Valco Instruments Co. Inc. TCD Instruction Manual page 10

Microvolume thermal conductivity detector
Table of Contents

Advertisement

Introduction
6
MAINS POWER connector
For connection to 115/230 VAC source.
CHART OUTPUT connector
Normally connected to a strip chart recorder. This output has an attenuated
range of 0-1 volt and 0 - 10 volts, with the signal scaled by the attenuation
factor set on the front panel. The output also has an internal signal reference
(-) at zero volts. For best noise performance, the shield (earth) and signal
reference (-) should not be connected together.
UNATTENUATED OUTPUT connector
Normally connected to a data acquisition system or other recording means.
For convenience, full-scale 0 - 1V and 0 - 10V outputs are provided, with an
internal signal reference (-) at zero volts. For best noise performance, the
shield (earth) and signal reference (-) should not be connected together.
DETECTOR connector
For connection to the detector control and heating system.
NOTE: These terminals are for connection only to
equipment having no accessible live parts.
Figure 4: Rear panel connectors