buttons on the instrument to set the probe parameters. 3. Press the probe Menu button again to close the Probe Control screen. TDP3500 and TDP4000 Quick Start User Manual...
Tektronix part number 016-1884-XX external connections Y-lead adapter Tektronix part number 196-3434-XX 0.25-in square pins for probe tip connections MicroCKT Test Tips (2) 0.25-in square pins-to-mini clips Tektronix part number 206-0569-XX Standard probe accessory TDP3500 and TDP4000 Quick Start User Manual...
Page 21
PROBE COMP terminals on the oscilloscope. 3. Press AUTOSET (or adjust the oscilloscope) to display the compensation waveform. A stable waveform indicates that your probe is functioning properly. TDP3500 and TDP4000 Quick Start User Manual...
The equipment must be warmed up for 20 minutes, and the calibration status of the host instrument must be pass. Required equipment The required equipment for calibration is the same as for the functional check. (See page 6, Required equipment.) TDP3500 and TDP4000 Quick Start User Manual...
3. If the Instrument Calibration Status is not pass, run the Signal Path Compensation routine: Disconnect all probes and signal sources from the oscilloscope, and select Calibrate. When Pass displays in the status box, proceed to the next step. TDP3500 and TDP4000 Quick Start User Manual...
Page 24
Calibration 4. Connect the probe to the oscilloscope PROBE COMP connector. 5. From the Vertical menu, select Probe Cal. TDP3500 and TDP4000 Quick Start User Manual...
Page 25
PROBE COMP connectors on the oscilloscope Verify that the selected channel on the left side of the Probe Setup menu matches the channel that the probe is plugged into TDP3500 and TDP4000 Quick Start User Manual...
Most differential measurements do not require a ground connection, however. More information on the ground connection is available. (See page 34, Probe grounding.) TDP3500 and TDP4000 Quick Start User Manual...
With low amplitude differential input signals, the common-mode input range is almost equal to the operating voltage input range. Higher amplitude differential voltages cause the common-mode input range to be reduced. TDP3500 and TDP4000 Quick Start User Manual...
Page 28
Basic operation Figure 1: Input voltage limits with no DC offset TDP3500 and TDP4000 Quick Start User Manual...
To set the probe offset, follow these steps: NOTE. See your oscilloscope manual for specific instructions on using the offset control. 1. Use the vertical position control to set a zero reference level on the oscilloscope display. TDP3500 and TDP4000 Quick Start User Manual...
Page 30
NOTE. The differential input range of the probe is 2.0 Vp-p, over an operating input voltage range of +5 V to –4 V. The offset range is +1 V to –1 V. TDP3500 and TDP4000 Quick Start User Manual...
Applications Applications The following example shows a simplified AC/DC circuit diagram that uses the TDP4000 probe and a DPO7000C oscilloscope. The oscilloscope is loaded with the DPOPWR power measurement software application to solve a typical test problem. Measure switching loss Measure the power loss at the switching device.
Page 32
3. Select the Power Device tab. 4. Select Switching Loss option and press the Configure button. 5. In the Source Configuration Panel, set the appropriate I-Probe Settings. 6. Select the Deskew button to deskew probes and channels. TDP3500 and TDP4000 Quick Start User Manual...
Page 33
Applications 7. Select Switching Loss measurement. 8. Select Run to acquire data and display the results. (See Figure 3 on page 19.) Figure 3: Switching loss results display TDP3500 and TDP4000 Quick Start User Manual...
Reorder Tektronix part number: 200-4236-XX TDP3500 and TDP4000 Quick Start User Manual...
Page 35
50 to 75 insertion cycles with the probe tip. Replace the adapter after exceeding these limits to avoid unreliable operation. Four adapters are included with the probe. Reorder Tektronix part number: 016-1884-XX quantity 4 TDP3500 and TDP4000 Quick Start User Manual...
Page 36
50 to 75 insertion cycles with the probe tip. Replace the adapter after exceeding these limits to avoid unreliable operation. Four adapters are included with the probe. Reorder Tektronix part number: 016-1885-XX quantity 4 TDP3500 and TDP4000 Quick Start User Manual...
Page 37
Use the solder-in adapter with the resistors and wires in the kit to create soldered test points to your circuit. The kit instructions include installation procedures and specifications for the adapter, resistors, and wires. Reorder Tektronix part number: 020-2505-XX TDP3500 and TDP4000 Quick Start User Manual...
Page 38
2. Connect the adapter to your circuit or to another adapter, such as the MicroCKT test tip. Two adapters are included with the probe. Reorder Tektronix part number: 196-3434-XX quantity 1 196-3468-XX quantity 2 TDP3500 and TDP4000 Quick Start User Manual...
Page 39
2. Turn the tip body to allow better probe orientation. 3. If necessary, slightly bend the sleeve of the tip to reduce stress on the connection. Three adapters are included with the probe. Reorder Tektronix part number: 206-0569-XX quantity 1 TDP3500 and TDP4000 Quick Start User Manual...
Page 40
When selecting the grounding connection, maintain as short a ground path as possible. (See page 34, Probe grounding.) Two ground leads are included with the probe. Reorder Tektronix part number: 196-3465-XX quantity 2 TDP3500 and TDP4000 Quick Start User Manual...
Page 41
Reorder Tektronix part number: 016-1315-XX quantity 5 colored pairs Antistatic wrist strap When using the probe, always work at an antistatic work station and wear the antistatic wrist strap. Reorder Tektronix part number: 006-3415-XX TDP3500 and TDP4000 Quick Start User Manual...
Page 42
1. Place the probe, accessory box, and manual in the carrying case. 2. Close the carrying case to transport the accessories to another location or for storage. Reorder Tektronix part number: 016-1952-XX TDP3500 and TDP4000 Quick Start User Manual...
Page 43
TDP4000 probes. Printed copies are available in the languages listed below. Tektronix part numbers: 071-2212-XX (English) 071-2213-XX (Japanese) 071-2214-XX (Simplified Chinese) 077-0295-XX (Russian) Calibration certificate A certificate of traceable calibration is provided with every instrument shipped. TDP3500 and TDP4000 Quick Start User Manual...
1. Push the IC Micro-Grabber onto the wire lead. 2. Press the lever to open the contacts. 3. Attach the IC Micro-Grabber to your circuit. Two adapters are included with the probe. Order Tektronix part number: SMK4 quantity 4 TDP3500 and TDP4000 Quick Start User Manual...
Page 45
2. Touch the adapter to your integrated circuit leads. One side of each adapter pin is insulated to prevent short-circuiting to adjacent pins. Four adapters are included with the probe. Reorder Tektronix part number: 016-1785-XX quantity 4 TDP3500 and TDP4000 Quick Start User Manual...
Page 46
It provides an SMA connector in the probe signal path for internal probe measurements. Instructions for use are included in the performance verification procedures. Order Tektronix part number: 067-1701-XX TDP3500 and TDP4000 Quick Start User Manual...
Option R3. Repair service 3 years Option R5. Repair service 5 years Manual Options Option L0. English language instruction manual Option L5. Japanese language instruction manual Option L7. Simplified Chinese language instruction manual TDP3500 and TDP4000 Quick Start User Manual...
CAUTION. To avoid damaging the circuitry under test, connect the probe ground (common) to a ground-reference point only. See the technical reference manual on the CD for more information. TDP3500 and TDP4000 Quick Start User Manual...
The lower the impedance of the probe relative to that of the source, the more the probe loads the circuit under test and reduces the signal amplitude. TDP3500 and TDP4000 Quick Start User Manual...
If the Status LED lights red after you connect the probe, an internal probe diagnostic fault may exist. Disconnect and reconnect the probe to restart the power-on diagnostic sequence. If the symptoms continue, the probe is defective, and must be returned to Tektronix for repair. TDP3500 and TDP4000 Quick Start User Manual...
75% isopropyl alcohol solution. A swab is useful for cleaning narrow spaces on the probe. Use only enough solution to dampen the swab or cloth. Do not use abrasive compounds on any part of the probe. TDP3500 and TDP4000 Quick Start User Manual...
Page 52
Maintenance TDP3500 and TDP4000 Quick Start User Manual...