instructions
SOURCE
SOURCE 15PHE PE 4.6/100
Quick Information
SOURCE 15PHE PE 4.6/100 is a pre-packed PEEK column for rapid
preparative hydrophobic interaction chromatography of proteins and
oligonucleotides.
Read the Instruction
The instructions on this page will help you get started quickly with
your new column.
The other side gives more in-depth information on optimisation and
trouble-shooting.
Column data
Matrix:
Polystyrene/divinyl benzene
Ligand:
Phenyl
Bead form:
Rigid, spherical porous, monodisperse
Particle size:
15 µm
Capacity:
At least 40 mg albumin/column
pH stability:
Long term
2–12
Temperature:
Regular use
+4 to +40 °C
Pressure over the column
Regular use
0.25–2.5 MPa
Flow rate:
Regular use
(water at room temperature)
0.5–2.5 ml/min
First time use
Equilibration of the column before initial use or after long term storage
or changing buffers:
1.
8 ml elution buffer.
2.
8 ml starting buffer.
Starting buffer:
50 mM phosphate buffer, 1.5 M (NH4)2SO4, pH 7.0
Elution buffer:
50 mM phosphate buffer, pH 7.0
Flow rate:
1.0 ml/min
Note: Before connecting the column, start the pump and remove
all air in the system, in particular in tubings and valves.
Try these conditions first
Before applying the sample, equilibrate the column.
Proceed according to the section "First time use".
Starting buffer*:
50 mM phosphate buffer, 1.5 M (NH
Elution buffer:
50 mM phosphate buffer, pH 7.0
Gradient:
0-100% B in 33 ml (20 CV)
Flow rate:
2.0 ml/min
* Use a lower concentration of ammonium sulphate if the protein of interest
begins to precipitate at this concentration.
Read the back of this instruction for information
on optimising a separation.
i
71-5002-34 AB 2002-08 • p1
Short term
1–14
Storage
+4 to +30 °C
Do not exceed
4 MPa
Maximum
5.0 ml/min
Injection
valve
)
SO
, pH 7.0
4
2
4
5 µm
Monitor
Buffer and solvent resistance
De-gas and filter all buffers through a 0.22 µm filter.
Daily use
Aqueous solutions pH 2–12
Urea, up to 8 M
Acetonitrile, up to 30% in aqueous buffers
Cleaning
Acetonitrile, up to 30%
Ethanol, up to 100%
Methanol, up to 100%
2-propanol, up to 100%
Hydrochloric acid, up to 1 M
Sodium hydroxide, up to 2 M
Acetic acid, up to 50%
Guanidine hydrochloride, up to 8 M
Anionic, cationic and non-ionic detergents
Avoid
Unfiltered solutions
Oxidising agents
Sample requirements/recommendations
Recommended
sample load:
Sample preparation:
Temperature*:
The sample should be dissolved in starting buffer.
*Hydrophobic interactions increases with increased
temperature. Results achieved at room temperature
may therefore not be reproduced in cold room, or
vice versa.
In Depth Information
Delivery/storage
The gel is delivered in 20% ethanol. If the column is to be stored for
more than two days after use, wash the column with at least 8 ml
distilled water and then equilibrate it with at least 8 ml 20% ethanol.
Instructions
SOURCE 15PHE PE 4.6/100
≤ 40 mg protein/column
Filtered through a 0.22 µm filter or
centrifuged at 10 000 g for 10 min
Ambient

Advertisement

loading
Need help?

Need help?

Do you have a question about the 15PHE PE 4.6/100 and is the answer not in the manual?

Questions and answers

Summary of Contents for Source 15PHE PE 4.6/100

  • Page 1 SOURCE SOURCE 15PHE PE 4.6/100 Instructions Quick Information SOURCE 15PHE PE 4.6/100 is a pre-packed PEEK column for rapid preparative hydrophobic interaction chromatography of proteins and oligonucleotides. Read the Instruction The instructions on this page will help you get started quickly with your new column.
  • Page 2 More rigorous cleaning: Reverse the flow direction and run at a flow rate of 0.2 ml/min with the SOURCE and Drop Design are trademarks of Amersham Biosciences Limited or its subsidiaries. following sequence of solutions: Amersham is a trademark of Nycomed Amersham plc. Milli-Q is a trademark of Millipore Corp.