arteriocyte magellan Instructions For Use Manual

Complete disposable kit
Hide thumbs Also See for magellan:

Advertisement

Quick Links

Complete Disposable Kit
Explanation of symbols on package labeling
l
Lot Number
j
Catalog Number
f
Use by
X
Quantity
k
Do Not Reuse
O
IQ
Fluid Path Sterilized Using Ethylene Oxide
O
IK
Fluid Path Sterilized Using Irradiation
O
IS
Fluid Path Sterilized Using Steam or Dry Heat
Not made with natural rubber latex
-
Nonpyrogenic
U.S.P.
United States Pharmacopeia
H
This Way Up
M
Manufacturer
F
Fragile, Handle with Care
Consult Instructions for Use
w
Caution, Consult Accompanying Documents
Caution: Federal law restricts this device to sale by or on
the order of a (licensed healthcare practitioner).
INTRODUCTION
Indications for Use
®
The Arteriocyte Medical Systems, Inc. MAGELLAN
Complete Disposable Kit is intended
®
for use only with the MAGELLAN
Autologous Platelet Separator Instrument. See
®
Magellan
Platelet Separator Indications for Use.
Contraindications
The Use of the Arteriocyte Medical Systems, Inc. MAGELLAN
contraindicated for a hemodynamically unstable or hypercoagulable patient.
Use of this product for pediatric patients should be approached carefully. Withdrawing
blood from pediatric patients should be at a physician's specific direction with attention
given to avoiding any significant reduction in the patient's blood volume.
Caution: Medications that adversely affect a patient's coagulation system may inhibit the
use of platelet separation system therapy.
Warning: Reprocessing may compromise the structural integrity of the device and/or lead
to device failure. Reuse of this single patient use device creates a potential risk of patient or
user infections. Contamination of the device may lead to injury, illness or death of the
patient.
Precautions
1. Only Arteriocyte Medical Systems sterilized disposable kits are approved for patient use
®
with the MAGELLAN
Platelet Separator Instrument. It is important that aseptic
technique be used to minimize the possibility of contamination of the disposable
components and/or patient.
2. Store all disposable components in a dry place away from extremes of environmental
conditions.
3. Materials used in the Arteriocyte Medical Systems, Inc. MAGELLAN
may be sensitive to chemicals (such as solvents and certain detergents). Under certain
adverse conditions, exposure to these chemicals (including vapors) may cause the
plastics to fail or malfunction. Visually inspect the contents of the disposable kit. Should
any evidence of damage to the components be found during inspection or setup, do not
© Arteriocyte Medical Systems, Inc.
All Rights Reserved
use the disposable components. Do not use silicone oils or greases near disposable
components.
System Description
The Arteriocyte Medical Systems, Inc. MAGELLAN
sterile fluid pathway single-patient-use components necessary for each platelet separation
procedure with the MAGELLAN
Cautions
1. Caution: Federal Law (USA) restricts this device to be sold by or on the order of
physician.
2. Refer to the System Operator's manual supplied with the instrument before performing
a platelet separation procedure using the components of this kit. Treat all blood and
fluids using Universal Precautions for bloodborne pathogens.
3. Each platelet separation procedure using the Arteriocyte Medical Systems Inc,
®
MAGELLAN
Platelet Separator Instrument requires one MAGELLAN
Separator Disposable Kit. The separation chamber and associated tubing can be used
with the same patient for up to three complete separation cycles.
4. Use only Arteriocyte Medical Systems disposable accessories.
5. Inspect the kit prior to use. Do not use the kit if any component or the tray is damaged
or opened.
6. Throughout procedure, make certain all tubing is free of any kinks, twists, or flat areas.
All components of this kit are sterile fluid pathway single-patient-use. Do not resterilize.
Do not use the ACD-A anticoagulant unless the solution is clear and the seal is intact.
The ACD-A anticoagulant supplied in this kit is not for preparation of blood products for
transfusion or for direct intravenous infusion. Discard the unused portion. Do not reuse.
How Supplied
Contents of the Arteriocyte Medical Systems, Inc. MAGELLAN
(see Figure 1):
A.
One (1) 50-mL bag ACD-A anticoagulant
B.
One (1) 60-mL syringe (Syringe 2)
C.
One (1) 10-mL syringe (Syringe 1)
D.
One (1) 5-mL syringe
E.
One (1) IV Site Prep Kit
F.
One (1) Separation chamber with attached tubing
G.
One (1) 18 gauge x 3.8 cm (1.5") needle
H.
One (1) phlebotomy needle set with luer-lock connector
I.
One (1) tubing clamp (For phlebotomy needle set)
J.
Four (4) syringe tip caps
K.
Instruction For Use
Note: If more than one separation cycle will be performed with the same patient, additional
1
10-mL and 60-mL
BD™ syringes are required and can be purchased from Arteriocyte
Medical Systems.
®
Disposables Kit is
Figure 1. Contents of the Arteriocyte Medical Systems, Inc. MAGELLAN
Disposable Kit.
INSTRUCTIONS
Caution: All components of this kit are sterile fluid pathway single-patient-use. Do not
resterilize.
1. Remove the cover from the separator disposable kit tray.
®
Disposables Kit
2. Be certain that all components are present and undamaged.
Caution: Do not use the kit if any component or the tray is damaged or open.
3. Remove the separation chamber package from the tray.
4. Peel open the lid on the chamber package.
®
Complete Disposable Kit consists of
®
Autologous Platelet Separator Instrument.
®
Platelet
®
Complete Disposable Kit
®
Complete
5. Holding the platelet separation chamber with the vent facing upward, thread the
attached tubing through the center of the chamber caddy (see Figure 2A and B).
Figure 2. Thread the separation chamber tubing through the center of the chamber
caddy.
6. Install the platelet separation chamber into the centrifuge caddy, making certain that
both ends of the chamber are properly located in the caddy notches (see Figure 3A).
Place T-connector in slot (see Figure 3B), and tubing under retainer on top surface of
caddy (see Figure 3C). Press tubing down into groove on outer edge of chamber caddy
(see Figure 3D).
Caution: Failure to install the separation chamber properly may result in error codes.
Figure 3. Install the separation chamber into the caddy and place tubing into slot and
under retainer.
7. Rotate the tubing collar so that its shape aligns with the opening in the support arm.
Slide the tubing collar into the support arm (see Figure 4A) and close the latch (see
Figure 4B).
Figure 4. Attach chamber tubing collar to the instrument tubing support arm.
8. Press tubing down into groove on support arm and place tubing through notch in
centrifuge ridge (see Figure 5).
Instructions for Use
Page 1 of 2
AMS95023LBL Rev8

Advertisement

loading
Need help?

Need help?

Do you have a question about the magellan and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for arteriocyte magellan

  • Page 1 Precautions ® Figure 1. Contents of the Arteriocyte Medical Systems, Inc. MAGELLAN Complete 1. Only Arteriocyte Medical Systems sterilized disposable kits are approved for patient use ® Disposable Kit. with the MAGELLAN Platelet Separator Instrument. It is important that aseptic...
  • Page 2 Arteriocyte Medical Systems, Inc. 15. Slowly draw the appropriate volume of patient blood. Gently mix with ACD-A throughout 2) The unused portion of the Product must be returned to Arteriocyte Medical Systems 45 South St. Hopkinton, MA 01748 USA - Internet: www.arteriocyte.com the blood draw for thorough distribution.
  • Page 3 AMS Confidential AMS95023LBL_REV 1_FC.fm 3/9/ 09 11:48 am AMS Confidential AMS Confidential ® MAGELLAN Ratio Dispenser Kit Instructions for Use AMS95125LBL Rev. 2 AMS95030LBL Rev. 2 M932798A001 Rev. 1.0...
  • Page 4 AMS Confidential AMS95125LBL Rev. 2 AMS95030LBL Rev. 2...
  • Page 5 AMS Confidential Explanation of symbols on package labeling AMS95125LBL Rev. 2 AMS95030LBL Rev. 2...
  • Page 6 AMS Confidential AMS95125LBL Rev. 2 AMS95030LBL Rev. 2...
  • Page 7 AMS Confidential AMS Confidential INTRODUCTION INTRODUCTION Intended use Intended use TO BE INSERTED AT THE END OF THE CONTRAINDICATIONS SECTION: (for English section) How Supplied Warning: Reprocessing may compromise the structural integrity of the device and/or lead to device failure. Reuse of this single patient use device creates a potential risk of patient or user infections.
  • Page 8 Accessories Cautions AMS Confidential INSTRUCTIONS 4. Snap the dispenser handle to the 12mL syringe (see Figure 1, number 1). English Instructions for Use 3 4 English Instructions for Use AMS95125LBL Rev. 2 AMS95030LBL Rev. 2...
  • Page 9: How Supplied

    AMS Confidential AMS Confidential INTRODUCTION Intended use How Supplied Accessories Cautions INSTRUCTIONS English Instructions for Use 3 AMS95125LBL Rev. 2 AMS95030CPS Rev. 1...
  • Page 10 AMS Confidential ARTERIOCYTE MEDICAL SYSTEMS INC. DISPOSABLES KIT LIMITED WARRANTY (U.S.) THE FOLLOWING LIMITED WARRANTY APPLIES TO UNITED STATES CUSTOMERS ONLY: Æ Arteriocyte Medical Systems, Inc. Magellan Ratio Dispenser 6 English Instructions for Use AMS95125LBL Rev. 2 AMS95030LBL Rev. 2...
  • Page 11 AMS Confidential English Instructions for Use 7 AMS95125LBL Rev. 2 AMS95030LBL Rev. 2...
  • Page 12 STATES. A. This LIMITED WARRANTY provides assurance for the customer who receives an Arteriocyte Medical Systems ® (“AMS”) Magellan® Ratio Dispenser Kit “Product”), that should the Product fail to function to specification, AMS will issues a credit equal to the original Product purchase price (but not to exceed the value of the replacement Product) against the purchase of any AMS replacement Product used for that customer.
  • Page 13 AMS Confidential AMS95125LBL Rev. 2 AMS95030LBL Rev. 2...
  • Page 14 AMS Confidential AMS Confidential Manufacturer: © Arteriocyte Medical Systems, Inc. 2011 AMS95125LBL Rev. 2 AMS95030LBL Rev. 2 AMS95023LBL AMS95125LBL Rev. 2 Rev. 2 AMS95030LBL Rev. 2...
  • Page 15 MAGELLAN ® Aspiration Kit Instructions For Use...
  • Page 17 Explanation of symbols on package labeling Caution, Consult Accompanying Documents Lot Number Catalog Number Sterilized Using Ethylene Oxide  Latex Free Nonpyrogenic Do Not Reuse Use by Fragile, Handle with Care This Way Up Quantity  DEHP Phthalate Open Here For US Audiences Only...
  • Page 18 Inspect the kit prior to use. Do not use the kit if any component of the tray is damaged or opened. Precautions Only Arteriocyte Medical Systems, Inc. sterilized disposable kits are approved for patient use with the MAGELLAN Autolo- ® gous Platelet Separator Instrument. It is important that aseptic technique be used to minimize the possibility of contamination of the disposable components and/or patient.
  • Page 19 The MAGELLAN Aspiration Kit consists of the following compo- ® nents: Two (2) 30 mL Syringes One (1) 60 mL Syringe One (1) 200 Micron Blood Component Filter One (1) Bone Marrow Aspiration Needle One (1) 3-Way Stopcock One (1) 18-G Needle...
  • Page 20 Disconnect the filtered blood component syringe from filter tub- ing and attach a sterile syringe cap. Follow AABB guidelines. Manufacturer: Arteriocyte Medical Systems, Inc. 45 South Street Hopkinton, MA 01748 Internet: www.arteriocyte.com Toll-free: 1-866-660-AMSI (2674) Fax: 508-497-8951...
  • Page 21: Intended Use

    Instructions for use of this Bone Marrow Aspiration Needle A. INTENDED USE The Bone Marrow Aspiration Needle is intended for use in aspirating bone marrow specimen. B. CONTRAINDICATIONS For use only for bone marrow aspiration as determined by a li- censed physician.
  • Page 22 Rotate the stylet cap counterclockwise, and pull the stylet out of the cannula. Attach a luer syringe to the cannula hub. Note: Avoid using glass syringes because they may result in an inadequate seal and may break and fragment. Apply negative pressure by quickly withdrawing the syringe plunger.
  • Page 23 (1) The Product must be used prior to its “Use By” date. (2) The unused portion of the Product must be returned to Arteriocyte Medical Systems and within 60 days after use and shall be the property of Arteriocyte Medical Systems.
  • Page 24 Systems to any representation, condition or warranty except this Limited Warranty. This Limited Warranty is provided by Arteriocyte Medical Systems, Inc., 45 South Street Hopkinton, MA 01748. It applies only in the United States. Areas outside the United States should contact their local Arteriocyte Medical Systems representative for exact terms of the...
  • Page 26 Manufacturer: Arteriocyte Medical Systems, Inc. 45 South Street Hopkinton, MA 01748 Internet: www.arteriocyte.com Toll-free: 1-866-660-AMSI (2674) Fax: 508-497-8951 ©Arteriocyte Medical Systems, Inc. All Rights Reserved AMS95140LBL Rev 2...